Literature DB >> 23411716

Angiographic CT: in vitro comparison of different carotid artery stents-does stent orientation matter?

Michael Lettau1, Martin Bendszus, Stefan Hähnel.   

Abstract

INTRODUCTION: Our aim was to evaluate the in vitro visualization of different carotid artery stents on angiographic CT (ACT). Of particular interest was the influence of stent orientation to the angiography system by measurement of artificial lumen narrowing (ALN) caused by the stent material within the stented vessel segment to determine whether ACT can be used to detect restenosis within the stent.
METHODS: ACT appearances of 17 carotid artery stents of different designs and sizes (4.0 to 11.0 mm) were investigated in vitro. Stents were placed in different orientations to the angiography system. Standard algorithm image reconstruction and stent-optimized algorithm image reconstruction was performed. For each stent, ALN was calculated.
RESULTS: With standard algorithm image reconstruction, ALN ranged from 19.0 to 43.6 %. With stent-optimized algorithm image reconstruction, ALN was significantly lower and ranged from 8.2 to 18.7 %. Stent struts could be visualized in all stents. Differences in ALN between the different stent orientations to the angiography system were not significant.
CONCLUSION: ACT evaluation of vessel patency after stent placement is possible but is impaired by ALN. Stent orientation of the stents to the angiography system did not significantly influence ALN. Stent-optimized algorithm image reconstruction decreases ALN but further research is required to define the visibility of in-stent stenosis depending on image reconstruction.

Entities:  

Mesh:

Year:  2013        PMID: 23411716     DOI: 10.1007/s00234-013-1152-y

Source DB:  PubMed          Journal:  Neuroradiology        ISSN: 0028-3940            Impact factor:   2.804


  25 in total

1.  Improved in-stent magnetic resonance angiography with high flip angle excitation.

Authors:  J M Meyer; A Buecker; E Spuentrup; K Schuermann; M Huetten; R D Hilgers; J J van Vaals; R W Guenther
Journal:  Invest Radiol       Date:  2001-11       Impact factor: 6.016

2.  Stent appearance at contrast-enhanced MR angiography: in vitro examination with 14 stents.

Authors:  M Lenhart; M Völk; C Manke; W R Nitz; M Strotzer; S Feuerbach; J Link
Journal:  Radiology       Date:  2000-10       Impact factor: 11.105

3.  Optimized intravenous Flat Detector CT for non-invasive visualization of intracranial stents: first results.

Authors:  Tobias Struffert; Stephan Kloska; Tobias Engelhorn; Yu Deuerling-Zheng; Sabine Ott; Marc Doelken; Marc Saake; Martin Köhrmann; Arnd Doerfler
Journal:  Eur Radiol       Date:  2010-08-14       Impact factor: 5.315

4.  Contrast-enhanced angiographic cone-beam CT of cerebrovascular stents: experimental optimization and clinical application.

Authors:  N V Patel; M J Gounis; A K Wakhloo; N Noordhoek; J Blijd; D Babic; D Takhtani; S-K Lee; A Norbash
Journal:  AJNR Am J Neuroradiol       Date:  2010-10-21       Impact factor: 3.825

5.  Carotid artery stents: in vitro comparison of different stent designs and sizes using CT angiography and contrast-enhanced MR angiography at 1.5T and 3T.

Authors:  M Lettau; A Sauer; S Heiland; S Rohde; M Bendszus; S Hähnel
Journal:  AJNR Am J Neuroradiol       Date:  2009-09-12       Impact factor: 3.825

6.  Flat-detector computed tomography in the assessment of intracranial stents: comparison with multi detector CT and conventional angiography in a new animal model.

Authors:  Tobias Struffert; Sabine Ott; Edyta Adamek; Marc Schwarz; Tobias Engelhorn; Stephan Kloska; Yu Deuerling-Zheng; Arnd Doerfler
Journal:  Eur Radiol       Date:  2011-03-02       Impact factor: 5.315

7.  64-slice multidetector coronary CT angiography: in vitro evaluation of 68 different stents.

Authors:  David Maintz; Harald Seifarth; Rainer Raupach; Thomas Flohr; Michael Rink; Torsten Sommer; Murat Ozgün; Walter Heindel; Roman Fischbach
Journal:  Eur Radiol       Date:  2005-12-07       Impact factor: 5.315

8.  Angiographic CT with intravenous contrast agent application for monitoring of intracranial flow diverting stents.

Authors:  Marc Saake; Tobias Struffert; Philipp Goelitz; Sabine Ott; Frank Seifert; Oliver Ganslandt; Arnd Doerfler
Journal:  Neuroradiology       Date:  2011-10-04       Impact factor: 2.804

9.  Silent embolism in diagnostic cerebral angiography and neurointerventional procedures: a prospective study.

Authors:  M Bendszus; M Koltzenburg; R Burger; M Warmuth-Metz; E Hofmann; L Solymosi
Journal:  Lancet       Date:  1999-11-06       Impact factor: 79.321

10.  In vitro comparison of different carotid artery stents: a pixel-by-pixel analysis using CT angiography and contrast-enhanced MR angiography at 1.5 and 3 T.

Authors:  Michael Lettau; Annett Sauer; Sabine Heiland; Stefan Rohde; Julia Reinhardt; Martin Bendszus; Stefan Hähnel
Journal:  Neuroradiology       Date:  2009-11-19       Impact factor: 2.804

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.