Abstract
In the era of communication technology, new options are now available for following-up patients implanted with pacemakers (PMs) and defibrillators (ICDs). Most major companies offer devices with wireless capabilities that communicate automatically with home transmitters, which then relay data to the physician, thereby allowing remote patient follow-up and monitoring. These systems are being widely used in the USA for remote follow-up, and have been more recently introduced in Europe, where their adoption is increasing. In this article, we describe the currently existing systems, review the available evidence in the literature regarding remote follow-up and monitoring of PMs and ICDs, and finally discuss some unresolved issues.Entities:
Mesh:
Year: 2009 PMID: 19470595 PMCID: PMC2686319 DOI: 10.1093/europace/eup110
Source DB: PubMed Journal: Europace ISSN: 1099-5129 Impact factor: 5.214
Minimum frequency of pacemakers/implantable cardioverter defibrillator/cardiac resynchronization therapy monitoring and possibilities of remote follow-up
| Direct follow-up | Remote follow-up | |
|---|---|---|
| Within 72 h of implantation | X | — |
| 2–12 weeks post-implantation | X | — |
| Every 3–12 months (PM/CRT-P) | X | X |
| Every 3–6 months (ICD/CRT-D) | X | X |
| Annually until battery depletion | X | — |
| Every 1–3 months at signs of battery depletion | X | X |
Adapted from Wilkoff et al.[1]
Comparison of different remote monitoring systems
| Biotronik Home Monitoring™ | Medtronic CareLink™ | Boston Scientific Latitude™ | St Jude Merlin.net™ | |
|---|---|---|---|---|
| Wireless communication with implanted device | Radiofrequency | Radiofrequency | Radiofrequency | Radiofrequency |
| Data transmission | GSM network | Analogue phoneline | Analogue phoneline | Analogue phoneline |
| Transmitter | Mobile | Stationary | Stationary | Stationary |
| Frequency of transmissions | Daily FU; Alert events | Scheduled FU; Alert events | Scheduled FU; Alert events | Scheduled FU; Alert events |
| Remote follow-up | Yes | Yes | Yes | Yes |
| Remote monitoring | Yes | Yes | Yes | Yes |
| Physician notification | SMS, e-mail, fax | SMS, e-mail | Fax, phone | Fax, e-mail, SMS |
| Feedback to patient via transmitter | LED indicating normal status or call to clinic | LED indicating normal status or call to clinic | Automatic text and audio messages | LED indicating call to clinic, automated phone calls |
| IEGM (real-time at remote follow-up) | 30 s (monthly periodic EGMs) | 10 s | 10 s | 30 s |
| IEGM (arrhythmic episodes) | All memorized episodes | All memorized episodes | All memorized episodes | All memorized episodes |
| Special features | Alerts fully configurable online | Automatic RA, RV and LV (only Consulta) pacing thresholds | Optional wireless weight scales and BP cuffs | Alerts fully configurable online |
| Automatic RV and LV thresholds (only Lumax 500/540) | Optivol® lung fluid status alert | Configurable data transmission to associated caregivers | Possibility of sending automated phone calls to patients | |
| Wireless PMs | Configurable red and yellow alerts | Configurable red and yellow alerts | Automatic RA, RV, and LV pacing thresholds (next generation of ICDs) | |
| Electronic health record data export capability |
RA, Right atrial; RV, right ventricular; LV, left ventricular; IEGM, intracardiac electrogram; BP, blood pressure.
Ongoing multicentre randomized studies on remote device monitoring and follow-up registered on www.clinicaltrials.gov
| Study | Sponsor | Device | Projected size | Primary outcome measures | Status |
|---|---|---|---|---|---|
| CONNECT[ | Medtronic | CRT-D, ICD-DR | 2000 | Reduction in time to clinical decision for arrhythmias, cardiovascular disease progression, and system issues | Active, not recruiting |
| effecT[ | Biotronik | CRT-D | 300 | Mortality and morbidity in patients with AF | Recruiting |
| EuroEco[ | Biotronik | ICD-VR, ICD-DR | 312 | Costs | Recruiting |
| EVATEL[ | French Ministry of Health | ICD-VR, ICD-DR | 1600 | Safety and cost-effectiveness | Recruiting |
| HomeCARE II[ | Biotronik | CRT-D, ICD-VR, ICD-DR | 300 | Performance of intrathoracic impedance for monitoring heart failure | Recruiting |
| IMPACT[ | Biotronik | CRT-D, ICD-DR | 2718 | Thromboembolic events and bleeding in patients with AF managed by pre-defined anticoagulation plan. | Recruiting |
| IN-TIME[ | Biotronik | CRT-D, ICD-VR, ICD-DR | 620 | Mortality, hospitalization for heart failure, clinical status | Recruiting |
| OptiLink-HF[ | Medtronic | CRT-D, ICD-VR, ICD-DR | 1000 | Intrathoracic impedance monitoring for reducing death or cardiovascular hospitalization | Recruiting |
| QUANTUM[ | Biotronik | ICD-VR, ICD-DR | 150 | Psychosomatic evaluation | Recruiting |
| TRUST[ | Biotronik | ICD-VR, ICD-DR | 1516 | Number of ICD follow-ups and safety | Active, not recruiting |
Data were retrieved on 22 February 2009.