Gamma inert sterilization: a solution to polyethylene oxidation?

Francisco J Medel1, Steven M Kurtz, William J Hozack, Javad Parvizi, James J Purtill, Peter F Sharkey, Daniel MacDonald, Matthew J Kraay, Victor Goldberg, Clare M Rimnac.   

Abstract

BACKGROUND: In the 1990s, oxidation was found to occur in ultra-high molecular weight polyethylene total joint replacement components following gamma irradiation and prolonged shelf aging in air. Orthopaedic manufacturers developed barrier packaging to reduce oxidation during and after radiation sterilization. The present study explores the hypothesis that polyethylene components sterilized in a low-oxygen environment undergo similar in vivo oxidative mechanisms as inserts sterilized in air. In addition, the potential influence of the different sterilization processes on the wear performance of the polyethylene components was examined.
METHODS: An analysis of oxidation, wear, and surface damage was performed for forty-eight acetabular liners and 123 tibial inserts. The mean implantation time was 12.3+/-3.7 years for thirty-one acetabular liners that had been gamma sterilized in air and 4.0+/-2.5 years for the seventeen acetabular liners that had been gamma sterilized in inert gas. The mean implantation time was 11.0+/-3.2 years for the twenty-six tibial inserts that had been sterilized in air and 2.8+/-2.2 years for the ninety-seven tibial inserts that had been gamma sterilized in inert gas. Oxidation and hydroperoxide levels were characterized in loaded and unloaded regions of the inserts.
RESULTS: Measurable oxidation and oxidation potential were observed in all cohorts. The oxidation and hydroperoxide levels were regional. Surfaces with access to body fluids were more heavily oxidized than protected bearing surfaces were. This variation appeared to be greater in historical (gamma-in-air-sterilized) components. Regarding wear performance, historical and conventional acetabular liners showed similar wear penetration rates, whereas a low incidence of delamination was confirmed for the conventional tibial inserts in the first decade of implantation.
CONCLUSIONS: The present study explores the impact of industry-wide changes in sterilization practices for polyethylene. We found lower oxidation and oxidation potential in the conventional acetabular liners and tibial inserts that had been gamma sterilized in inert gas as compared with the historical components that had been gamma sterilized in air. However, we also found strong evidence that conventional components undergo mechanisms of in vivo oxidation similar to those observed following gamma irradiation in air. In addition, gamma sterilization in inert gas did not provide polyethylene with a significant improvement in terms of wear resistance as compared with gamma sterilization in air, except for a lower incidence of delamination in the first decade of implantation for tibial inserts.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19339568      PMCID: PMC2665040          DOI: 10.2106/JBJS.H.00538

Source DB:  PubMed          Journal:  J Bone Joint Surg Am        ISSN: 0021-9355            Impact factor:   5.284


  28 in total

1.  Potential errors in FTIR measurement of oxidation in ultrahigh molecular weight polyethylene implants.

Authors:  F W Shen; Y J Yu; H McKellop
Journal:  J Biomed Mater Res       Date:  1999

2.  Effect of consolidation on adhesive and abrasive wear of ultra high molecular weight polyethylene.

Authors:  Rizwan M Gul; Frederick J McGarry; Charles R Bragdon; Orhun K Muratoglu; William H Harris
Journal:  Biomaterials       Date:  2003-08       Impact factor: 12.479

3.  Impact of gamma sterilization on clinical performance of polyethylene in the knee.

Authors:  J P Collier; D K Sperling; J H Currier; L C Sutula; K A Saum; M B Mayor
Journal:  J Arthroplasty       Date:  1996-06       Impact factor: 4.757

4.  Shelf life and in vivo duration. Impacts on performance of tibial bearings.

Authors:  B H Currier; J H Currier; J P Collier; M B Mayor; R D Scott
Journal:  Clin Orthop Relat Res       Date:  1997-09       Impact factor: 4.176

5.  Gamma sterilization of UHMWPE articular implants: an analysis of the oxidation problem. Ultra High Molecular Weight Poly Ethylene.

Authors:  V Premnath; W H Harris; M Jasty; E W Merrill
Journal:  Biomaterials       Date:  1996-09       Impact factor: 12.479

6.  The Otto Aufranc Award. Impact of gamma sterilization on clinical performance of polyethylene in the hip.

Authors:  L C Sutula; J P Collier; K A Saum; B H Currier; J H Currier; W M Sanford; M B Mayor; R E Wooding; D K Sperling; I R Williams
Journal:  Clin Orthop Relat Res       Date:  1995-10       Impact factor: 4.176

7.  Retrieval analysis of total knee prostheses: a method and its application to 48 total condylar prostheses.

Authors:  R W Hood; T M Wright; A H Burstein
Journal:  J Biomed Mater Res       Date:  1983-09

8.  Property changes of UHMW polyethylene hip cup endoprostheses during implantation.

Authors:  P Eyerer; Y C Ke
Journal:  J Biomed Mater Res       Date:  1984 Nov-Dec

9.  Analysis of retrieved implants: crystallinity changes in ultrahigh molecular weight polyethylene.

Authors:  E S Grood; R Shastri; C N Hopson
Journal:  J Biomed Mater Res       Date:  1982-07

10.  Oxidation in orthopaedic UHMWPE sterilized by gamma-radiation and ethylene oxide.

Authors:  L Costa; M P Luda; L Trossarelli; E M Brach del Prever; M Crova; P Gallinaro
Journal:  Biomaterials       Date:  1998 Apr-May       Impact factor: 12.479

View more
  16 in total

1.  In vivo oxidation contributes to delamination but not pitting in polyethylene components for total knee arthroplasty.

Authors:  Francisco J Medel; Steven M Kurtz; Javad Parvizi; Gregg R Klein; Matthew J Kraay; Clare M Rimnac
Journal:  J Arthroplasty       Date:  2010-09-28       Impact factor: 4.757

2.  Gamma-irradiation sterilization in an inert environment: a partial solution.

Authors:  Daniel J Berry; Barbara H Currier; Michael B Mayor; John P Collier
Journal:  Clin Orthop Relat Res       Date:  2012-07       Impact factor: 4.176

3.  Current modes of failure in TKA: infection, instability, and stiffness predominate.

Authors:  David H Le; Stuart B Goodman; William J Maloney; James I Huddleston
Journal:  Clin Orthop Relat Res       Date:  2014-03-11       Impact factor: 4.176

4.  Reasons for revision of first-generation highly cross-linked polyethylenes.

Authors:  Steven M Kurtz; Francisco J Medel; Daniel W MacDonald; Javad Parvizi; Matthew J Kraay; Clare M Rimnac
Journal:  J Arthroplasty       Date:  2010-06-11       Impact factor: 4.757

5.  In vivo oxidation and surface damage in retrieved ethylene oxide-sterilized total knee arthroplasties.

Authors:  Daniel MacDonald; Josa Hanzlik; Peter Sharkey; Javad Parvizi; Steven M Kurtz
Journal:  Clin Orthop Relat Res       Date:  2012-07       Impact factor: 4.176

6.  Do first-generation highly crosslinked polyethylenes oxidize in vivo?

Authors:  Daniel MacDonald; Ashlyn Sakona; Allyson Ianuzzi; Clare M Rimnac; Steven M Kurtz
Journal:  Clin Orthop Relat Res       Date:  2011-08       Impact factor: 4.176

7.  The effect of poly sterilization on wear, osteolysis and survivorship of a press-fit cup at 10-year followup.

Authors:  Charles A Engh; Cara C Powers; Henry Ho; Sarah E Beykirch-Padgett; Robert H Hopper; C Anderson Engh
Journal:  Clin Orthop Relat Res       Date:  2012-02       Impact factor: 4.176

8.  Polyethylene oxidation in total hip arthroplasty: evolution and new advances.

Authors:  Enrique Gómez-Barrena; Francisco Medel; José Antonio Puértolas
Journal:  Open Orthop J       Date:  2009-12-24

9.  Effects of Terminal Sterilization on PEG-Based Bioresorbable Polymers Used in Biomedical Applications.

Authors:  Divya Bhatnagar; Koustubh Dube; Vinod B Damodaran; Ganesan Subramanian; Kenneth Aston; Frederick Halperin; Meiyu Mao; Kurt Pricer; N Sanjeeva Murthy; Joachim Kohn
Journal:  Macromol Mater Eng       Date:  2016-07-11       Impact factor: 4.367

10.  Long-Term Implant Survivorship and Modes of Failure in Simultaneous Concurrent Bilateral Total Knee Arthroplasty.

Authors:  Taylor M Yong; Emily C Young; Ilda B Molloy; Brian M Fisher; Benjamin J Keeney; Wayne E Moschetti
Journal:  J Arthroplasty       Date:  2019-08-14       Impact factor: 4.757

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.