Literature DB >> 20875942

In vivo oxidation contributes to delamination but not pitting in polyethylene components for total knee arthroplasty.

Francisco J Medel1, Steven M Kurtz, Javad Parvizi, Gregg R Klein, Matthew J Kraay, Clare M Rimnac.   

Abstract

The aim of this study was to better understand how in vivo oxidation contributes to fatigue damage in total knee arthroplasty (TKA). A total of 119 tibial inserts were consecutively collected after revision surgery. Of the 119 polyethylene retrievals, 29 were gamma sterilized in air (historical), whereas the remaining 90 were gamma sterilized in nitrogen (conventional). Surface damage assessment and characterization of oxidation were performed on all the retrievals. Delamination was significantly more prevalent and extensive in the longer-term, highly oxidized, historical tibial inserts. Pitting damage, in contrast, seemed to be equally prevalent between both retrieval groups and was not correlated with in vivo oxidation. Our findings support our hypothesis that in vivo oxidation is a contributing factor to delamination, but not pitting, in TKA. Despite the lower oxidation displayed by conventional retrievals, this study provides strong evidence that delamination secondary to in vivo oxidation may occur during the second decade of implantation.
Copyright © 2011 Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20875942      PMCID: PMC3385511          DOI: 10.1016/j.arth.2010.07.010

Source DB:  PubMed          Journal:  J Arthroplasty        ISSN: 0883-5403            Impact factor:   4.757


  38 in total

Review 1.  Wear of ultra-high molecular weight polyethylene (UHMWPE) in total knee prostheses: a review of key influences.

Authors:  T M McGloughlin; A G Kavanagh
Journal:  Proc Inst Mech Eng H       Date:  2000       Impact factor: 1.617

2.  The causes of insert backside wear in total knee arthroplasty.

Authors:  Ray C Wasielewski
Journal:  Clin Orthop Relat Res       Date:  2002-11       Impact factor: 4.176

3.  The effect of conformity, thickness, and material on stresses in ultra-high molecular weight components for total joint replacement.

Authors:  D L Bartel; V L Bicknell; T M Wright
Journal:  J Bone Joint Surg Am       Date:  1986-09       Impact factor: 5.284

4.  Retrieval analysis of total knee prostheses: a method and its application to 48 total condylar prostheses.

Authors:  R W Hood; T M Wright; A H Burstein
Journal:  J Biomed Mater Res       Date:  1983-09

5.  Backside wear of Miller-Galante I and Insall-Burstein II tibial inserts.

Authors:  Naoya Taki; Victor M Goldberg; Matthew J Kraay; Clare M Rimnac
Journal:  Clin Orthop Relat Res       Date:  2004-11       Impact factor: 4.176

Review 6.  Polyethylene wear in total hip and knee arthroplasties.

Authors:  G Lewis
Journal:  J Biomed Mater Res       Date:  1997

7.  Evaluation of oxidation and fatigue damage of retrieved crossfire polyethylene acetabular cups.

Authors:  Barbara H Currier; John H Currier; Michael B Mayor; Kimberly A Lyford; John P Collier; Douglas W Van Citters
Journal:  J Bone Joint Surg Am       Date:  2007-09       Impact factor: 5.284

Review 8.  Osteolysis after total knee arthroplasty.

Authors:  Sanjay K Gupta; Alice Chu; Amar S Ranawat; John Slamin; Chitranjan S Ranawat
Journal:  J Arthroplasty       Date:  2007-09       Impact factor: 4.757

9.  Effect of fabrication method and resin type on performance of tibial bearings.

Authors:  B H Currier; J H Currier; J P Collier; M B Mayor
Journal:  J Biomed Mater Res       Date:  2000

10.  Rapid polyethylene failure of unicondylar tibial components sterilized with gamma irradiation in air and implanted after a long shelf life.

Authors:  Thomas F McGovern; Deborah J Ammeen; John P Collier; Barbara H Currier; Gerard A Engh
Journal:  J Bone Joint Surg Am       Date:  2002-06       Impact factor: 5.284

View more
  10 in total

1.  Surface damage versus tibial polyethylene insert conformity: a retrieval study.

Authors:  Markus A Wimmer; Michel P Laurent; Jeannie D Haman; Joshua J Jacobs; Jorge O Galante
Journal:  Clin Orthop Relat Res       Date:  2012-07       Impact factor: 4.176

2.  Experimental testing of total knee replacements with UHMW-PE inserts: impact of severe wear test conditions.

Authors:  Carmen Zietz; Joern Reinders; Jens Schwiesau; Alexander Paulus; Jan Philippe Kretzer; Thomas Grupp; Sandra Utzschneider; Rainer Bader
Journal:  J Mater Sci Mater Med       Date:  2015-02-26       Impact factor: 3.896

3.  Oxidative properties and surface damage mechanisms of remelted highly crosslinked polyethylenes in total knee arthroplasty.

Authors:  Daniel W MacDonald; Genymphas Higgs; Javad Parvizi; Gregg Klein; Mark Hartzband; Harlan Levine; Matthew Kraay; Clare M Rimnac; Steven M Kurtz
Journal:  Int Orthop       Date:  2013-02-10       Impact factor: 3.075

4.  Correlation between in vivo stresses and oxidation of UHMWPE in total hip arthroplasty.

Authors:  M Regis; P Bracco; L Giorgini; S Fusi; P Dalla Pria; L Costa; C Schmid
Journal:  J Mater Sci Mater Med       Date:  2014-06-29       Impact factor: 3.896

5.  In vivo oxidation and surface damage in retrieved ethylene oxide-sterilized total knee arthroplasties.

Authors:  Daniel MacDonald; Josa Hanzlik; Peter Sharkey; Javad Parvizi; Steven M Kurtz
Journal:  Clin Orthop Relat Res       Date:  2012-07       Impact factor: 4.176

Review 6.  Highly cross-linked polyethylene may not have an advantage in total knee arthroplasty.

Authors:  Vasileios I Sakellariou; Peter Sculco; Lazaros Poultsides; Timothy Wright; Thomas P Sculco
Journal:  HSS J       Date:  2013-08-10

7.  Quantification of structural changes of UHMWPE components in total joint replacements.

Authors:  Petr Fulin; David Pokorny; Miroslav Slouf; Martina Nevoralova; Tatana Vackova; Jiri Dybal; Jan Pilar
Journal:  BMC Musculoskelet Disord       Date:  2014-03-28       Impact factor: 2.362

8.  Can pin-on-disk testing be used to assess the wear performance of retrieved UHMWPE components for total joint arthroplasty?

Authors:  Steven M Kurtz; Daniel W MacDonald; Sevi Kocagöz; Mariya Tohfafarosh; Doruk Baykal
Journal:  Biomed Res Int       Date:  2014-09-11       Impact factor: 3.411

9.  Wear behavior of an unstable knee: stabilization via implant design?

Authors:  Jörn Reinders; Robert Sonntag; Jan Philippe Kretzer
Journal:  Biomed Res Int       Date:  2014-09-09       Impact factor: 3.411

10.  Retrieval analysis of Harris-Galante I and II acetabular liners in situ for more than 10 years.

Authors:  Keisha French; Rebecca Moore; Heather Gawel; Steven M Kurtz; Matthew J Kraay; Ke Xie; Victor M Goldberg; Clare M Rimnac
Journal:  Acta Orthop       Date:  2012-08-10       Impact factor: 3.717

  10 in total