Gamma-irradiation sterilization in an inert environment: a partial solution.

Daniel J Berry1, Barbara H Currier, Michael B Mayor, John P Collier.   

Abstract

BACKGROUND: In the mid to late 1990 s, to sterilize UHMWPE bearings, manufacturers changed from gamma-irradiation-in-air (gamma-air) sterilization, which initiated oxidation leading to bearing fatigue, to gamma-irradiation sterilization in an inert environment (gamma-inert). The change to gamma-inert sterilization reportedly prevented shelf oxidation before implantation but not in vivo oxidation. QUESTIONS/PURPOSES: We asked: (1) Has the change to gamma-inert sterilization prevented shelf oxidation that led to early in vivo fatigue damage in gamma-air-sterilized tibial inserts? And (2) has the change to gamma-inert sterilization prevented the occurrence of fatigue secondary to in vivo oxidation?
METHODS: We rated 183 retrieved gamma-air- and 175 retrieved gamma-inert-sterilized tibial inserts for clinical fatigue damage and analyzed 132 gamma-air- and 174 gamma-inert-sterilized tibial inserts for oxidation by Fourier transform infrared spectroscopy.
RESULTS: Oxidation led to decreased mechanical properties in shelf-aged gamma-air-sterilized tibial inserts. Barrier packaging prevented shelf oxidation in gamma-inert-sterilized tibial inserts. Gamma-air- and gamma-inert-sterilized inserts oxidized in vivo. Fatigue damage (delamination) occurred more frequently in inserts retrieved after longer time in vivo. Longer in vivo time correlated with higher oxidation and more accumulated cycles of use.
CONCLUSIONS: Published oxidation projections suggest gamma-inert-sterilized tibial inserts would reach the critical oxidation for the onset of fatigue after 11 to 14 years in vivo. These retrievals appear to follow the projected oxidation trends. Frequency of fatigue damage increased with increasing oxidation. CLINICAL RELEVANCE: Fatigue of tibial inserts becomes more likely, especially in active patients, after more than a decade of good clinical performance.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22033874      PMCID: PMC3369088          DOI: 10.1007/s11999-011-2150-1

Source DB:  PubMed          Journal:  Clin Orthop Relat Res        ISSN: 0009-921X            Impact factor:   4.176


  11 in total

Review 1.  Advances in the processing, sterilization, and crosslinking of ultra-high molecular weight polyethylene for total joint arthroplasty.

Authors:  S M Kurtz; O K Muratoglu; M Evans; A A Edidin
Journal:  Biomaterials       Date:  1999-09       Impact factor: 12.479

2.  Current sterilization and packaging methods for polyethylene.

Authors:  L S Bargmann; B C Bargmann; J P Collier; B H Currier; M B Mayor
Journal:  Clin Orthop Relat Res       Date:  1999-12       Impact factor: 4.176

3.  Gamma inert sterilization: a solution to polyethylene oxidation?

Authors:  Francisco J Medel; Steven M Kurtz; William J Hozack; Javad Parvizi; James J Purtill; Peter F Sharkey; Daniel MacDonald; Matthew J Kraay; Victor Goldberg; Clare M Rimnac
Journal:  J Bone Joint Surg Am       Date:  2009-04       Impact factor: 5.284

4.  Impact of gamma sterilization on clinical performance of polyethylene in the knee.

Authors:  J P Collier; D K Sperling; J H Currier; L C Sutula; K A Saum; M B Mayor
Journal:  J Arthroplasty       Date:  1996-06       Impact factor: 4.757

5.  Shelf life and in vivo duration. Impacts on performance of tibial bearings.

Authors:  B H Currier; J H Currier; J P Collier; M B Mayor; R D Scott
Journal:  Clin Orthop Relat Res       Date:  1997-09       Impact factor: 4.176

6.  Retrieval analysis of total knee prostheses: a method and its application to 48 total condylar prostheses.

Authors:  R W Hood; T M Wright; A H Burstein
Journal:  J Biomed Mater Res       Date:  1983-09

7.  The effect of alignment and BMI on failure of total knee replacement.

Authors:  Merrill A Ritter; Kenneth E Davis; John B Meding; Jeffery L Pierson; Michael E Berend; Robert A Malinzak
Journal:  J Bone Joint Surg Am       Date:  2011-09-07       Impact factor: 5.284

8.  Evaluation of oxidation and fatigue damage of retrieved crossfire polyethylene acetabular cups.

Authors:  Barbara H Currier; John H Currier; Michael B Mayor; Kimberly A Lyford; John P Collier; Douglas W Van Citters
Journal:  J Bone Joint Surg Am       Date:  2007-09       Impact factor: 5.284

9.  In vivo oxidation of gamma-barrier-sterilized ultra-high-molecular-weight polyethylene bearings.

Authors:  Barbara H Currier; John H Currier; Michael B Mayor; Kimberly A Lyford; Douglas W Van Citters; John P Collier
Journal:  J Arthroplasty       Date:  2007-03-28       Impact factor: 4.757

10.  Oxidation and oxidation potential in contemporary packaging for polyethylene total joint replacement components.

Authors:  Luigi Costa; Pierangiola Bracco; Elena Maria Brach del Prever; Steven M Kurtz; Paolo Gallinaro
Journal:  J Biomed Mater Res B Appl Biomater       Date:  2006-07       Impact factor: 3.368

View more
  6 in total

1.  A Novel Technique for Assessing Antioxidant Concentration in Retrieved UHMWPE.

Authors:  Barbara H Currier; Douglas W Van Citters
Journal:  Clin Orthop Relat Res       Date:  2017-05       Impact factor: 4.176

2.  Low-Dose Gamma Irradiation of Decellularized Heart Valves Results in Tissue Injury In Vitro and In Vivo.

Authors:  Meghana R K Helder; Ryan S Hennessy; Daniel B Spoon; Brandon J Tefft; Tyra A Witt; Ronald J Marler; Sorin V Pislaru; Robert D Simari; John M Stulak; Amir Lerman
Journal:  Ann Thorac Surg       Date:  2015-10-09       Impact factor: 4.330

3.  Can pin-on-disk testing be used to assess the wear performance of retrieved UHMWPE components for total joint arthroplasty?

Authors:  Steven M Kurtz; Daniel W MacDonald; Sevi Kocagöz; Mariya Tohfafarosh; Doruk Baykal
Journal:  Biomed Res Int       Date:  2014-09-11       Impact factor: 3.411

4.  The influence of malalignment and ageing following sterilisation by gamma irradiation in an inert atmosphere on the wear of ultra-high-molecular-weight polyethylene in patellofemoral replacements.

Authors:  Raman Maiti; Raelene M Cowie; John Fisher; Louise M Jennings
Journal:  Proc Inst Mech Eng H       Date:  2017-03-25       Impact factor: 1.617

5.  Tobacco Exposure Is Associated With Extremely Low Polyethylene Oxidation in Total Knee Arthroplasty Components.

Authors:  Samuel T Kunkel; Wayne E Moschetti; Paul Werth; Yale Fillingham; David Jevsevar; Douglas VanCitters; John Currier; Barbara Currier; Eric Henderson
Journal:  Arthroplast Today       Date:  2021-04-16

6.  Influence of Irradiation Temperature on Oxidative and Network Properties of X-Ray Cross-Linked Vitamin E Stabilized UHMWPE for Hip Arthroplasty.

Authors:  M A Mulliez; C Schilling; T M Grupp
Journal:  Biomed Res Int       Date:  2020-03-23       Impact factor: 3.411

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.