Central corneal thickness measured by the Orbscan II system, contact ultrasound pachymetry, and the Artemis 2 system.

Tania Paul1, Mira Lim, Christopher E Starr, Harriet O Lloyd, D Jackson Coleman, Ronald H Silverman.   

Abstract

PURPOSE: To compare central corneal thickness (CCT) measurements by the Orbscan II device, contact ultrasound (US) pachymetry, and the noncontact Artemis 2 scanning US system.
SETTING: Department of Ophthalmology, Weill Cornell Medical College, New York, New York, USA.
METHODS: The CCT in 40 eyes (20 normal subjects) was measured by the Orbscan II followed by contact US pachymetry and then the Artemis 2. Results were compared using analysis of variance (ANOVA), paired t tests, and Bland-Altman plots.
RESULTS: There was a significant difference in CCT measurements between the 3 modes (F = 32.84, P = .0001, 1-way ANOVA). Artemis 2 and US pachymetry measurements were highly correlated (r2 = 0.963, P < .0001), although Artemis 2 values were a mean of 11.2 microm +/- 6.6 (SD) thinner than pachymetry values. Artemis 2 and Orbscan II measurements were less well correlated (r2 = 0.851, P < .001); Orbscan II values were a mean of 7.5 +/- 15.7 microm thinner than Artemis 2 values. Orbscan II values showed a trend toward increasing underestimation of CCT in thinner corneas.
CONCLUSIONS: Ultrasound pachymetry and Artemis 2 CCT measurements were highly correlated; the 11 microm mean difference in measurements may be attributed to decentration, oblique incidence of the probe to the cornea, or possibly the effect of topical anesthesia with contact pachymetry. Although the mean difference between Orbscan II and Artemis 2 values was 7.5 microm, Orbscan values were less correlated than Artemis 2 values with contact US pachymetry and were prone to underestimation of the CCT in thinner corneas.

Entities:  

Mesh:

Year:  2008        PMID: 19006737      PMCID: PMC2601704          DOI: 10.1016/j.jcrs.2008.07.013

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  31 in total

1.  Comparison of corneal pachymetry using ultrasound and Orbscan II.

Authors:  Mohamed A Fakhry; Alberto Artola; José I Belda; Ma José Ayala; Jorge L Alió
Journal:  J Cataract Refract Surg       Date:  2002-02       Impact factor: 3.351

2.  Accuracy of Orbscan pachymetry measurements and DHG ultrasound pachymetry in primary laser in situ keratomileusis and LASIK enhancement procedures.

Authors:  N G Iskander; E Anderson Penno; N T Peters; H V Gimbel; M Ferensowicz
Journal:  J Cataract Refract Surg       Date:  2001-05       Impact factor: 3.351

3.  Reliability of pachymetric measurements using orbscan after excimer refractive surgery.

Authors:  Olivier Prisant; Nila Calderon; Philippe Chastang; Damien Gatinel; Thanh Hoang-Xuan
Journal:  Ophthalmology       Date:  2003-03       Impact factor: 12.079

4.  A comparison of two pachymetric systems: slit-scanning and ultrasonic.

Authors:  Maria Jesus Giraldez Fernandez; Alberto Diaz Rey; Alejandro Cerviño; Eva Yebra-Pimentel
Journal:  CLAO J       Date:  2002-10

5.  Central and peripheral corneal thickness measurement with Orbscan II and topographical ultrasound pachymetry.

Authors:  Jose Manuel González-Méijome; Alejandro Cerviño; Eva Yebra-Pimentel; Manuel A Parafita
Journal:  J Cataract Refract Surg       Date:  2003-01       Impact factor: 3.351

6.  Comparison of central corneal thickness measurements by specular microscopy, ultrasound pachymetry, and ultrasound biomicroscopy.

Authors:  Eric S Tam; David S Rootman
Journal:  J Cataract Refract Surg       Date:  2003-06       Impact factor: 3.351

7.  Corneal thickness values before and after oxybuprocaine 0.4% eye drops.

Authors:  Isabel Asensio; Saleh M Rahhal; Luis Alonso; José M Palanca-Sanfrancisco; Juan A Sanchis-Gimeno
Journal:  Cornea       Date:  2003-08       Impact factor: 2.651

8.  Central corneal thickness measurement with a retinal optical coherence tomography device versus standard ultrasonic pachymetry.

Authors:  M Bechmann; M J Thiel; A S Neubauer; S Ullrich; K Ludwig; K R Kenyon; M W Ulbig
Journal:  Cornea       Date:  2001-01       Impact factor: 2.651

9.  Effect of central corneal thickness on Goldmann applanation tonometry measures - a different result with different pachymeters.

Authors:  Michael J Doughty; Sven Jonuscheit
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2007-05-24       Impact factor: 3.117

10.  Arc-scanning very high-frequency digital ultrasound for 3D pachymetric mapping of the corneal epithelium and stroma in laser in situ keratomileusis.

Authors:  D Z Reinstein; R H Silverman; T Raevsky; G J Simoni; H O Lloyd; D J Najafi; M J Rondeau; D J Coleman
Journal:  J Refract Surg       Date:  2000 Jul-Aug       Impact factor: 3.573

View more
  12 in total

1.  Comparison of central corneal thickness measurements with three new optical devices and a standard ultrasonic pachymeter.

Authors:  Hasan Ali Bayhan; Seray Aslan Bayhan; Izzet Can
Journal:  Int J Ophthalmol       Date:  2014-04-18       Impact factor: 1.779

2.  Repeatability of layered corneal pachymetry with the artemis very high-frequency digital ultrasound arc-scanner.

Authors:  Dan Z Reinstein; Timothy J Archer; Marine Gobbe; Ronald H Silverman; D Jackson Coleman
Journal:  J Refract Surg       Date:  2009-11-16       Impact factor: 3.573

3.  Comparison of artemis 2 ultrasound and Visante optical coherence tomography corneal thickness profiles.

Authors:  Roxana Ursea; Matthew Feng; Raksha Urs; Arindam RoyChoudhury; Ronald H Silverman
Journal:  J Refract Surg       Date:  2012-11-30       Impact factor: 3.573

Review 4.  High-resolution ultrasound imaging of the eye - a review.

Authors:  Ronald H Silverman
Journal:  Clin Exp Ophthalmol       Date:  2008-12-09       Impact factor: 4.207

5.  Comparison of central corneal thickness with four different optical devices.

Authors:  Kuddusi Teberik; Mehmet Tahir Eski; Murat Kaya; Handan Ankaralı
Journal:  Int Ophthalmol       Date:  2017-10-11       Impact factor: 2.031

6.  Repeatability and interobserver reproducibility of Artemis-2 high-frequency ultrasound in determination of human corneal thickness.

Authors:  Kelechi C Ogbuehi; Uchechukwu L Osuagwu
Journal:  Clin Ophthalmol       Date:  2012-05-21

7.  Central Corneal Thickness in Highly Myopic Eyes: Inter-device Agreement of Ultrasonic Pachymetry, Pentacam and Orbscan II Before and After Photorefractive Keratectomy.

Authors:  Nader Nassiri; Kourosh Sheibani; Sare Safi; Saman Nassiri; Alireza Ziaei; Alireza Ziaee; Farnaz Haji; Shiva Mehravaran; Nariman Nassiri
Journal:  J Ophthalmic Vis Res       Date:  2014-01

8.  Role of percent peripheral tissue ablated on refractive outcomes following hyperopic LASIK.

Authors:  George Fatseas; Fiona Stapleton; Patrick Versace
Journal:  PLoS One       Date:  2017-02-02       Impact factor: 3.240

9.  Comparison of central corneal thickness measurements using ultrasound pachymetry, ultrasound biomicroscopy, and the Artemis-2 VHF scanner in normal eyes.

Authors:  Haya M Al-Farhan; Wafa'a Majed Al-Otaibi
Journal:  Clin Ophthalmol       Date:  2012-07-06

10.  Comparison of high-resolution Scheimpflug and high-frequency ultrasound biomicroscopy to anterior-segment OCT corneal thickness measurements.

Authors:  Anastasios John Kanellopoulos; George Asimellis
Journal:  Clin Ophthalmol       Date:  2013-11-20
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.