Kelechi C Ogbuehi1, Uchechukwu L Osuagwu.
Abstract
BACKGROUND: The purpose of this study was to assess the repeatability and limits of agreement of corneal thickness values measured by a high-frequency ultrasound (Artemis-2), hand-held ultrasound pachymeter (DGH-500) and a specular microscope (SP-3000P).Entities:
Keywords: Artemis-2 ultrasound; SP-3000P noncontact specular microscopy; cornea thickness; hand-held ultrasound; reproducibility
Year: 2012 PMID: 22693418 PMCID: PMC3367432 DOI: 10.2147/OPTH.S31690
Source DB: PubMed Journal: Clin Ophthalmol ISSN: 1177-5467
Mean central corneal thickness taken with Topcon SP-3000P noncontact specular microscope, ultrasound pachymeter DGH-500, and Artemis-2 very high-frequency ultrasound
| Device | Mean CCT μm (±SD) | Minimum | Maximum | 95% CI of mean |
|---|---|---|---|---|
| SP-3000P | 520 ± 29 | 466 | 567 | 509–530 |
| DGH-500 | 560 ± 33 | 507 | 613 | 548–571 |
| Artemis-2 | 558 ± 33 | 502 | 619 | 546–569 |
| SP-3000P | 520 ± 30 | 453 | 572 | 508–531 |
| USP | 561 ± 37 | 486 | 621 | 548–574 |
| Artemis-2 | 554 ± 34 | 506 | 616 | 542–566 |
Abbreviations: CCT, central corneal thickness; CI, confidence interval; SD, standard deviation; USP, ultrasound pachymeter.
Bland and Altman analysis of central corneal thickness measurements obtained by SP-3000P, Artemis-2, and DGH-500 ultrasound pachymetry
| Techniques | Mean difference (μm) | SD | 95% CI (μm) | CoR | |
|---|---|---|---|---|---|
| SP-3000P | −0.9 | 4.7 | −10.0 to 8.2 | >0.05 | ± 9.1 |
| SP-3000P | −0.9 | 5.5 | −11.6 to 9.9 | ± 10.8 | |
| Artemis-2 | −0.5 | 4.1 | −8.6 to 7.6 | >0.05 | ± 8.1 |
| Artemis-2 | −0.9 | 4.6 | −9.9 to 8.1 | ± 9.0 | |
| DGH-500 | 0.6 | 6.3 | −11.8 to 13.0 | >0.05 | ± 12.4 |
| DGH-500 | −1.7 | 6.2 | −13.9 to 10.4 | ± 12.1 |
Notes: Mean difference within session, standard deviations, 95% confidence intervals (mean ± 1.96 SD) in each technique with their CoR* are represented. P value is the between-observer paired t-test analysis.
SP-3000P is first session measurement;
SP-3000P is second session measurement;
statistically significant at the 0.05.
Abbreviations: CI, confidence interval; CoR, coefficients of repeatability; SD, standard deviation.
Figure 1Bland-Altman plot of mean difference in central corneal thickness measurements between Artemis-2 and SP-3000P noncontact specular microscope against their mean in session 1 and session 2.
Notes: Mean difference in session 1, P < 0.001; session 2, P < 0.001.
Abbreviations: CCT, central corneal thickness; CI, confidence interval.
Figure 2Bland-Altman plot of mean difference in central corneal thickness measurements between ultrasound pachymetry and Artemis-2 against their mean in session 1 and session 2.
Notes: Mean difference in session 1, P > 0.05; session 2, P > 0.05.
Abbreviations: CCT, central corneal thickness; CI, confidence interval.
Figure 3Bland-Altman plot of mean difference in central corneal thickness measurements between ultrasound pachymetry and SP-3000P noncontact specular microscope against their mean in session 1 and session 2.
Notes: Mean difference in session 1, P < 0.001; session 2, P < 0.001.
Abbreviations: CCT, central corneal thickness; CI, confidence interval.