Effect of central corneal thickness on Goldmann applanation tonometry measures - a different result with different pachymeters.

Michael J Doughty1, Sven Jonuscheit.   

Abstract

PURPOSE: To assess whether the use of Orbscan pachymetry, as an alternative to ultrasound pachymetry, had an effect on the IOP-CCT relationship for Goldmann tonometry in normal human eyes. PATIENTS AND METHODS: The right eye of 50 healthy subjects, aged 19 to 82 years, were assessed by Orbscan (three scans), and then central corneal thickness (CCT) measured by ultrasound pachymetry (three readings) and Goldmann tonometry performed (three readings).
RESULTS: Goldmann tonometry indicated average IOP values from 9.7 to 20 mm Hg (group mean 14.3 +/- 2.5 mm Hg) that were highly highly correlated with ultrasound pachymetry (r = 0.608, p < 0.001). Orbscan central zone pachymetry (2-mm sample zone) yielded higher values than ultrasound, averaging 0.591 +/- 0.044 mm compared to 0.523 +/- 0.037 mm (p < 0.001), with the differences between the instruments being proportional to the average thickness (p < 0.001; r = 0.461). Notwithstanding, the Goldmann tonometry values were still highly correlated with the Orbscan central zone thickness data (r = 0.595, p < 0.001); but, due to the difference in the thickness data generated by the two pachymeters, the absolute slope of the IOP-CCT relationship was slightly less if Orbscan measures were used, with or without use of the acoustic factor.
CONCLUSION: Orbscan measures of the thickness of a central corneal zone can be used to assess the impact of central corneal thickness on Goldmann tonometry data.

Entities:  

Mesh:

Year:  2007        PMID: 17522884     DOI: 10.1007/s00417-007-0601-x

Source DB:  PubMed          Journal:  Graefes Arch Clin Exp Ophthalmol        ISSN: 0721-832X            Impact factor:   3.117


  21 in total

1.  Corneal indentation during ultrasonic pachometry.

Authors:  O D Solomon
Journal:  Cornea       Date:  1999-03       Impact factor: 2.651

2.  Central corneal thickness in European (white) individuals, especially children and the elderly, and assessment of its possible importance in clinical measures of intra-ocular pressure.

Authors:  Michael J Doughty; Mohammed Laiquzzaman; Andreas Müller; Emil Oblak; Norman F Button
Journal:  Ophthalmic Physiol Opt       Date:  2002-11       Impact factor: 3.117

3.  [Applanation tonometry].

Authors:  H GOLDMANN; T SCHMIDT
Journal:  Ophthalmologica       Date:  1957-10       Impact factor: 3.250

4.  Corneal pachymetry in normal and keratoconic eyes: Orbscan II versus ultrasound.

Authors:  D Gherghel; S L Hosking; S Mantry; S Banerjee; S A Naroo; S Shah
Journal:  J Cataract Refract Surg       Date:  2004-06       Impact factor: 3.351

Review 5.  Orbscan computerized topography: attributes, applications, and limitations.

Authors:  Gerard Cairns; Charles N J McGhee
Journal:  J Cataract Refract Surg       Date:  2005-01       Impact factor: 3.351

6.  On the use of Orbscan II to assess the peripheral corneal thickness in humans: a comparison with ultrasound pachometry measures.

Authors:  Sven Jonuscheit; Michael J Doughty; Norman F Button
Journal:  Ophthalmic Physiol Opt       Date:  2007-03       Impact factor: 3.117

7.  The relative effects of corneal thickness and age on Goldmann applanation tonometry and dynamic contour tonometry.

Authors:  A Kotecha; E T White; J M Shewry; D F Garway-Heath
Journal:  Br J Ophthalmol       Date:  2005-12       Impact factor: 4.638

Review 8.  Human corneal thickness and its impact on intraocular pressure measures: a review and meta-analysis approach.

Authors:  M J Doughty; M L Zaman
Journal:  Surv Ophthalmol       Date:  2000 Mar-Apr       Impact factor: 6.048

Review 9.  Sources of error with use of Goldmann-type tonometers.

Authors:  M M Whitacre; R Stein
Journal:  Surv Ophthalmol       Date:  1993 Jul-Aug       Impact factor: 6.048

10.  Central and paracentral corneal pachymetry in patients with normal tension glaucoma and ocular hypertension.

Authors:  Jens F Jordan; Silke Joergens; Sven Dinslage; Thomas S Dietlein; Günter K Krieglstein
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2005-08-02       Impact factor: 3.535

View more
  5 in total

1.  An implantable microfluidic device for self-monitoring of intraocular pressure.

Authors:  Ismail E Araci; Baolong Su; Stephen R Quake; Yossi Mandel
Journal:  Nat Med       Date:  2014-08-24       Impact factor: 53.440

2.  [Mobile intraocular pressure measurement. From palpation to initial clinical experience with the handheld dynamic contour tonometer].

Authors:  U Schmid; C Kniestedt
Journal:  Ophthalmologe       Date:  2010-07       Impact factor: 1.059

3.  Central corneal thickness measured by the Orbscan II system, contact ultrasound pachymetry, and the Artemis 2 system.

Authors:  Tania Paul; Mira Lim; Christopher E Starr; Harriet O Lloyd; D Jackson Coleman; Ronald H Silverman
Journal:  J Cataract Refract Surg       Date:  2008-11       Impact factor: 3.351

Review 4.  A Meta-Analysis of Wearable Contact Lenses for Medical Applications: Role of Electrospun Fiber for Drug Delivery.

Authors:  Hamed Hosseinian; Samira Hosseini; Sergio O Martinez-Chapa; Mazhar Sher
Journal:  Polymers (Basel)       Date:  2022-01-03       Impact factor: 4.329

5.  Corneal biomechanical parameters and intraocular pressure: the effect of topical anesthesia.

Authors:  Kelechi C Ogbuehi
Journal:  Clin Ophthalmol       Date:  2012-06-15
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.