Effect of 2% fluorescein on Scheimpflug central corneal thickness measurements.

Stella Briggs1, Marwa Bin Moammar2.   

Abstract

AIM: To assess central corneal thickness (CCT) changes measured with Scheimpflug device following instillation of 2% fluorescein in normal subjects.
METHODS: This was a prospective randomized study of 60 hospital volunteers. After baseline CCT measurements of both eyes of 40 subjects were obtained using Scheimpflug system, a drop of preservative-free 2% fluorescein, was instilled in one eye and in other eye, one drop of normal saline (control). Measurements were repeated after 1, 2, 5, 10, 20, 30, 40, 50 and 60min (continuous assessment group). Twenty subjects had baseline CCT taken, then fluorescein was instilled in one eye and measurements were taken at 1min. Ten eyes had saline rinse after 1min and 10 other eyes did not, measurements were repeated at 2min (eye rinse group).
RESULTS: The mean baseline CCT for continuous assessment group was 546.2 ±32.1 µm (range, 489.0-606.0), control eyes was 546.6±30.7 µm (range, 489.0-602.0). At 1min after fluorescein instillation, CCT significantly increased by 37.0±34.0 µm (P<0.001), then decreased gradually, reaching baseline at 60min. CCT variations were not significant in control group (P>0.05). For eye rinse group, CCT mean differences between baseline and 2min were 18.2 µm (95 % CI: -54.7 to 18.3) with rinse and 26.5 µm (95% CI: -62.9 to 9.9) without rinse; paired sample tests were not significant (P>0.05).
CONCLUSION: The presence of fluorescein increased CCT value to a clinically relevant level of 6.8%. Eye rinse did not significantly reduce the effect at 2min post fluorescein timepoint.

Entities:  

Keywords:  Scheimpflug device; central corneal thickness; fluorescein

Year:  2016        PMID: 26949642      PMCID: PMC4761734          DOI: 10.18240/ijo.2016.02.10

Source DB:  PubMed          Journal:  Int J Ophthalmol        ISSN: 2222-3959            Impact factor:   1.779


  15 in total

1.  The repeatability of corneal thickness measures.

Authors:  M W Marsich; M A Bullimore
Journal:  Cornea       Date:  2000-11       Impact factor: 2.651

2.  Comparison of corneal pachymetry using ultrasound and Orbscan II.

Authors:  Mohamed A Fakhry; Alberto Artola; José I Belda; Ma José Ayala; Jorge L Alió
Journal:  J Cataract Refract Surg       Date:  2002-02       Impact factor: 3.351

3.  Accuracy of Orbscan optical pachymetry in corneas with haze.

Authors:  Francesco Boscia; Maria Gabriella La Tegola; Giovanni Alessio; Carlo Sborgia
Journal:  J Cataract Refract Surg       Date:  2002-02       Impact factor: 3.351

4.  Effect of fluorescein dye staining of the tear film on Scheimpflug measurements of central corneal thickness.

Authors:  Nino Hirnschall; Alja Crnej; Vinod Gangwani; Oliver Findl
Journal:  Cornea       Date:  2012-01       Impact factor: 2.651

5.  Repeatability and reproducibility of central corneal thickness measurement with Pentacam, Orbscan, and ultrasound.

Authors:  Birgit Lackner; Gerald Schmidinger; Stefan Pieh; Martin A Funovics; Christian Skorpik
Journal:  Optom Vis Sci       Date:  2005-10       Impact factor: 1.973

6.  Comparison of central corneal thickness measurements by rotating Scheimpflug camera, ultrasonic pachymetry, and scanning-slit corneal topography.

Authors:  Shiro Amano; Norihiko Honda; Yuki Amano; Satoru Yamagami; Takashi Miyai; Tomokazu Samejima; Miyuki Ogata; Kazunori Miyata
Journal:  Ophthalmology       Date:  2006-06       Impact factor: 12.079

7.  A clinical comparison of optical and ultrasonic pachometry.

Authors:  J Nissen; J O Hjortdal; N Ehlers; K Frost-Larsen; T Sørensen
Journal:  Acta Ophthalmol (Copenh)       Date:  1991-10

8.  Keratoconus: spatial variation of corneal thickenss as a diagnostic test.

Authors:  R B Mandell; K A Polse
Journal:  Arch Ophthalmol       Date:  1969-08

9.  A novel method for pachymetry mapping of human precorneal tear film using Pentacam with fluorescein.

Authors:  Hong Zhuang; Xingtao Zhou; Jianjiang Xu
Journal:  Invest Ophthalmol Vis Sci       Date:  2009-08-26       Impact factor: 4.799

10.  Relation between optical coherence tomography and optical pachymetry measurements of corneal swelling induced by hypoxia.

Authors:  Jianhua Wang; Desmond Fonn; Trefford L Simpson; Lyndon Jones
Journal:  Am J Ophthalmol       Date:  2002-07       Impact factor: 5.258

View more
  1 in total

1.  Effect of Proparacaine 0.375%-Sodium Fluorescein 0.25% Eye Drop Mixture and Fluorescein Strip on Anterior Segment Parameters.

Authors:  Mustafa Doğan; Mehmet Cem Sabaner; Mehmet Akif Erol
Journal:  J Ophthalmol       Date:  2018-09-05       Impact factor: 1.909

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.