Literature DB >> 18773260

Comparison of central corneal thickness measurements by Orbscan II and Pentacam after corneal refractive surgery.

Jumpei Matsuda1,2, Osamu Hieda3,4, Shigeru Kinoshita3.   

Abstract

PURPOSE: To compare the accuracy of postoperative pachymetry between Orbscan II (Bausch & Lomb) scanning-slit corneal topography/pachymetry and the Pentacam (Oculus) rotating Scheimpflug camera.
METHODS: Central corneal thickness (CCT) was determined in 24 patients (48 eyes) before and after laser in situ keratomileusis (LASIK) or Epipolis LASIK (Epi-LASIK) procedures. All eyes were examined by Orbscan II and Pentacam prior to refractive surgery and at the first, fourth, and twelfth week postoperatively. The residual CCT (RCCT) measured by each instrument was compared to the theoretical RCCT.
RESULTS: On the first, fourth, and twelfth week after the refractive surgery, the mean RCCT measurements by Orbscan II were 413 +/- 72, 435 +/- 65, and 440 +/- 69 microm, respectively. Those of Pentacam were 434 +/- 51, 436 +/- 53, and 438 +/- 50 microm, respectively. Orbscan II measurements at the postoperative first and fourth week were significantly smaller than the theoretical RCCT (P < 0.01, P < 0.01, paired t test). There was no statistical difference between the theoretical RCCT and the Pentacam measurements at any stage.
CONCLUSION: The Orbscan II measurement values of postoperative corneas tended to be thinner than the theoretical values, but not those of the Pentacam.

Entities:  

Mesh:

Year:  2008        PMID: 18773260     DOI: 10.1007/s10384-008-0550-x

Source DB:  PubMed          Journal:  Jpn J Ophthalmol        ISSN: 0021-5155            Impact factor:   2.447


  31 in total

1.  Factors affecting the forward shift of posterior corneal surface after laser in situ keratomileusis.

Authors:  T Baek; K Lee; F Kagaya; A Tomidokoro; S Amano; T Oshika
Journal:  Ophthalmology       Date:  2001-02       Impact factor: 12.079

2.  Accuracy of Orbscan pachymetry measurements and DHG ultrasound pachymetry in primary laser in situ keratomileusis and LASIK enhancement procedures.

Authors:  N G Iskander; E Anderson Penno; N T Peters; H V Gimbel; M Ferensowicz
Journal:  J Cataract Refract Surg       Date:  2001-05       Impact factor: 3.351

3.  Reliability of pachymetric measurements using orbscan after excimer refractive surgery.

Authors:  Olivier Prisant; Nila Calderon; Philippe Chastang; Damien Gatinel; Thanh Hoang-Xuan
Journal:  Ophthalmology       Date:  2003-03       Impact factor: 12.079

4.  Anterior chamber depth measurements in phakic and pseudophakic eyes: Pentacam versus ultrasound device.

Authors:  Gabor Nemeth; Attila Vajas; Bence Kolozsvari; Andras Berta; Laszlo Modis
Journal:  J Cataract Refract Surg       Date:  2006-08       Impact factor: 3.351

5.  Comparison of central corneal thickness measurements by rotating Scheimpflug camera, ultrasonic pachymetry, and scanning-slit corneal topography.

Authors:  Shiro Amano; Norihiko Honda; Yuki Amano; Satoru Yamagami; Takashi Miyai; Tomokazu Samejima; Miyuki Ogata; Kazunori Miyata
Journal:  Ophthalmology       Date:  2006-06       Impact factor: 12.079

6.  Central corneal thickness measurements using Orbscan II, Visante, ultrasound, and Pentacam pachymetry after laser in situ keratomileusis for myopia.

Authors:  Thomas Ho; Arthur C K Cheng; Srinivas K Rao; Silvania Lau; Chris K S Leung; Dennis S C Lam
Journal:  J Cataract Refract Surg       Date:  2007-07       Impact factor: 3.351

7.  Corneal changes after laser refractive surgery for myopia: comparison of Orbscan II and Pentacam findings.

Authors:  Hassan Hashemi; Shiva Mehravaran
Journal:  J Cataract Refract Surg       Date:  2007-05       Impact factor: 3.351

8.  Changes in the lens epithelium with respect to cataractogenesis: light microscopic and Scheimpflug densitometric analysis of the cataractous and the clear lens of diabetics and non-diabetics.

Authors:  Sergey I Tkachov; Christine Lautenschläger; Dirk Ehrich; Hans Gert Struck
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2005-09-21       Impact factor: 3.117

9.  Comparison of three methods of measuring corneal thickness and anterior chamber depth.

Authors:  Wolf Buehl; Danijela Stojanac; Stefan Sacu; Wolfgang Drexler; Oliver Findl
Journal:  Am J Ophthalmol       Date:  2006-01       Impact factor: 5.258

10.  Iatrogenic keratectasia after laser in situ keratomileusis.

Authors:  T Seiler; K Koufala; G Richter
Journal:  J Refract Surg       Date:  1998 May-Jun       Impact factor: 3.573

View more
  3 in total

1.  Comparison of corneal measurements in keratoconic eyes using rotating Scheimpflug camera and scanning-slit topography.

Authors:  Mohammad Naderan; Saeed Shoar; Morteza Naderan; Mohammad Amin Kamaleddin; Mohammad Taher Rajabi
Journal:  Int J Ophthalmol       Date:  2015-04-18       Impact factor: 1.779

2.  The distribution of orbscan indices in young population.

Authors:  Hassan Hashemi; Mehdi Khabazkhoob; Negareh Yazdani; Hadi Ostadimoghaddam; Payam Nabovati; Raheleh Moravej; AbbasAli Yekta
Journal:  J Curr Ophthalmol       Date:  2016-08-31

3.  Central Corneal Thickness Measurement Using Ultrasonic Pachymetry, Rotating Scheimpflug Camera, and Scanning-slit Topography Exclusively in Thin Non-keratoconic Corneas.

Authors:  Mehrdad Mohammadpour; Kazem Mohammad; Nasser Karimi
Journal:  J Ophthalmic Vis Res       Date:  2016 Jul-Sep
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.