Laser interference biometry versus ultrasound biometry in certain clinical conditions.

B A M Lege1, W Haigis.   

Abstract

PURPOSE: To compare laser interference biometry (LIB) with conventional ultrasound biometry in certain clinical conditions such as globe deformities, eccentric fixation, retinal detachment, macular edema or silicone oil-filled eyes.
SETTING: Department of Ophthalmology, Würzburg University Eye Hospital, Germany.
METHODS: We evaluated all patients who came to our university hospital for axial length measurement with our routine immersion biometry system (Grieshaber Biometry System) and compared the results with those obtained using the Zeiss IOLMaster of Carl Zeiss Jena, the commercially available LIB device.
RESULTS: Selected case reports demonstrate the advantages and disadvantages of LIB. Advantages of LIB were found in patients with asymmetrically shaped globes, eccentric fixation, silicone oil-filled eyes and a fearful/nervous disposition. Disadvantages of the system were revealed in cases of retinal detachment, severe opacities along the visual axis and poor patient cooperation.
CONCLUSION: We showed that LIB is a valuable addition to the choice of biometric devices, when used with medical understanding.

Entities:  

Mesh:

Year:  2003        PMID: 14648138     DOI: 10.1007/s00417-003-0672-2

Source DB:  PubMed          Journal:  Graefes Arch Clin Exp Ophthalmol        ISSN: 0721-832X            Impact factor:   3.117


  6 in total

1.  Pseudophakic correction factors for optical biometry.

Authors:  W Haigis
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2001-08       Impact factor: 3.117

2.  Mêasurement of corneal thickness by low-coherence interferometry.

Authors:  C K Hitzenberger
Journal:  Appl Opt       Date:  1992-11-01       Impact factor: 1.980

3.  Optical coherence tomography.

Authors:  A F Fercher
Journal:  J Biomed Opt       Date:  1996-04       Impact factor: 3.170

4.  Axial length measurements and fundus changes of the myopic eye.

Authors:  B J Curtin; D B Karlin
Journal:  Am J Ophthalmol       Date:  1971-01       Impact factor: 5.258

5.  Measurement of the axial length of cataract eyes by laser Doppler interferometry.

Authors:  C K Hitzenberger; W Drexler; C Dolezal; F Skorpik; M Juchem; A F Fercher; H D Gnad
Journal:  Invest Ophthalmol Vis Sci       Date:  1993-05       Impact factor: 4.799

6.  Comparison of immersion ultrasound biometry and partial coherence interferometry for intraocular lens calculation according to Haigis.

Authors:  W Haigis; B Lege; N Miller; B Schneider
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2000-09       Impact factor: 3.117

  6 in total
  17 in total

1.  Misleading axial length measurements with the IOLMaster due to a dense posterior vitreous surface membrane and a macular edema in a diabetic patient.

Authors:  Christoph Fidel Mayer; Navid Ardjomand; Werner Wackernagel; Michaela Velikay-Parel
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2011-11-03       Impact factor: 3.117

2.  Relationship between the retinal thickness of the macula and the difference in axial length.

Authors:  Tetsuo Ueda; Yoshiaki Nawa; Yoshiaki Hara
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2005-08-13       Impact factor: 3.117

3.  Prevalence of corneal astigmatism before cataract surgery.

Authors:  Mehran Mohammadi; Mohammad Naderan; Rozhin Pahlevani; Ali Jahanrad
Journal:  Int Ophthalmol       Date:  2016-02-24       Impact factor: 2.031

4.  A study to evaluate whether CTR increases refractive unpredictability between predicted and actual IOL position.

Authors:  V K Baranwal; Santosh Kumar; Avinash Mishra; Ajay K Dutta
Journal:  Med J Armed Forces India       Date:  2012-10-31

5.  Biometry and refractive outcome of eyes filled with silicone oil by standardized echography and partial coherence interferometry.

Authors:  Johannes Nepp; Katharina Krepler; Kerstin Jandrasits; Wolfgang Hauff; Georg Hanselmayer; Michaela Velikay-Parel; Karl C Ossoinig; Andreas Wedrich
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2005-10-20       Impact factor: 3.117

6.  Accuracy of axial length measurements from immersion B-scan ultrasonography in highly myopic eyes.

Authors:  Qing-Hua Yang; Bing Chen; Guang-Hua Peng; Zhao-Hui Li; Yi-Fei Huang
Journal:  Int J Ophthalmol       Date:  2014-06-18       Impact factor: 1.779

7.  Optical biometry intraocular lens power calculation using different formulas in patients with different axial lengths.

Authors:  Jia-Kang Wang; Shu-Wen Chang
Journal:  Int J Ophthalmol       Date:  2013-04-18       Impact factor: 1.779

8.  Partial Coherence Laser Interferometry in Highly Myopic versus Emmetropic Eyes.

Authors:  Gernot F Roessler; Yassin Djalali Talab; Thomas S Dietlein; Sven Dinslage; Niklas Plange; Peter Walter; Babac Ae Mazinani
Journal:  J Ophthalmic Vis Res       Date:  2014-04

9.  Comparison of two optical biometers in intraocular lens power calculation.

Authors:  Sheng Hui; Lu Yi
Journal:  Indian J Ophthalmol       Date:  2014-09       Impact factor: 1.848

10.  Comparison of a new optical biometer and a standard biometer in cataract patients.

Authors:  Pipat Kongsap
Journal:  Eye Vis (Lond)       Date:  2016-10-17