Literature DB >> 11585315

Pseudophakic correction factors for optical biometry.

W Haigis1.   

Abstract

INTRODUCTION: The IOLMaster of Carl Zeiss Jena, which has recently become available, is a combined instrument for biometry and intraocular lens (IOL) planning for cataract surgery utilizing partial coherence interferometry for measuring axial length. Whereas measurement data from classical ultrasound biometry, e.g. in pseudophakic eyes, need to be corrected by +0.4 to -0.8 mm--depending on the lens material--smaller corrections are expected to be necessary in optical biometry. Correction factors for various modern IOL materials were estimated theoretically and checked in first clinical measurements.
METHODS: Starting from the dispersion relation of PMMA and manufacturers' phase refractive index data at 546 nm, the group refractive indices of different IOL materials were estimated for the IOLMaster wavelength of lambda=780 nm. Then, for an average eye of 23.48 mm, the center thicknesses of emmetropizing lenses of different material were calculated. Finally, comparing the pseudophakic optical axial lengths thus deduced with the respective phakic value, individual material-specific correction factors were obtained.
RESULTS: Expressing the true axial length ALtrue by the length ALphak measured in phakic mode and a correction factor delta (ALtrue=Alphak+delta), we obtained values of delta=0.12+/-0.01 mm for all IOL materials considered (PMMA, silicone, collamer, AcrySof, MemoryLens). For aphakic eyes, delta=0.20 mm was deduced. DISCUSSION: The calculations suffered from the fact that hardly any information relating to optical material specifications of IOLs is available from lens manufacturers. Therefore, calculations had to be based on--nevertheless realistic--assumptions. Early clinical results support our theoretical findings. Thus, optical biometry seems to be more reliable and forgiving in pseudophakic eyes than classical ultrasound.

Entities:  

Mesh:

Substances:

Year:  2001        PMID: 11585315     DOI: 10.1007/s004170100323

Source DB:  PubMed          Journal:  Graefes Arch Clin Exp Ophthalmol        ISSN: 0721-832X            Impact factor:   3.117


  3 in total

1.  [Precision of a new device for biometric measurements in pseudophakic eyes].

Authors:  A L Hildebrandt; G U Auffarth; M P Holzer
Journal:  Ophthalmologe       Date:  2011-08       Impact factor: 1.059

2.  Laser interference biometry versus ultrasound biometry in certain clinical conditions.

Authors:  B A M Lege; W Haigis
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2003-11-25       Impact factor: 3.117

3.  Axial Length Change in Pseudophakic Eyes Measured by IOLMaster 700.

Authors:  Jiaqing Zhang; Zhenzhen Liu; Xiaozhang Qiu; Ling Jin; Lanhua Wang; Guangming Jin; Wei Wang; Xuhua Tan; Lixia Luo; Yizhi Liu
Journal:  Transl Vis Sci Technol       Date:  2021-05-03       Impact factor: 3.283

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.