Measurement of the axial length of cataract eyes by laser Doppler interferometry.

C K Hitzenberger1, W Drexler, C Dolezal, F Skorpik, M Juchem, A F Fercher, H D Gnad.   

Abstract

PURPOSE: To examine the applicability of the recently developed laser Doppler interferometry technique for measuring the axial length of cataract eyes in a realistic clinical situation. To determine the performance of the instrument as a function of cataract grade. To compare the results to those of ultrasound methods.
METHODS: A total of 196 cataract eyes of 100 patients were examined. The axial eye length was determined by laser Doppler interferometry and by two different ultrasound techniques, the applanation technique and the immersion technique. The cataract grade was determined by a commercial instrument that measures backscattered light.
RESULTS: Laser Doppler interferometry worked very well except in the cases of the highest cataract grades (4% of the eyes of this study were not measurable because of a too-high lens density). Only 3.5% of the other eyes were not measurable because of fixation problems of the patients. The precision of laser Doppler interferometry is not influenced by the cataract grade (except the highest grade). The standard deviation of the geometric eye length is approximately 20 microns. Linear regression analysis revealed a very good correlation of laser Doppler interferometry and ultrasonic measurements, but a systematic difference was found. The eye lengths measured by laser Doppler interferometry were about 0.18 mm longer than those measured by the immersion technique and about 0.47 mm longer than those measured by the applanation technique.
CONCLUSION: These differences are attributed to the laser Doppler interferometry results including the retinal thickness and indentation of the cornea by the applanation technique. The main advantages of the laser Doppler interferometry technique are high precision, high accuracy, and more comfort for the patient because it is a noncontact method, anesthesia is unnecessary, and the risk of corneal infection is avoided.

Entities:  

Mesh:

Year:  1993        PMID: 8491541

Source DB:  PubMed          Journal:  Invest Ophthalmol Vis Sci        ISSN: 0146-0404            Impact factor:   4.799


  14 in total

1.  IOLMaster biometry: refractive results of 100 consecutive cases.

Authors:  H Eleftheriadis
Journal:  Br J Ophthalmol       Date:  2003-08       Impact factor: 4.638

2.  Relationship between the retinal thickness of the macula and the difference in axial length.

Authors:  Tetsuo Ueda; Yoshiaki Nawa; Yoshiaki Hara
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2005-08-13       Impact factor: 3.117

3.  A new non-contact optical device for ocular biometry.

Authors:  J Santodomingo-Rubido; E A H Mallen; B Gilmartin; J S Wolffsohn
Journal:  Br J Ophthalmol       Date:  2002-04       Impact factor: 4.638

4.  Comparison of intraocular lens power prediction using immersion ultrasound and optical biometry with and without formula optimization.

Authors:  Gabor Nemeth; Attila Nagy; Andras Berta; Laszlo Modis
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2012-04-13       Impact factor: 3.117

5.  Agreement of IOL power and axial length obtained by IOLMaster 500 vs IOLMaster 500 with Sonolink connection.

Authors:  Sabong Srivannaboon; Chareenun Chirapapaisan; Patchara Nantasri; Mathinee Chongchareon; Pratuangsri Chonpimai
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2013-01-12       Impact factor: 3.117

6.  Reproducibility of a long-range swept-source optical coherence tomography ocular biometry system and comparison with clinical biometers.

Authors:  Ireneusz Grulkowski; Jonathan J Liu; Jason Y Zhang; Benjamin Potsaid; Vijaysekhar Jayaraman; Alex E Cable; Jay S Duker; James G Fujimoto
Journal:  Ophthalmology       Date:  2013-06-04       Impact factor: 12.079

7.  Visualization of 3-D high speed ultrahigh resolution optical coherence tomographic data identifies structures visible in 2D frames.

Authors:  Larry Kagemann; Hiroshi Ishikawa; Gadi Wollstein; Michelle Gabriele; Joel S Schuman
Journal:  Opt Express       Date:  2009-03-02       Impact factor: 3.894

8.  Laser interference biometry versus ultrasound biometry in certain clinical conditions.

Authors:  B A M Lege; W Haigis
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2003-11-25       Impact factor: 3.117

9.  Agreement and relationship between ultrasonic and partial coherence interferometry measurements of axial length and anterior chamber depth.

Authors:  Amany R Wissa; Sherein S Wahba; Maged M Roshdy
Journal:  Clin Ophthalmol       Date:  2012-01-31

10.  Refractive predictability of partial coherence interferometry and factors that can affect it.

Authors:  Seung Mo Kim; Joohyun Choi; Sangkyung Choi
Journal:  Korean J Ophthalmol       Date:  2009-03-09
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.