Biometry and refractive outcome of eyes filled with silicone oil by standardized echography and partial coherence interferometry.

Johannes Nepp1, Katharina Krepler, Kerstin Jandrasits, Wolfgang Hauff, Georg Hanselmayer, Michaela Velikay-Parel, Karl C Ossoinig, Andreas Wedrich.   

Abstract

BACKGROUND: We evaluated the validity of calculations for refractive outcome in cataract surgery in silicone oil-filled eyes. The retrosilicone space (RSS) was included in these calculations.
METHODS: In a prospective study the axial length (AL) of silicone oil-filled eyes was measured. with standardized A-scan echography (SAE) and partial coherence interferometry (PCI). Meldrum's formula was used to transform the velocity of ultrasound within the vitreous cavity. To investigate whether refractive outcome can be calculated accurately, we assessed the difference between precalculated and final refractive outcome. Furthermore, we determined the advantages and disadvantages of SAE and PCI. A minor aim was to assess whether the AL of the two eyes differed significantly.
RESULTS: In 85% of 117 eyes the difference between precalculated and postsurgical refraction was smaller than 1 diopter spherical and statistically not significant (p>0.2). The mean AL was 24.1 mm (range 20.0-31.4 mm). The difference in outcome between the two methods was without statistical significance: the AL difference was 0.4 (+/-2.6) mm on measurement with SAE and 0.04 (+/-0.46) mm with PCI. PCI has the advantage that it can be performed more easily, without contact, while echography is advantageous in the presence of advanced cataracts. In supine position an oil-free fluid space behind the silicone oil was detected with echography. The mean dimension of this space was 1.9 (+/-0.67) mm and it was taken into consideration for IOL calculation. The mean AL difference between the two eyes was 0.4 mm, but the difference was greater than 1 mm in 26% of the patients.
CONCLUSION: The AL of eyes filled with silicone oil can be measured reliably with SAE and PCI. In supine position the RSS has to be considered to obtain more accurate IOL calculations.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 15838667     DOI: 10.1007/s00417-004-1117-2

Source DB:  PubMed          Journal:  Graefes Arch Clin Exp Ophthalmol        ISSN: 0721-832X            Impact factor:   3.117


  16 in total

1.  Measurement of axial length of eyes with incomplete filling of silicone oil in the vitreous cavity using x ray computed tomography.

Authors:  Kazuo Takei; Yasuo Sekine; Fumiki Okamoto; Sachiko Hommura
Journal:  Br J Ophthalmol       Date:  2002-01       Impact factor: 4.638

2.  Evaluation of the practicality of optical biometry and applanation ultrasound in 253 eyes.

Authors:  Mana Tehrani; Frank Krummenauer; Eric Blom; H Burkhard Dick
Journal:  J Cataract Refract Surg       Date:  2003-04       Impact factor: 3.351

3.  Solid tissue model for the standardization of the echo-ophthalmograph 7200 MA (Kretztechnik).

Authors:  P Till
Journal:  Doc Ophthalmol       Date:  1976-10-15       Impact factor: 2.379

4.  Optics of the eye with air or silicone oil.

Authors:  E Stefánsson; J Tiedeman
Journal:  Retina       Date:  1988       Impact factor: 4.256

Review 5.  Standardized echography: basic principles, clinical applications, and results.

Authors:  K C Ossoinig
Journal:  Int Ophthalmol Clin       Date:  1979

6.  [Objective determination of refractive changes in silicon oil filled eyes--effect of head position on refraction].

Authors:  B Dick; S Pavlovic; F K Jacobi; D Eisenmann; K W Jacobi
Journal:  Klin Monbl Augenheilkd       Date:  1996-06       Impact factor: 0.700

7.  [Biometry--an exact method for the measurement of the axial length of the eye].

Authors:  W Hauff
Journal:  Wien Klin Wochenschr       Date:  1983-04-15       Impact factor: 1.704

8.  Cataract extraction after silicone oil repair of retinal detachments due to necrotizing retinitis.

Authors:  M L Meldrum; T M Aaberg; A Patel; J Davis
Journal:  Arch Ophthalmol       Date:  1996-07

9.  Partial coherence interferometry: a novel approach to biometry in cataract surgery.

Authors:  W Drexler; O Findl; R Menapace; G Rainer; C Vass; C K Hitzenberger; A F Fercher
Journal:  Am J Ophthalmol       Date:  1998-10       Impact factor: 5.258

10.  Cataract extraction in eyes filled with silicone oil.

Authors:  J Moisseiev; E Bartov; M Cahane; M Blumenthal; G Treister
Journal:  Arch Ophthalmol       Date:  1992-11
View more
  3 in total

1.  Changes in visual acuity, axial length, and refraction after removal of intraocular silicone oil following retinal reattachment surgery in Chinese patients: An open-label, prospective study.

Authors:  An-Huai Yang; Wei Jin; Yi-Qiao Xing
Journal:  Curr Ther Res Clin Exp       Date:  2009-06

2.  Accuracy and reliability of IOL master and A-scan immersion biometry in silicone oil-filled eyes.

Authors:  P Kunavisarut; P Poopattanakul; C Intarated; K Pathanapitoon
Journal:  Eye (Lond)       Date:  2012-08-10       Impact factor: 3.775

3.  Differences in Axial Length and IOL Power Based on Alternative A-Scan or Fellow-Eye Biometry in Macula-Off Rhegmatogenous Retinal Detachment Eyes.

Authors:  Rui Liu; Hongrong Li; Qingchen Li
Journal:  Ophthalmol Ther       Date:  2021-12-08
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.