Accuracy of axial length measurements from immersion B-scan ultrasonography in highly myopic eyes.

Qing-Hua Yang1, Bing Chen1, Guang-Hua Peng1, Zhao-Hui Li1, Yi-Fei Huang1.   

Abstract

AIM: To evaluate the accuracy of axial length (AL) measurements obtained from immersion B-scan ultrasonography (immersion B-scan) for intraocular lens (IOL) power calculation in patients with high myopia and cataracts.
METHODS: Immersion B-scan, contact A-scan ultrasonography (contact A-scan), and the IOLMaster were used to preoperatively measure the AL in 102 eyes from 102 patients who underwent phacoemulsification and IOL implantation. Patients were divided into two groups according to the AL: one containing patients with 22 mm≤AL<26 mm(group A) and the other containing patients with AL≥26 mm (group B). The mean error (ME) was calculated from the difference between the AL measurement methods predicted refractive error and the actual postoperative refractive error.
RESULTS: In group A, ALs measured by immersion B-scan (23.48±1.15) didn't differ significantly from those measured by the IOLMaster (23.52±1.17) or from those by contact A-scan (23.38±1.20). In the same group, the standard deviation (SD) of the mean error (ME) of immersion B-scan (-0.090±0.397 D) didn't differ significantly from those of IOLMaster (-0.095±0.411 D) and contact A-scan (-0.099±0.425 D). In group B, ALs measured by immersion B-scan (27.97±2.21 mm) didn't differ significantly from those of the IOLMaster (27.86±2.18 mm), but longer than those measured by Contact A-scan (27.75±2.23 mm, P=0.009). In the same group, the standard deviation (SD) of the mean error (ME) of immersion B-scan (-0.635±0.157 D) didn't differ significantly from those of the IOLMaster (-0.679±0.359 D), but differed significantly from those of contact A-scan (-0.953±1.713 D, P=0.028).
CONCLUSION: Immersion B-scan exhibits measurement accuracy comparable to that of the IOLMaster, and is thus a good alternative in measuring AL in eyes with high myopia when the IOLMaster can't be used, and it is more accurate than the contact A-scan.

Entities:  

Keywords:  axial length; high myopia; immersion; measurement; ultrasonography

Year:  2014        PMID: 24967188      PMCID: PMC4067656          DOI: 10.3980/j.issn.2222-3959.2014.03.10

Source DB:  PubMed          Journal:  Int J Ophthalmol        ISSN: 2222-3959            Impact factor:   1.779


  22 in total

1.  Evaluation of the practicality of optical biometry and applanation ultrasound in 253 eyes.

Authors:  Mana Tehrani; Frank Krummenauer; Eric Blom; H Burkhard Dick
Journal:  J Cataract Refract Surg       Date:  2003-04       Impact factor: 3.351

2.  Contact versus immersion biometry of axial length before cataract surgery.

Authors:  Michael P Hennessy; Derek G Chan
Journal:  J Cataract Refract Surg       Date:  2003-11       Impact factor: 3.351

3.  Biometry accuracy using zero- and negative-powered intraocular lenses.

Authors:  Robert E MacLaren; Mandeep S Sagoo; Marie Restori; Bruce D S Allan
Journal:  J Cataract Refract Surg       Date:  2005-02       Impact factor: 3.351

Review 4.  Calculation of intraocular lens power: a review.

Authors:  Thomas Olsen
Journal:  Acta Ophthalmol Scand       Date:  2007-04-02

5.  Comparison of ray-tracing method and thin-lens formula in intraocular lens power calculations.

Authors:  Haiying Jin; Tanja Rabsilber; Angela Ehmer; Andreas F Borkenstein; Il-Joo Limberger; Haike Guo; Gerd U Auffarth
Journal:  J Cataract Refract Surg       Date:  2009-04       Impact factor: 3.351

6.  B-mode-guided vector-A-mode versus A-mode biometry to determine axial length and intraocular lens power.

Authors:  O Bergès; M Puech; M Assouline; L Letenneur; M Gastellu-Etchegorry
Journal:  J Cataract Refract Surg       Date:  1998-04       Impact factor: 3.351

7.  Development of the SRK/T intraocular lens implant power calculation formula.

Authors:  J A Retzlaff; D R Sanders; M C Kraff
Journal:  J Cataract Refract Surg       Date:  1990-05       Impact factor: 3.351

8.  The accuracy of IOL power calculation formulas for eyes of axial length exceeding 24.5 mm.

Authors:  Wiktor Stopyra
Journal:  Klin Oczna       Date:  2013

9.  Intraocular lens power calculation by measuring axial length with partial optical coherence and ultrasonic biometry.

Authors:  Beatriz Machado Fontes; Bruno Machado Fontes; Elaine Castro
Journal:  Arq Bras Oftalmol       Date:  2011 May-Jun       Impact factor: 0.872

10.  Prediction error after pediatric cataract surgery with intraocular lens implantation: Contact versus immersion A-scan biometry.

Authors:  Rupal H Trivedi; M Edward Wilson
Journal:  J Cataract Refract Surg       Date:  2011-03       Impact factor: 3.351

View more
  6 in total

1.  Comparison of postoperative refractive outcome in eyes undergoing combined phacovitrectomy vs cataract surgery following vitrectomy.

Authors:  Paris G Tranos; Bruce Allan; Miltiadis Balidis; Athanasios Vakalis; Solon Asteriades; George Anogeianakis; Magda Triantafilla; Nikolaos Kozeis; Panagiotis Stavrakas
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2020-01-10       Impact factor: 3.117

2.  Accuracy of segmented measurement of axial length in ultra-high myopia filled with silicone oil using immersion B-scan ultrasonography.

Authors:  Qing-Hua Yang; Hong-Tao Zhang; Xiao-Qi Li; Bing Chen; Zhao-Hui Li; Yi-Fei Huang; Xin Jin; Ying Zhang; Li-Qiang Wang
Journal:  Int J Ophthalmol       Date:  2022-05-18       Impact factor: 1.645

3.  Predicting Axial Length From Choroidal Thickness on Optical Coherence Tomography Images With Machine Learning Based Algorithms.

Authors:  Hao-Chun Lu; Hsin-Yi Chen; Chien-Jung Huang; Pao-Hsien Chu; Lung-Sheng Wu; Chia-Ying Tsai
Journal:  Front Med (Lausanne)       Date:  2022-06-28

4.  Comparison axial length measurements from three biometric instruments in high myopia.

Authors:  Xiao-Gang Wang; Jing Dong; Yu-Lan Pu; Hui-Jun Liu; Qiang Wu
Journal:  Int J Ophthalmol       Date:  2016-06-18       Impact factor: 1.779

5.  Biometry characteristics in congenital cataract patients before surgery in a tertiary eye care centre in Nepal.

Authors:  Sabin Sahu; Paras Panjiyar
Journal:  Saudi J Ophthalmol       Date:  2019-07-17

Review 6.  Asymmetric diabetic retinopathy.

Authors:  Rajvardhan Azad; Sony Sinha; Prateek Nishant
Journal:  Indian J Ophthalmol       Date:  2021-11       Impact factor: 1.848

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.