Radiation exposure benefit of a lead cap in invasive cardiology.

E Kuon1, J Birkel, M Schmitt, J B Dahm.   

Abstract

BACKGROUND: Occupational head exposure to radiation in cardiologists may cause radiation induced cataracts and an increased risk of brain cancer.
OBJECTIVE: To determine the effectiveness of 0.5 mm lead equivalent caps, not previously used in invasive cardiology, in comparison with a 1.0 mm lead equivalent ceiling mounted lead glass screen.
DESIGN: An anthropomorphic Alderson-Rando phantom was used to represent the patient. Scatter entrance skin air kerma to the operator position (S-ESAK-O) was measured during fluoroscopy for all standard angulations and the S-ESAK-O per dose-area product (DAP) calculated, as applied to the phantom.
RESULTS: Measured mean (SD) left/right anterior oblique angulation ratios of S-ESAK-O without lead devices were 23.1 (10.1), and varied as a function of tube angulation, body height, and angle of incidence. S-ESAK-O/DAP decreased with incremental operator body height by 10 (3)% per 10 cm. A 1.0 mm lead glass shield reduced mean S-ESAK-O/DAP originating from coronary angiography from 1089 (764) to 54 (29) nSv/Gy x cm2. A 0.5 mm lead cap was effective in lowering measured levels to 1.8 (1.1) nSv/Gy x cm2. Both devices together enabled attenuation to 0.5 (0.1) nSv/Gy x cm2. The most advantageous line of vision for protection of the operator's eyes was > or = 60 degrees rightward.
CONCLUSIONS: Use of 0.5 mm lead caps proved highly effective, attenuating S-ESAK-O to 2.7 (2.0) x 10(-3) of baseline, and to 1.2 (1.4) x 10(-3) of baseline where there was an additional 1.0 mm lead glass shield. These results could vary according to the x ray systems used, catheterisation protocols, and correct use of radiation protection devices.

Entities:  

Mesh:

Substances:

Year:  2003        PMID: 12975420      PMCID: PMC1767881          DOI: 10.1136/heart.89.10.1205

Source DB:  PubMed          Journal:  Heart        ISSN: 1355-6037            Impact factor:   5.994


  23 in total

1.  Radiation exposure to medical staff in interventional and cardiac radiology.

Authors:  E Vañó; L González; E Guibelalde; J M Fernández; J I Ten
Journal:  Br J Radiol       Date:  1998-09       Impact factor: 3.039

2.  Significant reduction of radiation exposure to operator and staff during cardiac interventions by analysis of radiation leakage and improved lead shielding.

Authors:  Eberhard Kuon; Moritz Schmitt; Johannes B Dahm
Journal:  Am J Cardiol       Date:  2002-01-01       Impact factor: 2.778

3.  Clinical and technical determinants of the complexity of percutaneous transluminal coronary angioplasty procedures: analysis in relation to radiation exposure parameters.

Authors:  G Bernardi; R Padovani; G Morocutti; E Vaño; M R Malisan; M Rinuncini; L Spedicato; P M Fioretti
Journal:  Catheter Cardiovasc Interv       Date:  2000-09       Impact factor: 2.692

4.  Patient radiation exposure during coronary angiography and intervention.

Authors:  S G Fransson; J Persliden
Journal:  Acta Radiol       Date:  2000-03       Impact factor: 1.990

5.  Real-time quantification and display of skin radiation during coronary angiography and intervention.

Authors:  A den Boer; P J de Feijter; P W Serruys; J R Roelandt
Journal:  Circulation       Date:  2001-10-09       Impact factor: 29.690

6.  Radiation exposure to the operator performing cardiac angiography with U-arm systems.

Authors:  S Balter; F M Sones; R Brancato
Journal:  Circulation       Date:  1978-11       Impact factor: 29.690

Review 7.  Avoidance of radiation injuries from medical interventional procedures.

Authors:  J Valentin
Journal:  Ann ICRP       Date:  2000

8.  Effective dose to a patient undergoing coronary angiography.

Authors:  H Lobotessi; A Karoussou; V Neofotistou; A Louisi; V Tsapaki
Journal:  Radiat Prot Dosimetry       Date:  2001       Impact factor: 0.972

9.  Correlation of patient skin doses in cardiac interventional radiology with dose-area product.

Authors:  S van de Putte; F Verhaegen; Y Taeymans; H Thierens
Journal:  Br J Radiol       Date:  2000-05       Impact factor: 3.039

10.  Factors affecting patient radiation exposure during routine coronary angiography in a tertiary referral centre.

Authors:  A L Clark; A G Brennan; L J Robertson; J D McArthur
Journal:  Br J Radiol       Date:  2000-02       Impact factor: 3.039

View more
  13 in total

Review 1.  Occupational radiation doses to operators performing fluoroscopically-guided procedures.

Authors:  Kwang Pyo Kim; Donald L Miller; Amy Berrington de Gonzalez; Stephen Balter; Ruth A Kleinerman; Evgenia Ostroumova; Steven L Simon; Martha S Linet
Journal:  Health Phys       Date:  2012-07       Impact factor: 1.316

2.  Radiation exposure to cardiologists: how it could be reduced.

Authors:  E Vano
Journal:  Heart       Date:  2003-10       Impact factor: 5.994

3.  Radiation-reducing planning of cardiac catheterisation.

Authors:  E Kuon; J B Dahm; D M Robinson; K Empen; M Günther; W Wucherer
Journal:  Z Kardiol       Date:  2005-10

Review 4.  Cancer and non-cancer brain and eye effects of chronic low-dose ionizing radiation exposure.

Authors:  Eugenio Picano; Eliseo Vano; Luciano Domenici; Matteo Bottai; Isabelle Thierry-Chef
Journal:  BMC Cancer       Date:  2012-04-27       Impact factor: 4.430

Review 5.  The biological effects of diagnostic cardiac imaging on chronically exposed physicians: the importance of being non-ionizing.

Authors:  Maria Grazia Andreassi
Journal:  Cardiovasc Ultrasound       Date:  2004-11-22       Impact factor: 2.062

6.  Beam projections and radiation exposure in transradial and transfemoral approaches during coronary angiography.

Authors:  Ali Tarighatnia; Amirhossien Mohammadalian; Morteza Ghojazade; Leili Pourafkari; Alireza Farajollahi
Journal:  Anatol J Cardiol       Date:  2017-08-02       Impact factor: 1.596

7.  A Study on the Use of Radiation-Protective Apron among Interventionists in Radiology.

Authors:  Roshan Samuel Livingstone; Anna Varghese; Shyamkumar N Keshava
Journal:  J Clin Imaging Sci       Date:  2018-08-24

8.  Patient radiation doses in interventional cardiology procedures.

Authors:  Ioannis Pantos; Georgios Patatoukas; Demosthenes G Katritsis; Efstathios Efstathopoulos
Journal:  Curr Cardiol Rev       Date:  2009-01

9.  Fluoroscopic Cranial Radiation Exposure in Spine Surgery: A Prospective Single-Center Evaluation in Operating Room Personnel.

Authors:  Laila Malani Mohammad; James Messegee; M Omar Chohan; Christopher L Taylor
Journal:  Int J Spine Surg       Date:  2019-02-22

Review 10.  Radiation protection in the cardiac catheterization laboratory.

Authors:  Sylvia Marie R Biso; Mladen I Vidovich
Journal:  J Thorac Dis       Date:  2020-04       Impact factor: 3.005

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.