Masimo signal extraction pulse oximetry.

J M Goldman1, M T Petterson, R J Kopotic, S J Barker.   

Abstract

OBJECTIVE: To describe a new pulse oximetry technology and measurement paradigm developed by Masimo Corporation.
INTRODUCTION: Patient motion, poor tissue perfusion, excessive ambient light, and electrosurgical unit interference reduce conventional pulse oximeter (CPO) measurement integrity. Patient motion frequently generates erroneous pulse oximetry values for saturation and pulse rate. Motion-induced measurement error is due in part to widespread implementation of a theoretical pulse oximetry model which assumes that arterial blood is the only light-absorbing pulsatile component in the optical path.
METHODS: Masimo Signal Extraction Technology (SET) pulse oximetry begins with conventional red and infrared photoplethysmographic signals, and then employs a constellation of advanced techniques including radiofrequency and light-shielded optical sensors, digital signal processing, and adaptive filtration, to measure SpO2 accurately during challenging clinical conditions. In contrast to CPO which calculates O2 saturation from the ratio of transmitted pulsatile red and infrared light, Masimo SET pulse oximetry uses a new conceptual model of light absorption for pulse oximetry and employs the discrete saturation transform (DST) to isolate individual "saturation components" in the optical pathway. Typically, when the tissue under analysis is stationary, only the single saturation component produced by pulsatile arterial blood is present. In contrast, during patient motion, movement of non-arterial components (for example, venous blood) can be identified as additional saturation components (with a lower O2 saturation). When conditions of the Masimo model are met, the saturation component corresponding to the highest O2 saturation is reported by the instrument as SpO2.
CONCLUSION: The technological strategies implemented in Masimo SET pulse oximetry effectively permit continuous monitoring of SpO2 during challenging clinical conditions of motion and poor tissue perfusion.

Entities:  

Mesh:

Year:  2000        PMID: 12580205     DOI: 10.1023/a:1011493521730

Source DB:  PubMed          Journal:  J Clin Monit Comput        ISSN: 1387-1307            Impact factor:   2.502


  23 in total

1.  Pulse oximeters' reliability in detecting hypoxemia and bradycardia: comparison between a conventional and two new generation oximeters.

Authors:  B Bohnhorst; C S Peter; C F Poets
Journal:  Crit Care Med       Date:  2000-05       Impact factor: 7.598

2.  An update on pulse oximetry. Part II: limitations and future applications.

Authors:  K A Carlson; J S Jahr
Journal:  Anesthesiol Rev       Date:  1994 Mar-Apr

3.  Clinical evaluation of a prototype motion artifact resistant pulse oximeter in the recovery room.

Authors:  C Dumas; J A Wahr; K K Tremper
Journal:  Anesth Analg       Date:  1996-08       Impact factor: 5.108

4.  Major reduction in alarm frequency with a new pulse oximeter.

Authors:  B Bohnhorst; C F Poets
Journal:  Intensive Care Med       Date:  1998-03       Impact factor: 17.440

Review 5.  Pulse oximetry: analysis of theory, technology, and practice.

Authors:  M W Wukitsch; M T Petterson; D R Tobler; J A Pologe
Journal:  J Clin Monit       Date:  1988-10

6.  History of blood gas analysis. VII. Pulse oximetry.

Authors:  J W Severinghaus; Y Honda
Journal:  J Clin Monit       Date:  1987-04

7.  Crying wolf: false alarms in a pediatric intensive care unit.

Authors:  S T Lawless
Journal:  Crit Care Med       Date:  1994-06       Impact factor: 7.598

8.  Pulse oximetry monitoring can change routine oxygen supplementation practices in the postanesthesia care unit.

Authors:  R J DiBenedetto; S A Graves; N Gravenstein; C Konicek
Journal:  Anesth Analg       Date:  1994-02       Impact factor: 5.108

9.  A prospective study of intraoperative pulse oximetry failure.

Authors:  P R Freund; P T Overand; J Cooper; L Jacobson; S Bosse; B Walker; K L Posner; F W Cheney
Journal:  J Clin Monit       Date:  1991-07

10.  Postanesthesia monitoring revisited: frequency of true and false alarms from different monitoring devices.

Authors:  L Wiklund; B Hök; K Ståhl; A Jordeby-Jönsson
Journal:  J Clin Anesth       Date:  1994 May-Jun       Impact factor: 9.452

View more
  36 in total

1.  Decreased accuracy of pulse oximetry measurements during low perfusion caused by sepsis: Is the perfusion index of any value?

Authors:  Helmut D Hummler; Anja Engelmann; Frank Pohlandt; Josef Högel; Axel R Franz
Journal:  Intensive Care Med       Date:  2006-06-30       Impact factor: 17.440

2.  Pulse oximeter in a comparative test.

Authors:  Hartmut Gehring
Journal:  Intensive Care Med       Date:  2006-06-30       Impact factor: 17.440

3.  Adaptive motion artefact reduction in respiration and ECG signals for wearable healthcare monitoring systems.

Authors:  Zhengbo Zhang; Ikaro Silva; Dalei Wu; Jiewen Zheng; Hao Wu; Weidong Wang
Journal:  Med Biol Eng Comput       Date:  2014-10-02       Impact factor: 2.602

4.  Hypoxic events and concomitant factors in preterm infants on non-invasive ventilation.

Authors:  Omid Sadeghi Fathabadi; Timothy Gale; Kevin Wheeler; Gemma Plottier; Louise S Owen; J C Olivier; Peter A Dargaville
Journal:  J Clin Monit Comput       Date:  2016-02-20       Impact factor: 2.502

5.  Continuous noninvasive hemoglobin monitoring: the standard of care and future impact.

Authors:  Gerald J Kost; Nam K Tran
Journal:  Crit Care Med       Date:  2011-10       Impact factor: 7.598

6.  Algorithms for Monitoring Heart Rate and Respiratory Rate From the Video of a User's Face.

Authors:  Shourjya Sanyal; Koushik Kumar Nundy
Journal:  IEEE J Transl Eng Health Med       Date:  2018-04-12       Impact factor: 3.316

7.  Usefulness of perfusion index to detect the effect of brachial plexus block.

Authors:  Alparslan Kus; Yavuz Gurkan; Suna Kara Gormus; Mine Solak; Kamil Toker
Journal:  J Clin Monit Comput       Date:  2013-02-10       Impact factor: 2.502

8.  Motion resistant pulse oximetry in neonates.

Authors:  R Sahni; A Gupta; K Ohira-Kist; T S Rosen
Journal:  Arch Dis Child Fetal Neonatal Ed       Date:  2003-11       Impact factor: 5.747

9.  Comparison of two new generation pulse oximeters with arterial oxygen saturation in critically ill children.

Authors:  Bipin Jose; Rakesh Lodha; S K Kabra
Journal:  Indian J Pediatr       Date:  2014-03-15       Impact factor: 1.967

10.  Evaluation of biases in remote photoplethysmography methods.

Authors:  Ananyananda Dasari; Sakthi Kumar Arul Prakash; László A Jeni; Conrad S Tucker
Journal:  NPJ Digit Med       Date:  2021-06-03
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.