Literature DB >> 14602699

Motion resistant pulse oximetry in neonates.

R Sahni1, A Gupta, K Ohira-Kist, T S Rosen.   

Abstract

BACKGROUND: Pulse oximetry is widely used in neonates. However, its reliability is often affected by motion artefact. Clinicians confronted with questionable oxygen saturation (SpO(2)) values often estimate the reliability by correlating heart rate (HR) obtained with the oximeter with that obtained by electrocardiogram.
OBJECTIVE: To compare the effects of motion on SpO(2) and HR measurements made with Masimo signal extraction technology and those made with a Nellcor N-200.
DESIGN: Continuous pulse oximetry and HR monitoring were performed in 15 healthy, term infants (mean (SD) birth weight 3408 (458) g) undergoing circumcision, using Masimo and Nellcor pulse oximeters and a standard HR monitor (Hewlett-Packard). Simultaneous minute by minute behavioural activity codes were also assigned. Baseline data were collected for 10 minutes when the infant was quietly asleep and then continued during and after circumcision for a total duration of one hour. The oximeter HR and SpO(2) values were compared and related to HR values obtained by ECG during all three periods. The effect of behavioural activity on SpO(2) and HR was also evaluated.
RESULTS: When compared with results obtained with the Nellcor, the mean SpO(2) and HR were higher and the incidence of artefact lower with the Masimo during all three periods. Masimo HR more accurately predicted HR obtained with a standard monitor, with lower residual error. SpO(2) and HR values obtained with the Nellcor were lower and more variable during all behavioural states, especially crying, when excessive motion artefact was most likely.
CONCLUSIONS: The data suggest that Masimo signal extraction technology may offer improvement in pulse oximetry performance, particularly in clinical situations in which extreme motion artefacts are likely.

Entities:  

Mesh:

Substances:

Year:  2003        PMID: 14602699      PMCID: PMC1763244          DOI: 10.1136/fn.88.6.f505

Source DB:  PubMed          Journal:  Arch Dis Child Fetal Neonatal Ed        ISSN: 1359-2998            Impact factor:   5.747


  15 in total

1.  Clinical evaluation of a prototype motion artifact resistant pulse oximeter in the recovery room.

Authors:  C Dumas; J A Wahr; K K Tremper
Journal:  Anesth Analg       Date:  1996-08       Impact factor: 5.108

Review 2.  Physiology of oxygenation and its relation to pulse oximetry in neonates.

Authors:  W W Hay
Journal:  J Perinatol       Date:  1987       Impact factor: 2.521

3.  Neonatal pulse oximetry: accuracy and reliability.

Authors:  W W Hay; J M Brockway; M Eyzaguirre
Journal:  Pediatrics       Date:  1989-05       Impact factor: 7.124

4.  Routine oximetry. A fifth vital sign?

Authors:  T A Neff
Journal:  Chest       Date:  1988-08       Impact factor: 9.410

5.  Knowledge about pulse oximetry among medical and nursing staff.

Authors:  M D Stoneham; G M Saville; I H Wilson
Journal:  Lancet       Date:  1994-11-12       Impact factor: 79.321

6.  Masimo signal extraction pulse oximetry.

Authors:  J M Goldman; M T Petterson; R J Kopotic; S J Barker
Journal:  J Clin Monit Comput       Date:  2000       Impact factor: 2.502

7.  A scoring system for states of sleep and wakefulness in term and preterm infants.

Authors:  M Stefanski; K Schulze; D Bateman; R Kairam; T A Pedley; J Masterson; L S James
Journal:  Pediatr Res       Date:  1984-01       Impact factor: 3.756

8.  Crying wolf: false alarms in a pediatric intensive care unit.

Authors:  S T Lawless
Journal:  Crit Care Med       Date:  1994-06       Impact factor: 7.598

Review 9.  Pulse oximetry in neonatal medicine.

Authors:  W W Hay; E Thilo; J B Curlander
Journal:  Clin Perinatol       Date:  1991-09       Impact factor: 3.430

10.  Responses of premature infants to routine nursing interventions and noise in the NICU.

Authors:  L K Zahr; S Balian
Journal:  Nurs Res       Date:  1995 May-Jun       Impact factor: 2.381

View more
  7 in total

1.  Performance of Two Different Pulse Oximeters in Neonatal Transition.

Authors:  Sunita Jeyapal; Peter Prasanth Kumar Kommu; M Manikandan; Lalitha Krishnan
Journal:  Indian J Pediatr       Date:  2016-08-19       Impact factor: 1.967

2.  A higher incidence of intermittent hypoxemic episodes is associated with severe retinopathy of prematurity.

Authors:  Juliann M Di Fiore; Jeffrey N Bloom; Faruk Orge; Alison Schutt; Mark Schluchter; Vinay K Cheruvu; Michele Walsh; Neil Finer; Richard J Martin
Journal:  J Pediatr       Date:  2010-03-20       Impact factor: 4.406

3.  Prevention of retinopathy of prematurity in preterm infants through changes in clinical practice and SpO₂technology.

Authors:  Armando Castillo; Richard Deulofeut; Ann Critz; Augusto Sola
Journal:  Acta Paediatr       Date:  2010-10-15       Impact factor: 2.299

4.  SVM-Based Spectral Analysis for Heart Rate from Multi-Channel WPPG Sensor Signals.

Authors:  Jiping Xiong; Lisang Cai; Fei Wang; Xiaowei He
Journal:  Sensors (Basel)       Date:  2017-03-03       Impact factor: 3.576

5.  Noninvasive Monitoring during Interhospital Transport of Newborn Infants.

Authors:  Georg M Schmölzer; Megan O'Reilly; Po-Yin Cheung
Journal:  Crit Care Res Pract       Date:  2013-02-24

Review 6.  Safe oxygen saturation targeting and monitoring in preterm infants: can we avoid hypoxia and hyperoxia?

Authors:  Augusto Sola; Sergio G Golombek; María Teresa Montes Bueno; Lourdes Lemus-Varela; Claudia Zuluaga; Fernando Domínguez; Hernando Baquero; Alejandro E Young Sarmiento; Diego Natta; Jose M Rodriguez Perez; Richard Deulofeut; Ana Quiroga; Gabriel Lara Flores; Mónica Morgues; Alfredo García-Alix Pérez; Bart Van Overmeire; Frank van Bel
Journal:  Acta Paediatr       Date:  2014-07-28       Impact factor: 2.299

7.  Epidemiology of patient monitoring alarms in the neonatal intensive care unit.

Authors:  Taibo Li; Minoru Matsushima; Wendy Timpson; Susan Young; David Miedema; Munish Gupta; Thomas Heldt
Journal:  J Perinatol       Date:  2018-05-08       Impact factor: 2.521

  7 in total