Comparison of two new generation pulse oximeters with arterial oxygen saturation in critically ill children.

Bipin Jose1, Rakesh Lodha, S K Kabra.   

Abstract

OBJECTIVES: To compare the performance of two new generation pulse oximeters, one with enhanced signal extraction technology (SET) and other without enhanced SET in detecting hypoxemia and to correlate it with arterial blood gas analysis.
METHODS: Forty-eight patients, admitted to pediatric intensive care unit (PICU) of a teritiary care teaching hospital in India for critical care and support during the study period, who had an arterial catheter in situ were included. Children with those disease conditions known to interfere with pulse oximetry and blood gas analysis were excluded.184 set of observations were made during the study period. Each set had oxygen saturation (SpO2) measured from both the pulse oximeters and the corresponding arterial oxygen saturation (SaO2). The values were compared for occurrence of true and false alarms during periods of normal BP, hypotension and varying degrees of hypoxia.
RESULTS: The mean arterial SaO2 in the study was 94.4 % ± 4.9. The mean SpO2 recorded in conventional and enhanced signal extraction technology (SET) pulse oximeters were 94.9 % ± 4.5 and 97.2 % ± 4.7 respectively. Enhanced signal extraction technology pulse oximeter detected 4/27 (15 %) of true hypoxemic events and 1 event was a false alarm. Conventional pulse oximeter detected 11/27 (41 %) true hypoxemic events but recorded 6 false alarms.
CONCLUSIONS: Both pulse oximeters were not found to be performing satisfactorily in picking up hypoxemia in the study. There was good correlation with mean SpO2 from pulse oximeters and arterial SaO2. The reliability of pulse oximetry decreases with worsening hypoxemia and hypotension, and the sensitivity for picking up hypoxemia can be as low as 15 %.

Entities:  

Mesh:

Substances:

Year:  2014        PMID: 24627281     DOI: 10.1007/s12098-014-1381-z

Source DB:  PubMed          Journal:  Indian J Pediatr        ISSN: 0019-5456            Impact factor:   1.967


  12 in total

1.  The effects of motion artifact and low perfusion on the performance of a new generation of pulse oximeters in volunteers undergoing hypoxemia.

Authors:  Hartmut Gehring; Christoph Hornberger; Holger Matz; Ewald Konecny; Peter Schmucker
Journal:  Respir Care       Date:  2002-01       Impact factor: 2.258

2.  Does ambient light affect the accuracy of pulse oximetry?

Authors:  Robert R Fluck; Christine Schroeder; Greg Frani; Brad Kropf; Brenda Engbretson
Journal:  Respir Care       Date:  2003-07       Impact factor: 2.258

3.  False alarms and sensitivity of conventional pulse oximetry versus the Masimo SET technology in the pediatric postanesthesia care unit.

Authors:  S Malviya; P I Reynolds; T Voepel-Lewis; M Siewert; D Watson; A R Tait; K Tremper
Journal:  Anesth Analg       Date:  2000-06       Impact factor: 5.108

4.  Effects of methemoglobinemia on pulse oximetry and mixed venous oximetry.

Authors:  S J Barker; K K Tremper; J Hyatt
Journal:  Anesthesiology       Date:  1989-01       Impact factor: 7.892

5.  Limitations of pulse oximetry.

Authors:  A Huch; R Huch; V König; M R Neuman; D Parker; J Yount; D Lübbers
Journal:  Lancet       Date:  1988-02-13       Impact factor: 79.321

6.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

7.  Differences in pulse oximetry technology can affect detection of sleep-disorderd breathing in children.

Authors:  Robert T Brouillette; Jacinthe Lavergne; Andra Leimanis; Gillian M Nixon; Sylvia Ladan; Christine D McGregor
Journal:  Anesth Analg       Date:  2002-01       Impact factor: 5.108

8.  The pulse oximetry gap in carbon monoxide intoxication.

Authors:  R G Buckley; S E Aks; J L Eshom; R Rydman; J Schaider; P Shayne
Journal:  Ann Emerg Med       Date:  1994-08       Impact factor: 5.721

9.  Report of the Second Task Force on Blood Pressure Control in Children--1987. Task Force on Blood Pressure Control in Children. National Heart, Lung, and Blood Institute, Bethesda, Maryland.

Authors: 
Journal:  Pediatrics       Date:  1987-01       Impact factor: 7.124

10.  Oxygen saturation and outcomes in preterm infants.

Authors:  Ben J Stenson; William O Tarnow-Mordi; Brian A Darlow; John Simes; Edmund Juszczak; Lisa Askie; Malcolm Battin; Ursula Bowler; Roland Broadbent; Pamela Cairns; Peter Graham Davis; Sanjeev Deshpande; Mark Donoghoe; Lex Doyle; Brian W Fleck; Alpana Ghadge; Wendy Hague; Henry L Halliday; Michael Hewson; Andrew King; Adrienne Kirby; Neil Marlow; Michael Meyer; Colin Morley; Karen Simmer; Win Tin; Stephen P Wardle; Peter Brocklehurst
Journal:  N Engl J Med       Date:  2013-05-05       Impact factor: 91.245

View more
  2 in total

1.  Comparison of Two New Generation Pulse Oximeters with Arterial Oxygen Saturation in Critically Ill Children: Authors' Reply.

Authors:  Bipin Jose; Rakesh Lodha; Sushil K Kabra
Journal:  Indian J Pediatr       Date:  2015-05-14       Impact factor: 1.967

2.  Comparison of Two New Generation Pulse Oximeters with Arterial Oxygen Saturation in Critically Ill Children: Correspondence.

Authors:  Viswas Chhapola; Sandeep Kumar Kanwal
Journal:  Indian J Pediatr       Date:  2015-05-14       Impact factor: 1.967

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.