Pulse oximeters' reliability in detecting hypoxemia and bradycardia: comparison between a conventional and two new generation oximeters.

B Bohnhorst1, C S Peter, C F Poets.   

Abstract

OBJECTIVE: Pulse oximeters are increasingly used for patient monitoring; however, they are traditionally very prone to motion artifact. Newly developed instruments have lower false alarm rates. We wanted to know whether this is achieved at the expense of an increased proportion of false negative alarms such as missed or delayed identification of hypoxemia and/or bradycardia.
DESIGN: Observational study.
SETTING: Neonatal intensive care unit. PATIENTS: A total of 17 unsedated preterm infants (median gestational age at birth, 25 wks; range, 24-30 wks). INTERVENTION: Long-term recordings of transcutaneous partial pressure of oxygen (P(Tc)O2), heart rate, pulse oximeter saturation (SpO2), and pulse rate from a conventional oximeter and two new generation oximeters. MEASUREMENTS: Recordings were analyzed for episodes with P(Tc)O2 <40 torr or with heart rate <80 beats/min for >5 secs. Hypoxemia was considered identified if SpO2 had fallen to <85% within 2 mins of P(Tc)O2 reaching 40 torr, and bradycardia was considered identified if pulse rate had fallen to <80 beats/min within 2 mins of the heart rate reaching this threshold. MAIN
RESULTS: A total of 202 falls in P(Tc)O2 to <40 torr occurred; 174 (86%) were identified by all three oximeters. Of the remaining episodes, manual analysis of red and infrared absorption signals confirmed that SpO2 had indeed been <85% for > or =10 secs in 11 episodes; therefore, these episodes should have been identified by all three oximeters. None of these had been missed by the conventional oximeter, but 10 (5.4% of the total) were missed by one of the new generation instruments (Nellcor), and one (0.5%) was missed by the other (Masimo). Of 54 bradycardias, only 14 were identified by all three oximeters; 17 (32%) were missed by the conventional, 37 (69%) by the Nellcor, and 4 (7%) by the Masimo instrument.
CONCLUSION: One of the two new generation instruments investigated in this study missed 5.4% of hypoxemic episodes and 69% of bradycardias. It thus appears that this instrument's reduced false alarm rate is achieved at the expense of an unreliable and/or delayed identification of hypoxemia and bradycardia. The other instrument identified both conditions equally as or more reliably than a conventional pulse oximeter.

Entities:  

Mesh:

Year:  2000        PMID: 10834712     DOI: 10.1097/00003246-200005000-00050

Source DB:  PubMed          Journal:  Crit Care Med        ISSN: 0090-3493            Impact factor:   7.598


  15 in total

1.  Closed system endotracheal suctioning maintains lung volume during volume controlled mechanical ventilation.

Authors:  David Pogson; Peter Shirley; Elizabeth Connolly; Sharon Johnston
Journal:  Intensive Care Med       Date:  2001-11-29       Impact factor: 17.440

2.  Decreased accuracy of pulse oximetry measurements during low perfusion caused by sepsis: Is the perfusion index of any value?

Authors:  Helmut D Hummler; Anja Engelmann; Frank Pohlandt; Josef Högel; Axel R Franz
Journal:  Intensive Care Med       Date:  2006-06-30       Impact factor: 17.440

3.  Pulse oximeter in a comparative test.

Authors:  Hartmut Gehring
Journal:  Intensive Care Med       Date:  2006-06-30       Impact factor: 17.440

Review 4.  Pulse oximetry for monitoring infants in the delivery room: a review.

Authors:  J A Dawson; P G Davis; C P F O'Donnell; C O F Kamlin; C J Morley
Journal:  Arch Dis Child Fetal Neonatal Ed       Date:  2007-01       Impact factor: 5.747

5.  Online apnea-bradycardia detection based on hidden semi-Markov models.

Authors:  Miguel Altuve; Guy Carrault; Alain Beuchée; Patrick Pladys; Alfredo I Hernández
Journal:  Med Biol Eng Comput       Date:  2014-10-10       Impact factor: 2.602

6.  The incidence of hypoxemia during surgery: evidence from two institutions.

Authors:  Jesse M Ehrenfeld; Luke M Funk; Johan Van Schalkwyk; Alan F Merry; Warren S Sandberg; Atul Gawande
Journal:  Can J Anaesth       Date:  2010-07-31       Impact factor: 5.063

7.  Volume-controlled intermittent mandatory ventilation in preterm infants with hypoxemic episodes.

Authors:  Helmut D Hummler; Anja Engelmann; Frank Pohlandt; Axel R Franz
Journal:  Intensive Care Med       Date:  2006-02-24       Impact factor: 17.440

8.  Masimo signal extraction pulse oximetry.

Authors:  J M Goldman; M T Petterson; R J Kopotic; S J Barker
Journal:  J Clin Monit Comput       Date:  2000       Impact factor: 2.502

9.  Accuracy of pulse oximetry readings in an animal model of low perfusion caused by emerging pneumonia and sepsis.

Authors:  Helmut D Hummler; Anja Engelmann; Frank Pohlandt; Josef Högel; Axel R Franz
Journal:  Intensive Care Med       Date:  2004-01-13       Impact factor: 17.440

10.  Performance of a digital PCO2/SPO2 ear sensor.

Authors:  Serge Kocher; Roman Rohling; Andres Tschupp
Journal:  J Clin Monit Comput       Date:  2004-04       Impact factor: 2.502

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.