Literature DB >> 16810522

Decreased accuracy of pulse oximetry measurements during low perfusion caused by sepsis: Is the perfusion index of any value?

Helmut D Hummler1, Anja Engelmann, Frank Pohlandt, Josef Högel, Axel R Franz.   

Abstract

OBJECTIVE: To evaluate the effects of deteriorating perfusion caused by sepsis on the accuracy of pulse oximetry measurements using two more recently available techniques (Nellcor N-395 and Masimo Radical) and to evaluate the perfusion index as a marker of impaired peripheral perfusion to indicate that accuracy of pulse oximetry readings may be affected. DESIGN AND
SETTING: Interventional cohort study in a university animal research facility.
SUBJECTS: Thirty-seven adult anesthetized, ventilated rabbits.
INTERVENTIONS: Pneumonia/sepsis was induced by tracheal instillation of Escherichia coli. MEASUREMENTS AND
RESULTS: Oxygen saturation and perfusion index as a marker of peripheral perfusion were measured by pulse oximetry (SpO2) and recorded continuously for 8 h. Arterial oxygen saturation (SaO2) was measured every 30 min by CO oximetry, and bias (SpO2 - SaO2) was calculated at each time point for each device to assess time-dependent changes in bias. Bias increased significantly across time for both devices tested comparing the first with the second half of the experimental period. Bias measurements during the second half of the experimental time were beyond the +/-3% error limit in 21.4% of cases with Nellcor and in 22.6% with Masimo. A lower perfusion index was associated with increased bias, but sensitivity, specificity, and positive and negative predictive values of this marker for increased bias was very limited.
CONCLUSIONS: We conclude that accuracy of pulse oximetry measurements was considerably affected with both devices with progressively deteriorating hemodynamics in this animal model of severe sepsis. Perfusion index as a marker for increased risk of bias was not a useful tool.

Entities:  

Mesh:

Year:  2006        PMID: 16810522     DOI: 10.1007/s00134-006-0254-y

Source DB:  PubMed          Journal:  Intensive Care Med        ISSN: 0342-4642            Impact factor:   17.440


  18 in total

1.  The effects of motion artifact and low perfusion on the performance of a new generation of pulse oximeters in volunteers undergoing hypoxemia.

Authors:  Hartmut Gehring; Christoph Hornberger; Holger Matz; Ewald Konecny; Peter Schmucker
Journal:  Respir Care       Date:  2002-01       Impact factor: 2.258

2.  Pulse oximeters' reliability in detecting hypoxemia and bradycardia: comparison between a conventional and two new generation oximeters.

Authors:  B Bohnhorst; C S Peter; C F Poets
Journal:  Crit Care Med       Date:  2000-05       Impact factor: 7.598

3.  "Motion-resistant" pulse oximetry: a comparison of new and old models.

Authors:  Steven J Barker
Journal:  Anesth Analg       Date:  2002-10       Impact factor: 5.108

4.  Comparison of pulse oximeters: accuracy at low arterial pressure in volunteers.

Authors:  R J Falconer; B J Robinson
Journal:  Br J Anaesth       Date:  1990-10       Impact factor: 9.166

Review 5.  Recent developments in pulse oximetry.

Authors:  J W Severinghaus; J F Kelleher
Journal:  Anesthesiology       Date:  1992-06       Impact factor: 7.892

6.  Accuracy of pulse oximetry in patients with low systemic vascular resistance.

Authors:  C Secker; P Spiers
Journal:  Anaesthesia       Date:  1997-02       Impact factor: 6.955

7.  Accuracy of pulse oximetry in hypothermic neonates and infants undergoing cardiac surgery.

Authors:  P Iyer; P McDougall; P Loughnan; R B Mee; K Al-Tawil; J Carlin
Journal:  Crit Care Med       Date:  1996-03       Impact factor: 7.598

8.  Transesophageal versus surface pulse oximetry in intensive care unit patients.

Authors:  M N Vicenzi; H Gombotz; H Krenn; C Dorn; P Rehak
Journal:  Crit Care Med       Date:  2000-07       Impact factor: 7.598

9.  Accuracy of pulse oximetry readings in an animal model of low perfusion caused by emerging pneumonia and sepsis.

Authors:  Helmut D Hummler; Anja Engelmann; Frank Pohlandt; Josef Högel; Axel R Franz
Journal:  Intensive Care Med       Date:  2004-01-13       Impact factor: 17.440

10.  Reliability of conventional and new pulse oximetry in neonatal patients.

Authors:  William W Hay; Donna J Rodden; Shannon M Collins; Diane L Melara; Kathy A Hale; Lucy M Fashaw
Journal:  J Perinatol       Date:  2002 Jul-Aug       Impact factor: 2.521

View more
  10 in total

1.  Pulse oximeter in a comparative test.

Authors:  Hartmut Gehring
Journal:  Intensive Care Med       Date:  2006-06-30       Impact factor: 17.440

Review 2.  Year in review in Intensive Care Medicine, 2006. I. Experimental studies. Clinical studies: brain injury, renal failure and endocrinology.

Authors:  Peter Andrews; Elie Azoulay; Massimo Antonelli; Laurent Brochard; Christian Brun-Buisson; Daniel De Backer; Geoffrey Dobb; Jean-Yves Fagon; Herwig Gerlach; Johan Groeneveld; Duncan Macrae; Jordi Mancebo; Philipp Metnitz; Stefano Nava; Jerôme Pugin; Michael Pinsky; Peter Radermacher; Christian Richard
Journal:  Intensive Care Med       Date:  2006-12-19       Impact factor: 17.440

3.  A pilot study of the pleth variability index as an indicator of volume-responsive hypotension in newborn infants during surgery.

Authors:  Soyhan Bagci; Nicole Müller; Andreas Müller; Andreas Heydweiller; Peter Bartmann; Axel R Franz
Journal:  J Anesth       Date:  2012-11-07       Impact factor: 2.078

4.  Evaluating Innovative In-Ear Pulse Oximetry for Unobtrusive Cardiovascular and Pulmonary Monitoring During Sleep.

Authors:  Boudewijn Venema; Johannes Schiefer; Vladimir Blazek; Nikolai Blanik; Steffen Leonhardt
Journal:  IEEE J Transl Eng Health Med       Date:  2013-08-08       Impact factor: 3.316

5.  Recent advances in pulse oximetry.

Authors:  Maxime Cannesson; Pekka Talke
Journal:  F1000 Med Rep       Date:  2009-08-26

6.  Perfusion index versus non-invasive hemodynamic parameters during insertion of i-gel, classic laryngeal mask airway and endotracheal tube.

Authors:  Hosam M Atef; Salah Abd Fattah; Mohammed Emad Abd Gaffer; Ahamed Abd Al Rahman
Journal:  Indian J Anaesth       Date:  2013-03

7.  Utility of digital pulse oximetry in the screening of lower extremity arterial disease.

Authors:  Jung-Nam Kwon; Whan-Bong Lee
Journal:  J Korean Surg Soc       Date:  2012-01-27

8.  Perfusion index assessment during transition period of newborns: an observational study.

Authors:  Sezin Unal; Ebru Ergenekon; Selma Aktas; Serdar Beken; Nilgun Altuntas; Ebru Kazanci; Ferit Kulali; Ibrahim M Hirfanoglu; Esra Onal; Canan Turkyilmaz; Esin Koc; Yildiz Atalay
Journal:  BMC Pediatr       Date:  2016-10-07       Impact factor: 2.125

9.  A Retrospective Analysis of Oxygen Desaturation during Acoustic Respiratory Rate Monitoring in Non-ICU Patients following Tracheal Extubation after General Anesthesia.

Authors:  Hideaki Kawanishi; Satoki Inoue; Masahiko Kawaguchi
Journal:  Anesthesiol Res Pract       Date:  2017-04-12

Review 10.  Photoplethysmogram Analysis and Applications: An Integrative Review.

Authors:  Junyung Park; Hyeon Seok Seok; Sang-Su Kim; Hangsik Shin
Journal:  Front Physiol       Date:  2022-03-01       Impact factor: 4.566

  10 in total