Clinical evaluation of a prototype motion artifact resistant pulse oximeter in the recovery room.

C Dumas1, J A Wahr, K K Tremper.   

Abstract

The frequency and nature of spurious pulse oximetry readings were compared using both a conventional pulse oximeter (CPO) and a prototype Masimo signal extraction technology pulse oximeter (Masimo SET). At a university hospital, 50 ASA physical status I-IV adult patients who underwent general or spinal-epidural anesthesia were selected from a group of 250 patients on the basis of high-alarm generation with routine postoperative pulse oximetry. Pulse oximetry data were recorded simultaneously from both devices with a computer. Overall, the CPO alarm frequency (i.e., oxygen saturation < 90%, or complete signal loss) was once every 13 min, and 87% of these alarms were considered false. Alarms were considered false based on reference electrocardiographs (16 patients), arterial blood gases (7 patients), and clinical assessment. The prototype Masimo SET device alarm frequency was once every 30 min, and 59% of these were considered false. During arm motion with 15 patients, the CPO device produced spurious signals on 54 occasions compared with five for the prototype Masimo SET. The incidence of artifactual pulse oximetry events during patient motion appear to be substantially reduced with the prototype Masimo SET device, relative to a CPO device.

Entities:  

Mesh:

Substances:

Year:  1996        PMID: 8694304     DOI: 10.1097/00000539-199608000-00012

Source DB:  PubMed          Journal:  Anesth Analg        ISSN: 0003-2999            Impact factor:   5.108


  11 in total

1.  Making ICU alarms meaningful: a comparison of traditional vs. trend-based algorithms.

Authors:  R Schoenberg; D Z Sands; C Safran
Journal:  Proc AMIA Symp       Date:  1999

2.  A second generation technique for evaluating accuracy and reliability of second generation pulse oximeters.

Authors:  K K Tremper
Journal:  J Clin Monit Comput       Date:  2000       Impact factor: 2.502

3.  Influence of pulse oximeter settings on the frequency of alarms and detection of hypoxemia: Theoretical effects of artifact rejection, alarm delay, averaging, median filtering or a lower setting of the alarm limit.

Authors:  A T Rheineck-Leyssius; C J Kalkman
Journal:  J Clin Monit Comput       Date:  1998-04       Impact factor: 2.502

4.  Major reduction in alarm frequency with a new pulse oximeter.

Authors:  B Bohnhorst; C F Poets
Journal:  Intensive Care Med       Date:  1998-03       Impact factor: 17.440

5.  Masimo signal extraction pulse oximetry.

Authors:  J M Goldman; M T Petterson; R J Kopotic; S J Barker
Journal:  J Clin Monit Comput       Date:  2000       Impact factor: 2.502

6.  Monitoring of respiratory rate in postoperative care using a new photoplethysmographic technique.

Authors:  L Nilsson; A Johansson; S Kalman
Journal:  J Clin Monit Comput       Date:  2000       Impact factor: 2.502

7.  Sleep Overnight Monitoring for Apnea in Patients Hospitalized with Heart Failure (SOMA-HF Study).

Authors:  Sunil Sharma; Paul J Mather; Anindita Chowdhury; Suchita Gupta; Umer Mukhtar; Leslee Willes; David J Whellan; Atul Malhotra; Stuart F Quan
Journal:  J Clin Sleep Med       Date:  2017-10-15       Impact factor: 4.062

8.  Motion resistant pulse oximetry in neonates.

Authors:  R Sahni; A Gupta; K Ohira-Kist; T S Rosen
Journal:  Arch Dis Child Fetal Neonatal Ed       Date:  2003-11       Impact factor: 5.747

9.  Evaluation of finger and forehead pulse oximeters during mild hypothermic cardiopulmonary bypass.

Authors:  Ken Yamaura; Kazuo Irita; Tadashi Kandabashi; Kohei Tohyama; Shosuke Takahashi
Journal:  J Clin Monit Comput       Date:  2007-06-20       Impact factor: 1.977

10.  Pulse oximetry.

Authors: 
Journal:  Crit Care       Date:  1999       Impact factor: 9.097

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.