Influence of pulse oximeter settings on the frequency of alarms and detection of hypoxemia: Theoretical effects of artifact rejection, alarm delay, averaging, median filtering or a lower setting of the alarm limit.

A T Rheineck-Leyssius1, C J Kalkman.   

Abstract

OBJECTIVE: The potential benefit of a reduced frequency of false pulse oximeter low oxyhemoglobin saturation (SpO2) alarms is that the attention of personnel is only directed to patients who experience hypoxemia. The present study was undertaken to better understand the effects of different settings of the pulse oximeter on false (artifact) and true (hypoxemia) alarms.
METHODS: Using the original SpO2 data of 200 postoperative patients, we calculated off-line the effects of five methods (artifact rejection, alarm delay (2-44 s, 2 s increments), averaging (10-90 s), median filtering (10-90 s) and decreasing the alarm limit from 90% to 85%) on the number of (true- and false) alarms.
RESULTS: 830 episodes of hypoxemia (SpO2 < or = 90%) and 73 episodes of severe hypoxemia (SpO2 < or = 85%) occurred. With a SpO2 alarm limit of 90%, the alarm was triggered 1535 times (830 true, 705 false). Artifact rejection reduced alarms by almost 50%. An alarm delay of 6 s or an averaging or median filtering epoch of 10 s resulted in an alarm reduction of almost 50%. No differences were found in the reduction of alarms between averaging and median filtering. Changing the alarm limit to 85% reduced the number of alarms by 82%. A similar reduction of alarms was obtained with either an alarm delay of 18 s or an averaging or median filtering epoch of 42 s. However, an alarm limit of 85% reduced the number of false alarms less than the other three algorithms (p < 0.01).
CONCLUSIONS: The data from the present study suggest that in order to effectively suppress false alarms caused by pulse oximeter artifacts, it may be preferable to use a longer filtering epoch of approximately 40 s, rather than to decrease the lower alarm limit.

Entities:  

Mesh:

Year:  1998        PMID: 9676861     DOI: 10.1023/a:1007431305610

Source DB:  PubMed          Journal:  J Clin Monit Comput        ISSN: 1387-1307            Impact factor:   2.502


  15 in total

1.  Clinical evaluation of a prototype motion artifact resistant pulse oximeter in the recovery room.

Authors:  C Dumas; J A Wahr; K K Tremper
Journal:  Anesth Analg       Date:  1996-08       Impact factor: 5.108

2.  Effects of motion on the performance of pulse oximeters in volunteers.

Authors:  S J Barker; N K Shah
Journal:  Anesthesiology       Date:  1996-10       Impact factor: 7.892

3.  Adverse respiratory events in anesthesia: a closed claims analysis.

Authors:  R A Caplan; K L Posner; R J Ward; F W Cheney
Journal:  Anesthesiology       Date:  1990-05       Impact factor: 7.892

4.  Are there too many alarms in the intensive care unit? An overview of the problems.

Authors:  C Meredith; J Edworthy
Journal:  J Adv Nurs       Date:  1995-01       Impact factor: 3.187

5.  The median filter as a preprocessor for a patient monitor limit alarm system in intensive care.

Authors:  A Mäkivirta; E Koski; A Kari; T Sukuvaara
Journal:  Comput Methods Programs Biomed       Date:  1991 Feb-Mar       Impact factor: 5.428

6.  Crying wolf: false alarms in a pediatric intensive care unit.

Authors:  S T Lawless
Journal:  Crit Care Med       Date:  1994-06       Impact factor: 7.598

7.  Randomized evaluation of pulse oximetry in 20,802 patients: II. Perioperative events and postoperative complications.

Authors:  J T Moller; N W Johannessen; K Espersen; O Ravlo; B D Pedersen; P F Jensen; N H Rasmussen; L S Rasmussen; T Pedersen; J B Cooper
Journal:  Anesthesiology       Date:  1993-03       Impact factor: 7.892

8.  Alarm-inducing variability in cardiac postoperative data and the effects of prealarm delay.

Authors:  A Mäkivirta; E M Koski
Journal:  J Clin Monit       Date:  1994-05

9.  Prevention of intraoperative anesthesia accidents and related severe injury through safety monitoring.

Authors:  J H Eichhorn
Journal:  Anesthesiology       Date:  1989-04       Impact factor: 7.892

10.  Postanesthesia monitoring revisited: frequency of true and false alarms from different monitoring devices.

Authors:  L Wiklund; B Hök; K Ståhl; A Jordeby-Jönsson
Journal:  J Clin Anesth       Date:  1994 May-Jun       Impact factor: 9.452

View more
  12 in total

1.  Artifact detection in cardiovascular time series monitoring data from preterm infants.

Authors:  C Cao; I S Kohane; N McIntosh
Journal:  Proc AMIA Symp       Date:  1999

2.  Reduction of false arterial blood pressure alarms using signal quality assessment and relationships between the electrocardiogram and arterial blood pressure.

Authors:  W Zong; G B Moody; R G Mark
Journal:  Med Biol Eng Comput       Date:  2004-09       Impact factor: 2.602

3.  The Impact of Averaging Window Length on the"Desaturation Indexes during Overnight Pulse Oximetry at High-Altitude".

Authors:  Troy J Cross; Manda Keller-Ross; Amine Issa; Robert Wentz; Bryan Taylor; Bruce Johnson
Journal:  Sleep       Date:  2015-08-01       Impact factor: 5.849

4.  Masimo signal extraction pulse oximetry.

Authors:  J M Goldman; M T Petterson; R J Kopotic; S J Barker
Journal:  J Clin Monit Comput       Date:  2000       Impact factor: 2.502

5.  Evaluation of a wireless, portable, wearable multi-parameter vital signs monitor in hospitalized neurological and neurosurgical patients.

Authors:  Robert S Weller; Kristina L Foard; Timothy N Harwood
Journal:  J Clin Monit Comput       Date:  2017-12-06       Impact factor: 2.502

6.  Artifact detection in the PO2 and PCO2 time series monitoring data from preterm infants.

Authors:  C Cao; N McIntosh; I S Kohane; K Wang
Journal:  J Clin Monit Comput       Date:  1999-08       Impact factor: 2.502

7.  Motion resistant pulse oximetry in neonates.

Authors:  R Sahni; A Gupta; K Ohira-Kist; T S Rosen
Journal:  Arch Dis Child Fetal Neonatal Ed       Date:  2003-11       Impact factor: 5.747

Review 8.  Alarms in the intensive care unit: how can the number of false alarms be reduced?

Authors:  M C Chambrin
Journal:  Crit Care       Date:  2001-05-23       Impact factor: 9.097

9.  A conversion formula for comparing pulse oximeter desaturation rates obtained with different averaging times.

Authors:  Jan Vagedes; Anja Bialkowski; Cornelia Wiechers; Christian F Poets; Klaus Dietz
Journal:  PLoS One       Date:  2014-01-28       Impact factor: 3.240

10.  Development of a luminous textile for reflective pulse oximetry measurements.

Authors:  Marek Krehel; Martin Wolf; Luciano F Boesel; René M Rossi; Gian-Luca Bona; Lukas J Scherer
Journal:  Biomed Opt Express       Date:  2014-07-09       Impact factor: 3.732

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.