Reduction of false arterial blood pressure alarms using signal quality assessment and relationships between the electrocardiogram and arterial blood pressure.

W Zong1, G B Moody, R G Mark.   

Abstract

The paper presents an algorithm for reducing false alarms related to changes in arterial blood pressure (ABP) in intensive care unit (ICU) monitoring. The algorithm assesses the ABP signal quality, analyses the relationship between the electrocardiogram and ABP using a fuzzy logic approach and post-processes (accepts or rejects) ABP alarms produced by a commercial monitor. The algorithm was developed and evaluated using unrelated sets of data from the MIMIC database. By rejecting 98.2% (159 of 162) of the false ABP alarms produced by the monitor using the test set of data, the algorithm was able to reduce the false ABP alarm rate from 26.8% to 0.5% of ABP alarms, while accepting 99.8% (441 of 442) of true ABP alarms. The results show that the algorithm is effective and practical, and its use in future patient monitoring systems is feasible.

Entities:  

Mesh:

Year:  2004        PMID: 15503972     DOI: 10.1007/bf02347553

Source DB:  PubMed          Journal:  Med Biol Eng Comput        ISSN: 0140-0118            Impact factor:   2.602


  9 in total

1.  Artifact detection in cardiovascular time series monitoring data from preterm infants.

Authors:  C Cao; I S Kohane; N McIntosh
Journal:  Proc AMIA Symp       Date:  1999

2.  PhysioBank, PhysioToolkit, and PhysioNet: components of a new research resource for complex physiologic signals.

Authors:  A L Goldberger; L A Amaral; L Glass; J M Hausdorff; P C Ivanov; R G Mark; J E Mietus; G B Moody; C K Peng; H E Stanley
Journal:  Circulation       Date:  2000-06-13       Impact factor: 29.690

3.  Multiple signal integration by decision tree induction to detect artifacts in the neonatal intensive care unit.

Authors:  C L Tsien; I S Kohane; N McIntosh
Journal:  Artif Intell Med       Date:  2000-07       Impact factor: 5.326

4.  Influence of pulse oximeter settings on the frequency of alarms and detection of hypoxemia: Theoretical effects of artifact rejection, alarm delay, averaging, median filtering or a lower setting of the alarm limit.

Authors:  A T Rheineck-Leyssius; C J Kalkman
Journal:  J Clin Monit Comput       Date:  1998-04       Impact factor: 2.502

5.  Robust sensor fusion improves heart rate estimation: clinical evaluation.

Authors:  J M Feldman; M H Ebrahim; I Bar-Kana
Journal:  J Clin Monit       Date:  1997-11

6.  Poor prognosis for existing monitors in the intensive care unit.

Authors:  C L Tsien; J C Fackler
Journal:  Crit Care Med       Date:  1997-04       Impact factor: 7.598

7.  The median filter as a preprocessor for a patient monitor limit alarm system in intensive care.

Authors:  A Mäkivirta; E Koski; A Kari; T Sukuvaara
Journal:  Comput Methods Programs Biomed       Date:  1991 Feb-Mar       Impact factor: 5.428

8.  Crying wolf: false alarms in a pediatric intensive care unit.

Authors:  S T Lawless
Journal:  Crit Care Med       Date:  1994-06       Impact factor: 7.598

9.  Preliminary clinical trials of a computer-based cardiac arrest alarm.

Authors:  A D Crew; K D Stoodley; R Lu; S Old; M Ward
Journal:  Intensive Care Med       Date:  1991       Impact factor: 17.440

  9 in total
  24 in total

1.  Is the Sequence of SuperAlarm Triggers More Predictive Than Sequence of the Currently Utilized Patient Monitor Alarms?

Authors:  Yong Bai; Duc Do; Quan Ding; Jorge Arroyo Palacios; Yalda Shahriari; Michele M Pelter; Noel Boyle; Richard Fidler; Xiao Hu
Journal:  IEEE Trans Biomed Eng       Date:  2016-06-30       Impact factor: 4.538

2.  Model-based data integration in clinical environments.

Authors:  Thomas Heldt; George C Verghese
Journal:  Annu Int Conf IEEE Eng Med Biol Soc       Date:  2010

3.  Predicting electrocardiogram and arterial blood pressure waveforms with different Echo State Network architectures.

Authors:  Allan Fong; Ranjeev Mittu; Raj Ratwani; James Reggia
Journal:  AMIA Annu Symp Proc       Date:  2014-11-14

Review 4.  Smart health monitoring systems: an overview of design and modeling.

Authors:  Mirza Mansoor Baig; Hamid Gholamhosseini
Journal:  J Med Syst       Date:  2013-01-15       Impact factor: 4.460

Review 5.  A review of methods for the signal quality assessment to improve reliability of heart rate and blood pressures derived parameters.

Authors:  Nicolò Gambarotta; Federico Aletti; Giuseppe Baselli; Manuela Ferrario
Journal:  Med Biol Eng Comput       Date:  2016-02-23       Impact factor: 2.602

Review 6.  -Omic and Electronic Health Record Big Data Analytics for Precision Medicine.

Authors:  Po-Yen Wu; Chih-Wen Cheng; Chanchala D Kaddi; Janani Venugopalan; Ryan Hoffman; May D Wang
Journal:  IEEE Trans Biomed Eng       Date:  2016-10-10       Impact factor: 4.538

7.  False alarm reduction in critical care.

Authors:  Gari D Clifford; Ikaro Silva; Benjamin Moody; Qiao Li; Danesh Kella; Abdullah Chahin; Tristan Kooistra; Diane Perry; Roger G Mark
Journal:  Physiol Meas       Date:  2016-07-25       Impact factor: 2.833

8.  The Effect of Signal Quality on Six Cardiac Output Estimators.

Authors:  T Chen; Gd Clifford; Rg Mark
Journal:  Comput Cardiol       Date:  2009-01-01

9.  Reducing false alarm rates for critical arrhythmias using the arterial blood pressure waveform.

Authors:  Anton Aboukhalil; Larry Nielsen; Mohammed Saeed; Roger G Mark; Gari D Clifford
Journal:  J Biomed Inform       Date:  2008-03-21       Impact factor: 6.317

10.  Robust parameter extraction for decision support using multimodal intensive care data.

Authors:  G D Clifford; W J Long; G B Moody; P Szolovits
Journal:  Philos Trans A Math Phys Eng Sci       Date:  2009-01-28       Impact factor: 4.226