PURPOSE: To investigate agreement between 2 swept source OCT biometers, IOL Master700 and Anterion, in various ocular biometry and intraocular lens (IOL) calculations of primary angle-closure disease (PACD).Entities:
Mesh:
Year: 2022 PMID: 35312733 PMCID: PMC8936461 DOI: 10.1371/journal.pone.0265844
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.240
Number of PACD patients and their demographic data.
| Data | PACD |
|---|---|
| Number of patients | 45 |
| • Male | 16 (35.6%) |
| • Female | 29 (64.4%) |
| Number of eyes | 69 |
| Age (years) | 63.27±8.01 (48–77) |
| Diagnosis, N eye (%) | |
| • PACS | 45 (65.22%) |
| • PAC | 11 (15.94%) |
| • PACG | 13 (18.84%) |
| Visual acuity (logMAR) | 0.32±0.48 (0.0–2.7) |
| Spherical equivalent (D) | 0.45±1.97 (-4.0–5.5) |
| Intraocular pressure (mmHg) | 16.54±4.27 (6–34) |
| Cup-to-disc ratio | 0.62±0.19 (0.3–1.0) |
PACD: Primary angle closure disease, PACS: Primary angle closure suspect, PAC: Primary angle closure
PACG: Primary angle closure glaucoma, D: diopter(s)
Comparison of various biometry and intraclass correlation coefficient between IOLMaster700 and Anterion in PACD.
| Parameter | IOLMaster 700 | Anterion | Mean difference ±SD | P value | ICC | Confident limit | |
|---|---|---|---|---|---|---|---|
| Mean ± SD | Mean ± SD | lower | upper | ||||
| ACD (mm) | 2.48 ± 0.37 | 2.58 ± 0.33 | -0.10 ± 0.09 | <0.001 | 0.932 | 0.529 | 0.978 |
| AL (mm) | 22.94 ± 0.83 | 22.95 ± 0.83 | -0.01 ± 0.02 | 0.003 | 1.000 | 1 | 1 |
| K1 (D) | 43.66 ± 1.22 | 43.47 ± 1.18 | 0.20 ± 0.29 | <0.001 | 0.953 | 0.861 | 0.979 |
| K2 (D) | 44.62 ± 1.37 | 44.37 ± 1.28 | 0.25 ± 0.27 | <0.001 | 0.950 | 0.778 | 0.98 |
| LT (mm) | 5.00 ± 0.32 | 5.04 ± 0.30 | -0.04 ± 0.10 | 0.003 | 0.947 | 0.905 | 0.97 |
| WTW (mm) | 11.71 ± 0.44 | 11.52 ± 0.51 | 0.19 ± 0.28 | <0.001 | 0.775 | 0.477 | 0.888 |
| LAF | 2.18 ± 0.17 | 2.20 ± 0.17 | -0.02 ± 0.04 | 0.005 | 0.967 | 0.941 | 0.981 |
| IOL power (D) | 22.30 ± 2.23 | 22.64 ± 2.34 | -0.34 ± 0.34 | <0.001 | 0.978 | 0.831 | 0.993 |
| EPR (D) | 0.00 ± 0.11 | -0.02 ± 0.11 | 0.01 ± 0.17 | 0.488 | -0.249 | -0.461 | -0.11 |
Values are presented as mean ± SD,
* = significant P value <0.05
a p-value by independent t-test
b p-value by Mann-Whitney U test
PACD: Primary angle closure disease, ACD: Anterior chamber depth, AL: Axial length, K: Keratometry, LT: Lens thickness, WTW: White-to-white corneal diameter, LAF: Lens-axial length factor, EPR: Estimated post-operative residual refraction, ICC: intraclass correlation coefficient, IOL: intraocular lens, D: diopter(s)
Comparison of various biometry and intraclass correlation coefficient in PACD with ACD ≤ 2.4 mm, between IOLMaster700 and Anterion.
| Parameter | IOLMaster 700 | Anterion | Mean difference ±SD | P value | ICC | Confident limit | |
|---|---|---|---|---|---|---|---|
| Mean ± SD | Mean ± SD | lower | upper | ||||
| ACD (mm) | 2.13 ± 0.27 | 2.25 ± 0.18 | -0.13 ± 0.14 | <0.001 | 0.701 | 0.176 | 0.885 |
| AL (mm) | 22.75 ± 0.80 | 22.76 ± 0.80 | -0.01 ± 0.02 | 0.017 | 1.000 | 0.999 | 1 |
| K1 (D) | 43.78 ± 1.38 | 43.56 ± 1.34 | 0.22 ± 0.25 | <0.001 | 0.971 | 0.824 | 0.991 |
| K2 (D) | 44.80 ± 1.68 | 44.53 ± 1.61 | 0.27 ± 0.23 | <0.001 | 0.978 | 0.712 | 0.994 |
| LT (mm) | 5.15 ± 0.23 | 5.17 ± 0.21 | -0.02 ± 0.14 | 0.583 | 0.819 | 0.626 | 0.917 |
| WTW (mm) | 11.54 ± 0.50 | 11.31 ± 0.60 | 0.23 ± 0.33 | <0.001 | 0.764 | 0.399 | 0.903 |
| LAF | 2.27 ± 0.11 | 2.28 ± 0.11 | -0.00 ± 0.06 | 0.706 | 0.862 | 0.703 | 0.939 |
| IOL power (D) | 22.60 ± 1.55 | 23.00 ± 1.68 | -0.40 ± 0.36 | <0.001 | 0.948 | 0.556 | 0.985 |
| EPR (D) | 0.00 ± 0.11 | -0.02 ± 0.13 | 0.02 ± 0.19 | 0.584 | -0.269 | -0.627 | 0.159 |
Values are presented as mean ± SD
* = significant P value <0.05
a p-value by independent t-test,
b p-value by Mann-Whitney U test
PACD: Primary angle closure disease, ACD: Anterior chamber depth, AL: Axial length, K: Keratometry, LT: Lens thickness, WTW: White-to-white corneal diameter, LAF: Lens-axial length factor, EPR: Estimated post-operative residual refraction, ICC: intraclass correlation coefficient, IOL: intraocular lens, D: diopter(s)
Comparison of various biometry and intraclass correlation coefficient in PACD with ACD >2.4 mm, between IOLMaster700 and Anterion.
| Parameter | IOLMaster 700 | Anterion | Mean difference ±SD | P value | ICC | Confident limit | |
|---|---|---|---|---|---|---|---|
| Mean ± SD | Mean ± SD | lower | upper | ||||
| ACD (mm) | 2.68 ± 0.25 | 2.77 ± 0.24 | -0.08 ± 0.02 | <0.001 | 0.945 | 0.007 | 0.989 |
| AL (mm) | 23.07 ± 0.85 | 23.07 ± 0.85 | -0.01 ± 0.02 | 0.01 | 1.000 | 1 | 1 |
| K1 (D) | 43.64 ± 1.14 | 43.45 ± 1.11 | 0.19 ± 0.27 | <0.001 | 0.958 | 0.859 | 0.982 |
| K2 (D) | 44.57 ± 1.17 | 44.32 ± 1.07 | 0.25 ± 0.29 | <0.001 | 0.944 | 0.717 | 0.98 |
| LT (mm) | 4.92 ± 0.34 | 4.97 ± 0.32 | -0.05 ± 0.06 | <0.001 | 0.973 | 0.864 | 0.99 |
| WTW (mm) | 11.79 ± 0.38 | 11.62 ± 0.43 | 0.17 ± 0.25 | <0.001 | 0.753 | 0.423 | 0.884 |
| LAF | 2.14 ± 0.18 | 2.18 ± 0.18 | -0.02 ± 0.02 | <0.001 | 0.984 | 0.916 | 0.994 |
| IOL power (D) | 22.03 ± 2.53 | 22.35 ± 2.64 | -0.31 ± 0.33 | <0.001 | 0.985 | 0.889 | 0.995 |
| EPR (D) | 0.00 ± 0.10 | -0.01 ± 0.10 | 0.01 ± 0.16 | 0.718 | -0.226 | -0.501 | 0.084 |
Values are presented as mean ± SD
* = significant P value <0.05
a p-value by independent t-test
b p-value by Mann-Whitney U test
PACD: Primary angle closure disease, ACD: Anterior chamber depth, AL: Axial length, K: Keratometry, LT: Lens thickness, WTW: White-to-white corneal diameter, LAF: Lens-axial length factor, EPR: Estimated post-operative residual refraction, ICC: intraclass correlation coefficient, IOL: intraocular lens, D: diopter(s)
Fig 1The Bland-Altman plot of the IOL power using parameters from IOLMaster700 and Anterion.
The solid line represents the mean difference, whereas dotted lines on both sides represent the upper and lower 95% LoA. A: In all PACD eyes, The plot demonstrates that only two eyes from total 69 eyes were out of the 95% LoAs, indicating excellent agreement in IOL power. B: In the eyes with ACD≤2.4 mm, The plot demonstrates that only one eye from 24 eyes were out of the 95% LoAs, indicating excellent agreement in IOL power. C: In the eyes with ACD>2.4 mm. The plot demonstrates that only one eye from 45 eyes were out of the 95% LoAs, indicating excellent agreement in IOL power.
Fig 2The EPR between the devices’ measurements.
The Bland-Altman plot of the EPR using parameters from IOLMaster700 and Anterion of all PACD eyes. The solid line represents the mean difference, whereas dotted lines on both sides represent the upper and lower 95% LoA. A: In all PACD eyes, the plot demonstrates that all cases were within the 95% LoAs, indicating excellent agreement in EPR. B: In the eyes with ACD ≤ 2.4 mm, all cases were within the 95% LoAs, which indicates excellent agreement between the devices. C: In the eyes with ACD > 2.4 mm, the plot demonstrates that only one eye was out of the 95% LoAs, indicating excellent agreement between the devices.