Literature DB >> 32860152

Agreement of white-to-white measurements with swept-source OCT, Scheimpflug and color LED devices.

Pedro Tañá-Rivero1, Salvador Aguilar-Córcoles1, José Luís Rodríguez-Prats1, Robert Montés-Micó2, Ramón Ruiz-Mesa1.   

Abstract

PURPOSE: To assess the interchangeability of different devices for measuring white-to-white (WTW) distance.
METHODS: WTW distance was measured in 53 eyes of 53 patients using Anterion swept-source optical coherence topographer (SS-OCT), IOLMaster 700 SS-OCT, Pentacam HR Scheimpflug and Cassini color LED. Statistical analysis was done by means of the Friedman test and the post hoc Tukey test. The Bland-Altman analysis was applied to carry out pairwise comparisons with the average difference, 95% confidence interval of the average difference and limits of agreement 95% (LoA).
RESULTS: WTW values obtained by the Anterion, IOLMaster 700, Pentacam HR and Cassini were: 11.84 ± 0.41 mm, 11.96 ± 0.41 mm, 11.68 ± 0.38 mm and 12.65 ± 0.52 mm, respectively. Statistically significant differences were found in all pairwise comparison (p < 0.001). The lowest mean difference was found between the Anterion and IOLMaster 700 (- 0.11 mm) and the highest between the Pentacam HR and Cassini (- 0.96 mm). The widest LoA ranges were those that compared any device with the Cassini. LoA ranges when the other three devices were compared among them were similar: Anterion versus IOLMaster 700, Anterion versus Pentacam HR and IOLMaster versus Pentacam HR (about 0.2 mm).
CONCLUSIONS: Our results show that there were statistically significant differences in WTW measurement among the four devices, but under a clinical point of view, we believe that Anterion and IOLMaster 700 may be considered interchangeable and so Anterion and Pentacam HR, however, IOLMaster 700 and Pentacam HR may not and neither is Cassini with any of the other three devices.

Entities:  

Keywords:  Color LED; OCT; Scheimpflug; White-to-white

Mesh:

Year:  2020        PMID: 32860152     DOI: 10.1007/s10792-020-01552-9

Source DB:  PubMed          Journal:  Int Ophthalmol        ISSN: 0165-5701            Impact factor:   2.031


  1 in total

Review 1.  Device interchangeability on anterior chamber depth and white-to-white measurements: a thorough literature review.

Authors:  Alberto Domínguez-Vicent; Cari Pérez-Vives; Teresa Ferrer-Blasco; Santiago García-Lázaro; Robert Montés-Micó
Journal:  Int J Ophthalmol       Date:  2016-07-18       Impact factor: 1.779

  1 in total
  6 in total

1.  Agreement of total corneal power between 2 swept-source optical coherence tomography and Scheimpflug tomography in normal and keratoconic patients.

Authors:  Rosepon Asawaworarit; Vannarut Satitpitakul; Parichart Taweekitikul; Krit Pongpirul
Journal:  PLoS One       Date:  2022-05-24       Impact factor: 3.752

2.  Agreement in anterior segment measurements between swept-source and Scheimpflug-based optical biometries in keratoconic eyes: a pilot study.

Authors:  Evangelia Chalkiadaki; Panos S Gartaganis; Thomas Ntravalias; Ioannis Giannakis; Evangelos Manousakis; Efthymios Karmiris
Journal:  Ther Adv Ophthalmol       Date:  2022-03-31

3.  Evaluation of ocular biometry in primary angle-closure disease with two swept source optical coherence tomography devices.

Authors:  Boonsong Wanichwecharungruang; Anyarak Amornpetchsathaporn; Wisakorn Wongwijitsook; Kittipong Kongsomboon; Somporn Chantra
Journal:  PLoS One       Date:  2022-03-21       Impact factor: 3.240

4.  A prediction model for sulcus-to-sulcus diameter in myopic eyes: a 1466-sample retrospective study.

Authors:  Qiu-Jian Zhu; Wei-Jian Zhu; Wen-Jing Chen; Lie Ma; You Yuan
Journal:  BMC Ophthalmol       Date:  2022-07-16       Impact factor: 2.086

5.  Agreement between swept-source optical biometry and Scheimpflug-based tomography in eyes with previous myopic refractive surgery.

Authors:  Efthymios Karmiris; Panos S Gartaganis; Thomas Ntravalias; Evangelos Manousakis; Ioannis Giannakis; Evangelia Chalkiadaki
Journal:  Saudi J Ophthalmol       Date:  2022-08-29

6.  Anterior-Segment Swept-Source Ocular Coherence Tomography and Scheimpflug Imaging Agreement for Keratometry and Pupil Measurements in Healthy Eyes.

Authors:  Francisco Pérez-Bartolomé; Carlos Rocha-De-Lossada; José-María Sánchez-González; Silvia Feu-Basilio; Josep Torras-Sanvicens; Jorge Peraza-Nieves
Journal:  J Clin Med       Date:  2021-12-10       Impact factor: 4.241

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.