Abstract

PURPOSE: To assess the repeatability of ocular biometric parameters using a high-resolution imaging device.
METHODS: 74 healthy right eyes were included in this study. Five-repeated measurements were taken with ANTERION high-resolution swept-source optical coherence tomographer (SS-OCT) to measure: corneal thickness (central and at 2, 4 and 6-mm diameters), aqueous depth (AD), lens thickness (LT), anterior chamber volume (ACV), axial length (AL), and pupil (diameter and position). The intrasubject standard deviation (Sw), coefficient of repeatability (CoR) and intraclass correlation coefficient (ICC) were calculated. Bland-Altman method was applied to analyze the difference between the first and the last measurement. The average and the difference between both measurements were calculated for all parameters.
RESULTS: We have not found statistically significant differences between repeated measurements (p > 0.05). The mean difference for corneal thickness was between -0.08 and 0.28 μm. For AD and LT was 0.004 and -0.004 μm, respectively. ACV mean difference was -0.03 mm3 and for AL was 0.001 mm. Pupil diameter and position mean differences ranged between -0.008 and 0.009 mm. Overall, most ocular parameters had a Sw <1 and a CoR <2 in their respective units, and an ICC >0.92.
CONCLUSIONS: The ANTERION high-resolution SS-OCT device provides good repeatability for different ocular biometric measurements.

Entities:  

Keywords:  ANTERION; Repeatability; anterior segment; biometry; optical coherence tomography

Year:  2020        PMID: 32425075     DOI: 10.1080/17434440.2020.1772050

Source DB:  PubMed          Journal:  Expert Rev Med Devices        ISSN: 1743-4440            Impact factor:   3.166


  7 in total

1.  Repeatability of a new swept-source optical coherence tomographer and agreement with other three optical biometers.

Authors:  Shi-Ming Cheng; Jia-Sheng Zhang; Xu Shao; Ze-Tong Wu; Tian-Tian Li; Po Wang; Jun-Hai Lin; A-Yong Yu
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2022-02-16       Impact factor: 3.117

2.  Comparison of acquisition rate and agreement of axial length with two swept-source optical coherence tomographers and a partial coherence interferometer.

Authors:  Shi-Ming Cheng; Wen-Tao Yan; Jia-Sheng Zhang; Tian-Tian Li; Xi Li; A-Yong Yu
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2022-04-30       Impact factor: 3.535

3.  Repeatability of automated measurements by a new anterior segment optical coherence tomographer and biometer and agreement with standard devices.

Authors:  Domenico Schiano-Lomoriello; Kenneth J Hoffer; Irene Abicca; Giacomo Savini
Journal:  Sci Rep       Date:  2021-01-13       Impact factor: 4.379

4.  Refractive prediction of four different intraocular lens calculation formulas compared between new swept source optical coherence tomography and partial coherence interferometry.

Authors:  Mi Yeon Song; Sung Rae Noh; Kook Young Kim
Journal:  PLoS One       Date:  2021-05-04       Impact factor: 3.240

5.  Lens-vault analysis and its correlation with other biometric parameters using swept-source OCT.

Authors:  Pedro Tañá-Rivero; Ramón Ruiz-Mesa; Salvador Aguilar-Córcoles; Cristina Tello-Elordi; María Ramos-Alzamora; Robert Montés-Micó
Journal:  J Optom       Date:  2021-11-01

6.  Clinical evaluation of ocular biometry of dual Scheimpflug analyzer, GALILEI G6 and swept source optical coherence tomography, ANTERION.

Authors:  Boonsong Wanichwecharungruang; Anyarak Amornpetchsathaporn; Kittipong Kongsomboon; Wisakorn Wongwijitsook; Kornkamol Annopawong; Somporn Chantra
Journal:  Sci Rep       Date:  2022-03-04       Impact factor: 4.379

7.  Evaluation of ocular biometry in primary angle-closure disease with two swept source optical coherence tomography devices.

Authors:  Boonsong Wanichwecharungruang; Anyarak Amornpetchsathaporn; Wisakorn Wongwijitsook; Kittipong Kongsomboon; Somporn Chantra
Journal:  PLoS One       Date:  2022-03-21       Impact factor: 3.240

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.