Shuying Yang1, Yuan Qin1, Xueyan Guo2, Yaqi Li2.
Abstract
Purpose: To examine the effects of fluid contamination on the reverse torque value (RTV) of abutment screws. 484 titanium fixtures were mounted into the stainless-steel holders.Entities:
Mesh:
Substances:
Year: 2022 PMID: 35299888 PMCID: PMC8923780 DOI: 10.1155/2022/4111710
Source DB: PubMed Journal: Biomed Res Int Impact factor: 3.411
Figure 1The cyclic loading device (SD Mechatronic, Feld-krichen, Westerham, Germany).
Descriptive statistics and torque loss values of all groups.
| N | Mean | Std. deviation | Std. error | 95% confidence interval for mean | Minimum | Maximum | Torque loss | ||
|---|---|---|---|---|---|---|---|---|---|
| Lower bound | Upper bound | ||||||||
| Control | 44 | 18.00 | 1.78 | 0.62 | 17.56 | 20.43 | 17.00 | 23.00 | 36.70% |
| Blood | 44 | 22.12 | 1.56 | 0.51 | 19.93 | 22.29 | 19.00 | 23.00 | 29.7% |
| Saliva | 44 | 21.56 | 1.43 | 0.44 | 19.53 | 21.58 | 19.00 | 22.00 | 31.70% |
| Flouride | 44 | 24.00 | 1.02 | 0.37 | 23.14 | 24.85 | 23.00 | 26.00 | 20.00% |
| CHX | 44 | 21.89 | 1.25 | 0.38 | 21.99 | 23.78 | 21.00 | 25.00 | 23.70% |
| Blood+saliva | 44 | 22.00 | 1.54 | 0.42 | 19.16 | 21.53 | 18.00 | 25.00 | 29.70% |
| Blood+flouride | 44 | 20.11 | 1.58 | 0.71 | 20.73 | 24.29 | 17.00 | 24.00 | 28.7% |
| Blood+CHX | 44 | 22.56 | 1.73 | 0.54 | 21.53 | 22.18 | 19.00 | 23.00 | 30.80% |
| Saliva+flouride | 44 | 23.51 | 1.19 | 0.47 | 23.51 | 23.15 | 21.00 | 25.00 | 28.00% |
| Saliva+CHX | 44 | 23.89 | 1.92 | 0.68 | 20.89 | 24.58 | 22.00 | 26.00 | 29.60% |
| Flouride+CHX | 44 | 21.02 | 1.35 | 0.58 | 21.19 | 22.48 | 20.00 | 25.00 | 28.10% |
Comparison of the mean difference of reverse torque value in the studied groups.
| Tukey HSD | |||||
|---|---|---|---|---|---|
| Study groups | Mean difference | 95% confidence interval | |||
| Lower bound | Upper bound |
| |||
| Control | Blood | -2.11 | -4.04 | -0.18 | 0.041 |
| Control | Saliva | -1.56 | -3.48 | 0.37 | 0.16 |
| Control | Fluoride | -5.00 | -6.93 | -3.07 | 0.033 |
| Control | CHX | -3.89 | -5.82 | -1.96 | 0.011 |
| Blood | Saliva | 0.56 | -1.37 | 2.48 | 0.92 |
| Blood | Fluoride | -2.89 | -4.82 | -0.96 | 0.001 |
| Blood | CHX | -1.78 | -3.71 | 0.15 | 0.08 |
| Saliva | Fluoride | -3.44 | -5.37 | -1.52 | <0.001 |
| Saliva | CHX | -2.33 | -4.26 | -0.41 | 0.01 |
| Fluoride | CHX | 1.11 | -0.82 | 3.04 | 0.48 |
| Blood+saliva | Control | -2.14 | -7.11 | 5.89 | 0.04 |
| Blood+fluoride | Control | -2.08 | -3.96 | -0.10 | 0.04 |
| Blood+CHX | Control | -4.1 | -2.30 | 0.48 | 0.021 |
| Saliva+fluoride | Control | -3.5 | -4.29 | -0.38 | 0.01 |
| Saliva+CHX | Control | -3.75 | -7.9 | -2.07 | 0.01 |
| Fluoride+CHX | Control | -5.12 | -6.95 | -1.30 | 0.031 |
The mean difference is significant at the 0.05 level.
Figure 2Comparison of the mean difference of reverse torque value in the studied groups.