Torque removal evaluation of prosthetic screws after tightening and loosening cycles: an in vitro study.

Mayra Cardoso1, Marcelo Ferreira Torres, Eduardo José Veras Lourenço, Daniel de Moraes Telles, Renata Cristina Silveira Rodrigues, Ricardo Faria Ribeiro.   

Abstract

OBJECTIVES: The aim of this study was to evaluate the variation in removal torque of implant prosthetic abutment screws after successive tightening and loosening cycles, in addition to evaluating the influence of the hexagon at the abutment base on screw removal torque.
MATERIAL AND METHODS: Twenty hexagonal abutments were tightened to 20 regular external hex implants with a titanium alloy screw, with an insertion torque of 32 N cm, measured with a digital torque gauge. The implant/abutment/screw assemblies were divided into two groups: (1) abutments without hexagon at the base and (2) abutments with a hexagon at the base. Each assembly received a provisional restoration and was submitted to mechanical loading cycles. After this, the screws were removed and the removal torque was measured. This sequence was repeated 10 times, then the screw was replaced by a new one, and another cycle was performed. Linear regression analysis was performed.
RESULTS: Removal torque values tended to decrease as the number of insertion/removal cycles increased, for both groups. Comparisons of the slopes and the intercepts between groups showed no statistical difference. There was no significant difference between the mean values of last five cycles and the 11th cycle. Within the limitations of this in vitro study, it was concluded that (1) repeated insertion/removal cycles promoted gradual reduction in removal torque of screws, (2) replacing the screw with a new one after 10 cycles did not increase resistance to loosening, and (3) removal of the hexagon from the abutment base had no effect on the removal torque of the screws.
© 2011 John Wiley & Sons A/S.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21457352     DOI: 10.1111/j.1600-0501.2011.02165.x

Source DB:  PubMed          Journal:  Clin Oral Implants Res        ISSN: 0905-7161            Impact factor:   5.977


  6 in total

1.  Removal torque evaluation of three different abutment screws for single implant restorations after mechanical cyclic loading.

Authors:  T Paepoemsin; P A Reichart; P Chaijareenont; F P Strietzel; P Khongkhunthian
Journal:  Oral Implantol (Rome)       Date:  2016-11-16

2.  Effect of cyclic loading and retightening on reverse torque value in external and internal implants.

Authors:  Woong-Rae Cho; Yoon-Hyuk Huh; Chan-Jin Park; Lee-Ra Cho
Journal:  J Adv Prosthodont       Date:  2015-08-18       Impact factor: 1.904

3.  Correlation between microleakage and screw loosening at implant-abutment connection.

Authors:  Cem Sahin; Simel Ayyildiz
Journal:  J Adv Prosthodont       Date:  2014-02-14       Impact factor: 1.904

4.  Stability of the prosthetic screws of three types of craniofacial prostheses retention systems.

Authors:  Antonio Gabriel Lanata-Flores; Eder Alberto Sigua-Rodriguez; Douglas Rangel Goulart; Veber Luiz Bomfim-Azevedo; Sergio Olate; José Ricardo de Albergaria-Barbosa
Journal:  J Korean Assoc Oral Maxillofac Surg       Date:  2016-12-27

5.  An In Vitro Study of Fluid Contaminations Influences on Reverse Torque Values of Implant-Abutment Connections.

Authors:  Shuying Yang; Yuan Qin; Xueyan Guo; Yaqi Li
Journal:  Biomed Res Int       Date:  2022-03-08       Impact factor: 3.411

6.  Healing Abutment Distortion in Implant Prostheses: An In Vitro Study.

Authors:  Chandrashekhar Pandey; Bishwa Prakash Bhattarai; Apiwat Riddhabhaya; Natthamet Wongsirichat; Dinesh Rokaya
Journal:  J Funct Biomater       Date:  2022-06-21
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.