The effect of 3 torque delivery systems on gold screw preload at the gold cylinder-abutment screw joint.

Keson B Tan1, Jack I Nicholls.   

Abstract

PURPOSE: This study measured the gold screw preload at the gold cylinder-abutment screw joint interface obtained by 3 torque delivery systems.
MATERIALS AND METHODS: Using a precalibrated, strain-gauged standard abutment as the load cell, 3 torque delivery systems tested were shown to have significant differences in gold screw preload when a gold cylinder was attached.
RESULTS: Mean preloads measured were 291.2 N for hand torque drivers set at 10 Ncm, 340.3 N for electronic torque controllers at low setting/10 Ncm, 384.4 N for electronic torque controllers at high setting/10 Ncm; and 140.8 N for hand-tightening with a prosthetic slot screwdriver. Significant differences in screw preload were also found between operators using a hand torque driver. DISCUSSION: Hand-tightening delivered insufficient preload and cannot be recommended for final gold screw tightening. Different electronic torque controller units set at 10 Ncm induced mean gold screw preloads that ranged from 264.1 N to as high as 501.2 N.
CONCLUSION: Electronic torque controllers should be regularly recalibrated to ensure optimal output.

Entities:  

Mesh:

Substances:

Year:  2002        PMID: 11958399

Source DB:  PubMed          Journal:  Int J Oral Maxillofac Implants        ISSN: 0882-2786            Impact factor:   2.804


  5 in total

1.  An experimental and theoretical composite model of the human mandible.

Authors:  R De Santis; F Mollica; R Esposito; L Ambrosio; L Nicolais
Journal:  J Mater Sci Mater Med       Date:  2005-12       Impact factor: 3.896

2.  Effect of dynamic cyclic loading on screw loosening of retightened versus new abutment screw in both narrow and standard implants (in-vitro study).

Authors:  Eman Mohammed Nasr Attiah; Attiah Ali AlGendy; Tamer Mohamed Nasr Mostafa
Journal:  Int J Implant Dent       Date:  2020-07-28

3.  Accuracy of different electronic torque drivers: A comparative evaluation.

Authors:  Byeong-Dae Ko; KeunBaDa Son; Seok Hyon Kang; Kyu-Bok Lee
Journal:  J Adv Prosthodont       Date:  2019-12-18       Impact factor: 1.904

4.  An In Vitro Study of Fluid Contaminations Influences on Reverse Torque Values of Implant-Abutment Connections.

Authors:  Shuying Yang; Yuan Qin; Xueyan Guo; Yaqi Li
Journal:  Biomed Res Int       Date:  2022-03-08       Impact factor: 3.411

5.  Digital evaluation of axial displacement by implant-abutment connection type: An in vitro study.

Authors:  Sung-Jun Kim; KeunBaDa Son; Kyu-Bok Lee
Journal:  J Adv Prosthodont       Date:  2018-10-22       Impact factor: 1.904

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.