Literature DB >> 16650006

Effect of casting procedures on screw loosening in UCLA-type abutments.

Stefania C Kano1, Paul Binon, Gerson Bonfante, Donald A Curtis.   

Abstract

BACKGROUND: Screw loosening of implant restorations continues to be a complication in implant prosthodontics. Screw joints are subjected to a loss of initially applied torque because of friction and component misfit. It has been suggested that the loss of applied torque is less in machined metal abutments than in cast plastic abutments.
PURPOSE: This study compared the loss of applied torque (detorque) values in machined titanium and in cast UCLA-type abutments for external hex abutment/implant interface.
MATERIALS AND METHODS: Four groups of 12 samples each were evaluated: (1) machined titanium abutments, (2) premachined palladium abutments cast with palladium, (3) plastic abutments cast with nickel-chromium, and (4) plastic abutments cast with cobalt-chromium. Each abutment was torqued to 30 Ncm according to the manufacturer's instructions and detorqued three times. The mean loss of applied torque (detorque) was recorded as a percentage of the torque applied. Group means were calculated and compared using ANOVA and Tukey's LSD test.
RESULTS: Mean detorque values were (1) 92.3 +/- 2.9%, (2) 81.6 +/- 5.0%, (3) 86.4 +/- 4.6%, and (4) 84.0 +/- 7.0%. Machined abutments demonstrated significantly greater detorque values compared with all cast groups (p < 0.05). No significant differences were found among cast groups.
CONCLUSION: Machined abutments retained a significantly greater percentage of torque compared with cast abutments. Casting procedures decrease the percentage of applied torque, which may influence final screw joint stability. Copyright 2006 by The American College of Prosthodontists.

Entities:  

Mesh:

Year:  2006        PMID: 16650006     DOI: 10.1111/j.1532-849X.2006.00078.x

Source DB:  PubMed          Journal:  J Prosthodont        ISSN: 1059-941X            Impact factor:   2.752


  8 in total

1.  Peri-implantitis and the prosthodontist.

Authors:  A Dawood; B Marti Marti; S Tanner
Journal:  Br Dent J       Date:  2017-09-08       Impact factor: 1.626

2.  Comparison of fit accuracy and torque maintenance of zirconia and titanium abutments for internal tri-channel and external-hex implant connections.

Authors:  Hakimeh Siadat; Elaheh Beyabanaki; Niloufar Mousavi; Marzieh Alikhasi
Journal:  J Adv Prosthodont       Date:  2017-08-16       Impact factor: 1.904

3.  Sealing agent reduces formation of single and dual-species biofilms of Candida albicans and Enterococcus faecalis on screw joints at the abutment/implant interface.

Authors:  Cecília Alves de Sousa; Jadison Junio Conforte; Karina Sampaio Caiaffa; Cristiane Duque; Wirley Gonçalves Assunção
Journal:  PLoS One       Date:  2019-10-22       Impact factor: 3.240

4.  Analysis of Torque Maintenance and Fracture Resistance after Fatigue in Retention Screws Made of Different Metals for Screw-Retained Implant-Borne Prosthesis Joints.

Authors:  Maria Beatriz Bello Taborda; Gabriela Sumie Yaguinuma Gonçalves; Cecília Alves de Sousa; Wirley Gonçalves Assunção
Journal:  Int J Dent       Date:  2021-11-18

5.  An In Vitro Study of Fluid Contaminations Influences on Reverse Torque Values of Implant-Abutment Connections.

Authors:  Shuying Yang; Yuan Qin; Xueyan Guo; Yaqi Li
Journal:  Biomed Res Int       Date:  2022-03-08       Impact factor: 3.411

6.  Accuracy combining different brands of implants and abutments.

Authors:  María-Fernanda Solá-Ruíz; Eduardo Selva-Otaolaurruchi; Gisela Senent-Vicente; Inés González-de-Cossio; Vicente Amigó-Borrás
Journal:  Med Oral Patol Oral Cir Bucal       Date:  2013-03-01

7.  Tightening torque of implant abutment using hand drivers against torque wrench and its effect on the internal surface of implant.

Authors:  Y Sameera; Rathika Rai
Journal:  J Indian Prosthodont Soc       Date:  2020-04-07

8.  Comparison of the Effect of Four Different Abutment Screw Torques on Screw Loosening in Single Implant-Supported Prosthesis after the Application of Mechanical Loading.

Authors:  Behnaz Ebadian; Amirhossein Fathi; Saba Khodadad
Journal:  Int J Dent       Date:  2021-07-19
  8 in total