Implant abutment screw reverse torque values before and after plasma cleaning.

Costanza Micarelli1, Luigi Canullo, Paolo Baldissara, Marco Clementini.   

Abstract

This in vitro study analyzed the reverse torque (RevT) of abutment screws following different cleaning treatments. A convenience sample of 50 customized titanium abutment screw complexes was divided into five groups: cleaning by steam (control group), cleaning by Argon plasma (test groups 1 and 2 [with chlorhexidine gel]), and replacement of old screws with new ones (test groups 3 and 4 [with chlorhexidine gel]). Abutments were screwed onto implants and tested for RevT. The RevT of the test groups was significantly higher than that of the control group. No statistically significant difference between test groups was noted except between groups 2 and 3.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 23837162     DOI: 10.11607/ijp.3396

Source DB:  PubMed          Journal:  Int J Prosthodont        ISSN: 0893-2174            Impact factor:   1.681


  3 in total

1.  Microscopical and chemical surface characterization of CAD/CAM zircona abutments after different cleaning procedures. A qualitative analysis.

Authors:  Peter Gehrke; Astrid Tabellion; Carsten Fischer
Journal:  J Adv Prosthodont       Date:  2015-04-23       Impact factor: 1.904

2.  The efficacy of convenient cleaning methods applicable for customized abutments: an in vitro study.

Authors:  Sunjai Kim; Changhun Choi; Yunsu Cha; Jae-Seung Chang
Journal:  BMC Oral Health       Date:  2021-02-18       Impact factor: 2.757

3.  An In Vitro Study of Fluid Contaminations Influences on Reverse Torque Values of Implant-Abutment Connections.

Authors:  Shuying Yang; Yuan Qin; Xueyan Guo; Yaqi Li
Journal:  Biomed Res Int       Date:  2022-03-08       Impact factor: 3.411

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.