Development of a Simple DNA Extraction Method and Candida Pan Loop-Mediated Isothermal Amplification Assay for Diagnosis of Candidemia.

Da Hye Lim1, Hyunseul Jee2, Kyung Chul Moon3, Chae Seung Lim1, Woong Sik Jang3.   

Abstract

To reduce the morbidity and mortality of candidemia patients through rapid treatment, the development of a simple, rapid molecular diagnostic method that is based on nucleic acid extraction and is superior to conventional methods for detecting Candida in the blood is necessary. We developed a multiplex Candida Pan/internal control (IC) loop-mediated isothermal amplification (LAMP) assay and a simple DNA extraction boiling protocol using Chelex-100 that could extract yeast DNA in blood within 20 min. The Chelex-100/boiling method for DNA extraction showed comparable efficiency to that of the commercial QIAamp UCP Pathogen Mini Kit using Candida albicans qPCR. In addition, the Candida Pan/IC LAMP assay showed superior sensitivity to that of general Candida Pan and species qPCRs against clinical DNA samples extracted with the QIAamp UCP Pathogen Mini Kit and Chelex-100/boiling method. The Candida Pan/IC LAMP assay followed by Chelex-100/boiling-mediated DNA extraction showed high sensitivity (100%) and specificity (100%) against clinical samples infected with Candida. These results suggest that the Candida Pan/IC LAMP assay could be used as a rapid molecular diagnostic test for candidemia.

Entities:  

Keywords:  Candida albicans; Candida spp.; Chelex-100; DNA extraction; multiplex LAMP

Year:  2022        PMID: 35215055      PMCID: PMC8878442          DOI: 10.3390/pathogens11020111

Source DB:  PubMed          Journal:  Pathogens        ISSN: 2076-0817


1. Introduction

Candida species are among the top five infectious bloodstream pathogens and remain the most common cause of invasive fungal infections [1]. Invasive candidiasis, which occurs when Candida spp. infect other tissues and organs, infects more than 250,000 people worldwide each year and causes more than 50,000 deaths [2,3]. The reported candidemia mortality ranges from 30% to 60% with up to 30 days of increase in the length of hospital stay for survivors [4]. Early diagnosis is critical for appropriate patient management and for improving the outcomes of candidemia. Blood cultures, the current diagnostic gold standard, are limited by low sensitivity, ranging from 21% to 71% [5], and a slow turnaround, usually exceeding 48 h [6,7,8]. Therefore, various non-culture-based diagnostic methods such as immunoassay (mannan, anti-mannan antibodies and (1-3)-β-d-glucan (BDG) assay) [9,10,11] and PCR [8] have been developed; however, detection methods using mannan, anti-mannan antibodies and BDG have been reported to have low specificity [10,12] and PCR-based diagnosis is time-consuming, although the detection specificity is high. Therefore, the development of a fast diagnosis system is required for the detection of fungi in blood. Recently, several isothermal amplification techniques for Candida have been proposed as molecular diagnostic methods to overcome these limitations, including loop-mediated isothermal amplification (LAMP) [13], nucleic acid sequence-based amplification (NASBA) [14] and rolling circle amplification [15]. Among these isothermal amplification methods, LAMP is the most extensively investigated method for Candida detection [13,16,17]. Inàcio et al. reported a LAMP technique for the amplification of the 26S rRNA gene in clinically relevant Candida yeasts [13]. Fallahi et al. developed a C. albicans-specific LAMP assay using fluorescence detection [16]. Hongling et al. established multiple pathogen loop-mediated isothermal amplification (LAMP) using microfluidic chip technology for Staphylococcus aureus, Escherichia coli, Pneumoniae Klebsiella, Shigella, methicillin-resistant Staphylococcus aureus (MRSA) and C. albicans [17]. LAMP is composed of six primers, of which four primers contain six parts of the target gene sequence and two loop primers react with the target gene to form a loop structure and then robustly amplify the target gene at 58–65 °C using Bst or Bsm polymerases with DNA-strand displacement activity [18,19]. To shorten the time for diagnosis, fast nucleic acid extraction from Candida in blood is needed. The sensitivity of any molecular diagnostic method for the detection of fungal pathogens depends on the lysis efficiency of fungal cells from blood samples and purification of DNA without PCR inhibitors [20]. In particular, the breakdown of the fungal cell wall is a crucial step for lysis of the cell entity and isolation of genomic DNA. Current fungal DNA extraction protocols involve enzymatic [21,22], chemical or physical disruption steps [23,24], bead beating using glass [25], or ceramic beads [26] to disrupt the fungal cell wall. Unlike nucleic acid extraction of animal cells or viruses, the additional cell wall disruption step makes rapid nucleic acid extraction from fungi more difficult. In this study, we developed a Chelex-100/boiling DNA extraction method (within 20 min) and Candida Pan/IC LAMP assay (40 min) for the rapid diagnosis of candidemia. The performance of the Chelex-100 DNA extraction method was compared and evaluated with that of the QIAamp UCP Pathogen Mini Kit using general Candida qPCR. In addition, the sensitivity and specificity of the Candida Pan/IC LAMP assay were compared with those of two general qPCRs (Candida Pan and Candida species) for two kinds of clinical sample DNAs, which were extracted using the QIAamp UCP Pathogen Mini Kit and Chelex-100/boiling DNA extraction.

2. Materials and Methods

2.1. Clinical Samples

A total of 136 clinical whole blood samples were collected from Candida-infected (n = 36) and non-infected patients (normal control, n = 100) at Korea University Guro Hospital from January 2019 to August 2021. All clinical samples were confirmed by VITEK 2 COMPACT system (bioMérieux, Durham, NC, USA) using a VITEK®2 YST ID card (bioMérieux, Durham, NC, USA). True positives included clinical blood samples infected with C. albicans (n = 9), C. glabrata (n = 9), C. tropicalis (n = 9) and C. parapsilosis (n = 9). For the cross-reactivity test, the cultured bacteria samples, including Escherichia coli, Enterococcus faecium, Klebsiella spp., Staphylococcus aureus and Staphylococcus epidermidis, were obtained from the Korea University Guro Hospital. The study was conducted in accordance with the guidelines of the Declaration of Helsinki and was approved by the Institutional Review Board of Korea University Guro Hospital (2020GR0512).

2.2. Isolation of Genomic DNA from Candida Strains

Candida albicans (CCARM 14029), C. krusei (CCARM 14017), C. tropicalis (CCARM14019), C. parapsilosis (CCARM14016), C. auris (KCTC17850) and C. glabrata (KCTC 7219) were obtained from the Culture Collection of Antimicrobial-Resistant Microbes (CCARM; Seoul, Korea) and Korean Collection for Type Cultures (KCTC; Daejeon, Korea). For DNA extraction from Candida cell stocks, the yeasts were grown on yeast peptone dextrose (YPD) broth (Difco BD, Milan, Italy) at 180 rpm and 37 °C overnight. After cell counting with phase-contrast microscopy (40 × power) using a counting grid, DNA was extracted from the Candida cells (~2 × 108 cells/mL) using the QIAamp UCP Pathogen Mini Kit according to the manufacturer’s manual.

2.3. Isolation of Candida Genomic DNA from Whole Blood

DNA extraction was performed using two different methods, the QIAamp UCP Pathogen Mini Kit (Qiagen, Hilden, Germany) and a newly developed simple boiling method using Chelex-100. First, DNA was extracted from the blood samples using the QIAamp UCP Pathogen Mini Kit and Lysing Matrix C tube (MP Biomedicals, Illkirch, France) according to the manufacturer’s instructions, and Candida DNA was eluted with 100 μL of elution buffer. Second, Candida DNA was extracted from the blood samples using the boiling method with Chelex-100 (Figure 1). Briefly, 200 µL of 2× red blood cell lysis buffer (Bio Basic, Toronto, ON, Canada) was added to 200 µL of whole blood sample. The sample was vortexed for 15 s first, followed by incubation at room temperature for 3 min and, later, centrifuged at 10,000 rpm for 5 min. After removing the supernatant, 500 µL of 10% Chelex-100 Resin (Bio-Rad Laboratories, Hercules, CA, USA) solution (10 mM Tris-HCl and 1 mM EDTA; pH 8.0) was added to the pellets. The suspension was boiled for 10 min at 100 °C in a heat block and then vortexed for 15 s three times. After filtration with a 3 µm Polycarbonate track-etched membrane filter (Whatman, Marlborough, Mass, USA), the supernatant was transferred to a new tube for subsequent experiments. The filtration process was performed using a SEPARA® tube (GVS, Bologna, Italy). A 3 µm membrane was attached to the filtering unit using instant adhesive (UNITECH, Gyeonggi, Korea) after removing the existing filter (0.2 µm).
Figure 1

Schematic of Chelex-100/boiling DNA extraction method.

2.4. Primer Design

The Candida Pan LAMP primer set was designed within the conserved regions of partial ITS1, 5.8S rRNA gene and partial ITS2 of 6 Candida species (C. albicans MT640022.1_70-499, C. glabrata MT548912.1_350-885, C. krusei MZ507554.1_50-538, C. tropicalis LC639851.1_50-601, C. parapsilosis LC641867.1_130-786 and C. auris OL455790.1_1-300). For internal control, the LAMP primer set was designed within the conserved human glucose 6 phosphatase dehydrogenase (G6PD) gene. All LAMP primers, including two outer primers (forward primer F3 and backward primer B3), two inner primers (forward inner primer FIP and backward inner primer BIP) and two loop primers (forward loop primer FLP and backward loop primer BLP), were designed using Primer Explorer software (Version 4; Eiken Chemical Co., Tokyo, Japan). For the multiplex LAMP assay, a dye-labeled artificial nucleic acid + BLP sequence probe and a quencher-labeled displacement probe complementary to the artificial nucleic acid sequence were used. In this study, two types of artificial nucleic acids (35mers and 32mers) were used for multiplexing different fluorescence (Cy5 and Texas Red) quenched by BHQ2 and BHQ1, respectively. A Cy5-labeled 35-artificial oligomer-Ca Pan BLP was designed for Ca Pan BLP probe 1 and a Texas Red-labeled 32-artificial oligomer-internal control BLP was designed for internal control BLP probe 2. The quencher-labeled 35-oligonucleotide (BHQ2) or 30-oligonucleotide (BHQ1) were complementary to the artificial nucleic acid sequences of Ca Pan BLP probe 1 and internal control BLP probe 2, respectively. All primers were assessed for specificity before use in the LAMP assays via a BLAST search of sequences in GenBank (National Center for Biotechnology Information (NCBI), Bethesda, MD). All LAMP primers and probes were synthesized by Macrogen Inc. (Seoul, Korea; Table 1).
Table 1

The LAMP primer sets and qPCR primer sets used in the study.

TargetNameSequence (5′-3′)Length (mer)Reference
Candida Pan (Ca Pan, partial ITS1, 5.8S rRNA gene and partial ITS2)Ca Pan F3AAA ACT TTC AAC GGA T19Present study
Ca Pan B3ACG CTC AAA CAG GCA15
Ca Pan FIPCAA KTC ARA YTA WKT ATC GCA STT CCT CTT GGT TCT CGC ATC G43
Ca Pan BIPCGT GAA TCA TCG AAR YYT TT TTC GCT GCG CTC TTC ATT GGC GCA ATG TGC GT53
Ca Pan FLPACG TAT CGC ATT TCG CTG C19
Ca Pan BLPTTC GCT GCG CTC TTC A16
Ca Pan BLP_CY5 probe1[CY5]-GTC AGT GCA GGC TCC CGT GTT AGG ACG AGG GTA GGT TCG CTG CGC TCT TCA51
Internal control (IC, G6PD)IC G6PD F3TGT CAC CAG CAA CAT CTC GA20Present study
IC G6PD B3TCC TCA GGG AAG CAA ATG AC18
IC G6PD FIPATA GCA GAG AGG CTG CCT ACG GTT TTG ATG TCC CCT GTC CCA45
IC G6PD BIPAAG AAA AGC AGA CGC AGC TTT TTG GGG CTG TTT GCG GAT T43
IC G6PD FLPGGG GTG GCC ATG GAG TGC18
IC G6PD BLPTCC CAA CCT CAA TGC CCT GC20
IC G6PD BLP TEX probe 2[Texas red] –CGG GCC CGT ACA AAG GGA ACA CCC ACA CTC CGT CCC AAC CTC AAT GCC CTG C52
Quencher probe 1 CCT ACC CTC GTC CTA ACA CGG GAG CCT GCA CTG AC-BHQ235
Quencher probe 2 GAG TGT GGG TGT TCC CTT TGT ACG GGC CCG-BHQ130
Candida Pan (Ca Pan) RT-PCRCP PCR FCCT GTTT GAG CGT CRT TT17[27]
CP PCR RTCC GCT TAT TGA TAT18
C. albicans(CA) RT-PCRCA PCR FCTT GGT ATT TTG CAT GTT GCT CTC24[28]
CA PCR RGTC AGA GGC TAT AAC ACA CAG CAG24
CA PCR probe[FAM] - TTT ACC GGG CCA GCA TCG GTT T – BHQ122
C. glabrata(CG) RT-PCRCF PCR FGCG CCC CTT GCC TCT C16[28]
CF PCR RCCC AGG GCT ATA ACA CTC TAC ACC24
CF PCR probe[HEX] – TGG GCT TGG GAC TCT CGC AGC – BHQ121
C. tropicalis(CT) RT-PCRCT PCR FGCG GTA GGA GAA TTG CGT T19[28]
CT PCR RTCA TTA TGC CAA CAT CCT AGG TTT A25
CT PCR probe[CY5] – CGC AGT CCT CAG TCT AGG CTG GCA G – BHQ225
C. krusei(CK) RT-PCRCK PCR FCTCA GAT TTG AAA TCG TGC TTT G23[28]
CK PCR RGGG GCT CTC ACC CTC CTG18
CK PCR probe[TEX] – CAC GAG TTG TAG ATT GCA GGT TGG AGT CTG – BHQ130
C. parapsilosis(CP) RT-PCRCP PCR FGAT CAG ACT TGG TAT TTT GTA TGT TAC TCT C31[28]
CP PCR RCAG AGC CAC ATT TCT TTG CAC21
CP PCR probe[FAM] – CCT CTA CAG TTT ACC GGG CCA GCA TCA – BHQ127
C. auris(CR) RT-PCRCR PCR FCGT GAT GTC TTC TCA CCA ATC T22[29]
CR PCR RTAC CTG ATT TGA GGC GAC AAC21

2.5. The Candida Pan/IC LAMP Assay

The Candida Pan/IC LAMP assay was performed using a Mmiso DNA amplification kit (Mmonitor, Deagu, South Korea). For the multiplex Candida Pan/IC LAMP assay, the reaction mixture was prepared with 12.5 μL of 2× reaction buffer, 1.25 μL of Candida Pan LAMP primer mix (20×), 0.3125 μL of internal control LAMP primer mix (20×), 1.25 μL of 9 μM quencher 1 solution for quenching the Candida Pan LAMP probe, 0.3125 μL of 9 μM quencher 2 solution for quenching the IC LAMP probe and 2 μL of sample DNA (with a final reaction volume of 25 μL). The composition of the Candida Pan LAMP primer mix (20×) included two outer primers at 4 µM (F3 and B3), two inner primers at 32 µM (FIP and BIP), 10 µM loopF primer (FLP), 4 µM loopB primer (BLP) and 6 µM loopB Cy5 probe. The composition of the internal control LAMP primer mix included two outer primers at 4 µM (F3 and B3), two inner primers at 32 µM (FIP and BIP), 10 µM loopF primer (FLP), 4 µM loopB primer (BLP) and 6 µM loopB Texas Red probe. The LAMP assay was run on a CFX 96 Touch Real-Time PCR Detection System (Bio-Rad Laboratories, Hercules, CA, USA) at 58 °C for 40 min. In the LAMP assay, negative controls (human blood DNA and distilled water) were used to set the baseline.

2.6. Real-Time PCR

To evaluate the performance of the Candida Pan/IC LAMP assay, real-time PCR was performed with Candida Pan [27] and Candida species real-time PCR primer sets [28,29] using the iQ Multiplex Powermix (Bio-Rad Laboratories, California, USA) on the CFX96 Touch Real-Time PCR Detection System (Bio-Rad Laboratories). The PCR cycling conditions of Candida Pan real-time PCR primer set were as follows: inactivation at 95 °C for 3 min, 39 cycles of denaturation at 95 °C for 15 s and annealing with fluorescence detection at 62 °C for 25 s. The PCR cycling conditions of the Candida species real-time PCR primer set were as follows: inactivation at 50 °C for 2 min and 95 °C for 10 min, 39 cycles of denaturation at 95 °C for 15 s and annealing with fluorescence detection at 60 °C for 1 min.

2.7. Limit of Detection (LOD) Tests

The LOD of the Candida Pan/IC LAMP was determined using six Candida spp., including C. albicans, C. krusei, C. tropicalis, C. parapsilosis, C. auris and C. glabrata. Candida DNA (1.0 × 107 cells) was serially diluted 10-fold, from 1.0 × 107 cells/μL to 1.0 × 100 cells/μL and used to determine the LOD of the multiplex Candia Pan/IC LAMP assay. In addition, the LOD of the Candida Pan/IC LAMP was tested with serially diluted blood samples spiked with C. albicans (from 107 to 100). The LOD of the Candida Pan/IC LAMP assay was compared with that of the conventional Candida Pan RT-PCR. All tests were repeated three times and determined as the minimum concentration in a 10-fold dilution series, at which three of three replicates were amplified.

2.8. Statistical Analysis

The confidence intervals (CI) for sensitivity and specificity were set at 95%. The sensitivity, specificity and 95% CI for the assays were calculated using a diagnostic test evaluation calculator program (https://www.medcalc.org/calc/diagnostic_test.php, accessed on 21 December 2021).

3. Results

3.1. Optimization of the Chelex-100/Boiling Method for Nucleic Acid Extraction

To optimize the Chelex-100/boiling method, different concentrations of the Chelex 100 Resin solutions (0%, 5% and 10%) were tested using C. albicans real-time PCR and the Candida Pan/IC LAMP assay for C. albicans DNA extracted from the whole blood samples spiked with Candida cells (total cell concentration of 107/mL) (Figure 2A). For DNA extraction with 0%, 5% and 10% Chelex-100 resin solutions, real-time PCR and the Candida Pan/IC LAMP assay showed Ct 28.84/26.16/25.05 and Ct 13.59/13.49/12.75, respectively. Thus, the 10% Chelex-100 resin solution was determined to be the optimum concentration of Chelex-100 for the Chelex-100/boiling method. Next, the performance of the Chelex-100/boiling method was compared with that of the commercial QIAamp UCP Pathogen Mini Kit (Qiagen, Hilden, Germany) against whole blood samples spiked with C. albicans using C. albicans qPCR and Candida Pan/IC LAMP assay (Figure 2B, Table 2). As a result, the detection limits of the C. albicans qPCR primer set were 105/mL and 104/mL in DNA samples extracted using the Chelex-100/boiling and QIAamp UCP Pathogen Mini Kit, respectively. Interestingly, the Candida Pan/IC LAMP assay showed lower detection limits (104/mL and 103/mL) than those obtained by qPCR for DNA extracted using the Chelex-100/boiling and QIAamp UCP Pathogen Mini Kit, respectively. Although the Chelex-100/boiling method showed lower efficiency than the commercial QIAamp UCP Pathogen Mini Kit, Candida Pan/IC LAMP, followed by Chelex-100/boiling DNA extraction, showed similar results to qPCR for DNA samples extracted using the QIAamp UCP Pathogen Mini Kit. These results suggest that the Candida Pan/IC LAMP with Chelex-100/boiling DNA extraction is useful for the rapid diagnosis of candidemia.
Figure 2

Optimization of the Chelex-100/boiling DNA extraction method. (A) Different concentration ratios of Chelex-100 (0%, 5% and 10%). (B) Comparison of detection limit of Candida albicans qPCR and Candida Pan/IC LAMP assay on two kinds of 10-fold serial diluted DNA samples extracted by Chelex-100/boiling and QIAamp UCP Pathogen Mini Kit, respectively.

Table 2

Limit of detection (LOD) tests of Candida species qPCRs and the Candida Pan/IC LAMP assay for two Candida-spiked blood DNA samples extracted by QIAamp UCP Pathogen Mini Kit and Chelex-100/boiling method.

DNA ExtractionMethod PCR AnalysisPrimer SetsTotal Concentration (cells/mL)
107106105104103102101DW *
QIAamp UCPPathogen Mini KitqPCR C. albicans 22.4826.1531.2936.27N/AN/AN/AN/A
MultiplexRT LAMPCy5 (c. pan)15.8016.5318.3522.5425.14N/AN/AN/A
Tex (IC)25.0127.2027.0124.5325.10N/AN/AN/A
Chelex-100/boilingqPCR C. albicans 24.1228.1131.99N/AN/AN/AN/AN/A
MultiplexRT LAMPCy5 (c. pan)16.3617.7821.0423.43N/AN/AN/AN/A
Tex (IC)23.4322.2518.6318.59N/AN/AN/AN/A

* DW, distilled water; N/A, not available.

3.2. Comparison of Detection Limits of the Candida Pan/IC LAMP Assay and Two qPCR (Pan and Candida Species) against Candida Species

To confirm the performance of the Candida Pan/IC LAMP assay, the detection limit of the LAMP assay was compared with that of Candida Pan/Candida species qPCRs for six Candida species, including C. albicans, C. glabrata, C. tropicalis, C. krusei, C. parapsilosis and C. auris. DNA samples from all Candida species were extracted using the QIAamp UCP Pathogen Mini Kit (Table 3). For C. albicans and C. krusei, the Candida Pan/IC LAMP assay showed the lowest detection limit (103) among the three tested assays. The detection limits of Candida Pan qPCR and Candida species qPCR (C. albicans and C. krusei) [28] were 105/106 and 104/105 for C. albicans and C. krusei, respectively. For C. glabrata and C. tropicalis, the Candida Pan/IC LAMP assay and Candida species qPCR (C. glabrata and C. tropicalis) [28] showed the same detection limit (105). The detection limit of Candida Pan qPCR was 106/106 for C. glabrata and C. tropicalis, respectively. qPCR specific for C. parapsilosis [28] and C. auris [29] showed the lowest detection limit (104) among the three tested assays. The detection limits of Candida Pan qPCR were 106 and 107 for C. parapsilosis and C. auris, respectively. The Candida Pan/IC LAMP assay showed the same detection limits (105) for C. parapsilosis and C. auris.
Table 3

Limit of detection (LOD) tests of Pan/Candida species qPCRs and the Candida Pan/IC LAMP assay for 6 Candida spp.

CandidaSpeciesPrimer SetsTotal Concentration (cells/mL)
107106105104103102101DW *
C. albicans qPCRCandida Pan24.1732.7136.66N/AN/AN/AN/AN/A
C. albicans 22.5026.6031.3539.78N/AN/AN/AN/A
LAMPCy5 (c. pan)15.2016.7522.4527.4428.01N/AN/AN/A
Tex (IC)N/AN/AN/AN/AN/AN/AN/AN/A
C. glabrata qPCRCandida Pan22.0231.08N/AN/AN/AN/AN/AN/A
C. glabrata 21.0728.2535.50N/AN/AN/AN/AN/A
LAMPCy5 (c. pan)15.0417.4.23.76N/AN/AN/AN/AN/A
Tex (IC)N/AN/AN/AN/AN/AN/AN/AN/A
C. tropicalis qPCRCandida Pan22.2730.38N/AN/AN/AN/AN/AN/A
C. tropicalis 23.6929.2236.38N/AN/AN/AN/AN/A
LAMPCy5 (c. pan)15.7618.3626.04N/AN/AN/AN/AN/A
Tex (IC)N/AN/AN/AN/AN/AN/AN/AN/A
C. krusei qPCRCandida Pan20.6028.44N/AN/AN/AN/AN/AN/A
C. krusei 22.4628.4238.84N/AN/AN/AN/AN/A
LAMPCy5 (c. pan)17.0019.3125.3434.45N/AN/AN/AN/A
Tex (IC)N/AN/AN/AN/AN/AN/AN/AN/A
C. parapsilosis qPCRCandida Pan20.3824.8534.73N/AN/AN/AN/AN/A
C. parapsilosis 21.3425.8031.4836.97N/AN/AN/AN/A
LAMPCy5 (c. pan)17.8020.3428.92N/AN/AN/AN/AN/A
Tex (IC)N/AN/AN/AN/AN/AN/AN/AN/A
C. auris qPCRCandida Pan35.51N/AN/AN/AN/AN/AN/AN/A
C. auris 21.5925.1430.2136.63N/AN/AN/AN/A
LAMPCy5 (c. pan)15.9717.5421.58N/AN/AN/AN/AN/A
Tex (IC)N/AN/AN/AN/AN/AN/AN/AN/A

* DW, distilled water; N/A, not available.

3.3. Sensitivity and Specificity of the Candida Pan/IC LAMP Assay with Two qPCR Assays against Candida Clinical Sample DNA Extracted by QIAamp UCP Pathogen Mini Kit and Chelex-100/Boiling Method

To confirm the clinical performance of the Chelex-100/boiling DNA extraction method and the Candida Pan/IC LAMP assay, DNA extraction was performed using two different methods, the QIAamp UCP Pathogen Mini Kit and Chelex-100/boiling method; the sensitivities of the Candida Pan/IC LAMP assay were compared with those of the Candida Pan qPCR and specific Candida species qPCR for 36 clinical samples from patients infected with C. albicans (9), C. glabrata (9), C. tropicalis (9) and C. parapsilosis (9) (Table 4). Within 40 min, the Candida Pan/IC LAMP assay showed 100% sensitivity for two kinds of Candida clinical sample DNA extracted by QIAamp UCP Pathogen Mini Kit and Chelex-100/boiling method, respectively. Conventional Candida Pan and Candida species qPCR showed the same sensitivity (86.11%) for 36 Candida clinical sample DNA extracted using the QIAamp UCP Pathogen Mini Kit. However, for Candida clinical sample DNA extracted by the Chelex-100/boiling method, the Candida Pan qPCR and Candida species qPCR showed 22% and 44% sensitivity, respectively. For 100 negative clinical samples (non-infected), the specificity of the three assays was 100% (Table 4).
Table 4

Comparison of sensitivity and specificity of multiplex Candida Pan/IC LAMP assay with reference pan and mono Candida species RT-PCR against Candida and non-infectious clinical samples.

Clinical SamplesQIAamp UCP Pathogen Mini KitBoiling and Filtering Method
qPCRMultiplexLAMPqPCRMultiplexLAMP
CandidapanCandidaSpeciesCy5(C. pan)Tex(IC)CandidapanCandidaSpeciesCy5(C. pan)Tex(IC)
Candida Spp.(n = 36)P/N31/531/536/027/98/2816/2036/028/8
Sensitivity (95% CI-86.11%[70.50–95.33]86.11%[70.50–95.33]100%[90.26–100.00]75.00%[57.80–87.88]22.22%[10.12–39.15]44.44%[27.94–61.90]100%[90.26–100.00]77.77%[60.85–89.88]
C. albicans(n = 9)P/N8/18/19/06/31/85/49/09/0
Sensitivity88.89%88.89%100%66.67%11.11%55.56%100%100%
C. glabrata(n = 9)P/N8/18/19/07/22/73/69/06/3
Sensitivity88.89%88.89%100%77.78%22.22%33.33%100%66.67%
C. tropicalis(n = 9)P/N6/36/39/08/12/73/69/06/3
Sensitivity66.67%66.67%100%88.89%22.22%33.33%100%66.67%
C. parapsilosis(n = 9)P/N9/09/09/06/33/65/49/07/2
Sensitivity100%100%100%66.67%33.33%55.56%100%77.78%
Non-infection(n = 100)P/N0/1000/1000/100100/00/1000/1000/100100/0
Sensitivity(95% CI)N/AN/AN/A100%[96.38–100.00]N/AN/AN/A100%[96.38–100.00]
Specificity(95% CI)100%[96.38–100.00]100%[96.38–100.00]100%[96.38–100.00]N/A 100%[96.38–100.00]100%[96.38–100.00]100%[96.38–100.00]N/A

3.4. Cross-Reactivity Test

To confirm the possibility of cross-reactivity of the Candida Pan/IC LAMP assay, the Candida Pan/IC LAMP assay was tested with other bacterial samples, including Escherichia coli, Enterococcus faecium, Klebsiella spp., Staphylococcus aureus and Staphylococcus epidermidis samples (Table 5). The Candida Pan/IC LAMP assay showed no cross-reactivity with other fungal infection samples, suggesting that the LAMP assay can specifically detect Candida species.
Table 5

Cross-reactivity of the Candida Pan/IC LAMP against other bacterial infection samples.

Candida Pan/IC LAMP
Cy5(Candida pan)Texas Red(Internal Control)
SamplesCtRFU *CtRFU *
Escherichia coli N/A19.1N/A23.3
Enterococcus faecium N/A31.1N/A22.6
Klebsiella spp. N/A20.0N/A28.3
Staphylococcus aureus N/A44.0N/A39.0
Staphylococcus epidermidis N/A57.7N/A42.6
Human whole blood DNAN/A−17.929.684186
Distilled waterN/A1.76N/A−0.889

* RFU, relative fluorescence units.

4. Discussion

Invasive candidiasis (IC) is a serious cause of morbidity and mortality [26,30]. In the hospital, Candida spp. account for 8–9% of all nosocomial bloodstream infections and the risk is higher in intensive care unit (ICU) patients and cancer patients [31,32]. Candidemia has an associated mortality rate of up to 25% and a fast diagnosis followed by early adequate antifungal therapy can significantly reduce premature mortality [33,34]. In this study, we developed a Chelex-100/boiling DNA extraction method (within 20 min) and Candida Pan/IC LAMP assay (40 min) for the rapid diagnosis of candidemia. The Chelex-100/boiling method showed a slightly lower efficiency than the commercial QIAamp UCP Pathogen Mini Kit; however, the Chelex-100/boiling DNA extraction method (within 20 min) extracted DNA 3–4 times faster than commercial QIAamp UCP Pathogen Mini kits (60–90 min). In addition, Chelex-100/boiling DNA extraction followed by the Candida Pan/IC LAMP assay showed similar results to the reference qPCR for DNA samples extracted using the QIAamp UCP Pathogen Mini Kit. Interestingly, the LAMP assay showed superior performance (100% sensitivity) compared with the conventional reference qPCRs (86.11% sensitivity) against 36 numbers of Candida clinical DNA samples extracted by both the Chelex-100/boiling method and commercial QIAamp UCP Pathogen Mini Kit. In addition, the Candida Pan/IC LAMP assay showed 100% specificity against 100 non-infected clinical samples. As fungal cell walls are difficult to break with conventional extraction methods, obtaining DNA from fungi is more difficult than extracting nucleic acids from bacteria or mammalian cells [35]. For this reason, nucleic acid extraction methods to detect fungus-infected blood generally require additional procedures (e.g., mechanical, enzymatic and/or chemical methods) to disrupt the fungal cell wall [36]. Therefore, these nucleic acid extraction methods consist of a complicated procedure and it takes a long time to extract the nucleic acid [37]. Nucleic acid extraction using cheliex-100 is known as a fast and easy method for nucleic acid extraction from various samples, such as forensics, blood, parasites, virus and bacteria [38,39]. The first protocol for DNA extraction using Chelex-100 was developed by Walsh et al. [40]. Chelex-100 has been mainly used in forensics in conjunction with thermal denaturation to extract nucleic acids from trace cells or blood. Boiling the sample not only releases the DNA from the cells into the solution, but also promotes the binding of Chelex-100 to the magnesium ion, which is a cofactor of deoxyribonuclease. DNA degradation is prevented because the binding of Chelex-100 to magnesium ions results in the inactivation of deoxyribonuclease. In this study, after removing red blood cells from the blood using a red blood cell lysis solution, cells that were not lysed were separated by centrifugation. After heating the separated cells using a Chelex-100 solution, the nucleic acids were separated using a 3 µm filter. Additionally, a bead beating step was added to the Chelex-100/boiling method to further increase Candida cell destruction. However, there was no significant difference in the ct value of the LAMP assay, so this step was excluded (Figure S1). The Candida Pan LAMP primer set was designed using Primer Explorer software by multialigning conserved regions of partial ITS1, 5.8S rRNA gene and partial ITS2 genes of C. albicans, C. glabrata, C. tropicalis, C. krusei, C. parapsilosis and C. auris. However, it was not possible to design the BLP primer because B1 and B2 of the designed primer set were too close. The LOOP primer explosively amplified the LAMP reaction. In this study, for the development of the rapid Candida Pan LAMP primer set, an artificial sequence (TTCGCTGCGCTCTTCA) capable of reacting with the LOOP primer was added between B1 and B2 to design the BIP primer. Indeed, the Candida Pan LAMP primer set with the BLP primer of the same artificial sequence showed an increased reaction rate compared to the primer set without BLP (Figure S2). In addition, we used this BLP + Cy5-labeled 35-artificial oligomer as a Candida Pan probe; since the BLP of this artificial sequence is not involved in the Candida target gene, it only responds to the operation of BIP without a non-specific reaction. Our study has a limitation. The Candida Pan/IC LAMP assay was performed with a relatively small sample size of positive Candida clinical samples (36), which resulted in widened confidence intervals for our estimates of diagnostic accuracy. However, considering that the Candida Pan/IC LAMP assay showed higher sensitivity than the two conventional PCR and 100% specificity to the negative samples, the Candida Pan/IC LAMP assay is sufficiently competitive for commercial development.

5. Conclusions

In this study, we developed a fast candidemia detection system including Chelex-100/boiling DNA extraction and the Candida Pan/IC LAMP assay, which is capable of diagnosing Candida species in blood within 1 h. In a sensitivity test with Candida clinical samples, the Candida Pan/IC LAMP assay showed superior performance to the two reference qPCRs. Thus, Chelex-100/boiling DNA extraction followed by the Candida Pan/IC LAMP assay could serve as a useful fast molecular diagnostic test for Candida spp. in blood.
  40 in total

1.  Loop-mediated isothermal amplification of DNA.

Authors:  T Notomi; H Okayama; H Masubuchi; T Yonekawa; K Watanabe; N Amino; T Hase
Journal:  Nucleic Acids Res       Date:  2000-06-15       Impact factor: 16.971

2.  Comparison of six DNA extraction methods for recovery of fungal DNA as assessed by quantitative PCR.

Authors:  David N Fredricks; Caitlin Smith; Amalia Meier
Journal:  J Clin Microbiol       Date:  2005-10       Impact factor: 5.948

3.  Performance of Candida real-time polymerase chain reaction, β-D-glucan assay, and blood cultures in the diagnosis of invasive candidiasis.

Authors:  M Hong Nguyen; Mark C Wissel; Ryan K Shields; Martin A Salomoni; Binghua Hao; Ellen G Press; Ryan M Shields; Shaoji Cheng; Dimitra Mitsani; Aniket Vadnerkar; Fernanda P Silveira; Steven B Kleiboeker; Cornelius J Clancy
Journal:  Clin Infect Dis       Date:  2012-03-19       Impact factor: 9.079

4.  Diagnosis of Candida albicans: conventional diagnostic methods compared to the loop-mediated isothermal amplification (LAMP) assay.

Authors:  Shirzad Fallahi; Milad Babaei; Ali Rostami; Hadi Mirahmadi; Zahra Arab-Mazar; Asghar Sepahvand
Journal:  Arch Microbiol       Date:  2019-10-22       Impact factor: 2.552

Review 5.  Comparison between different methods of DNA isolation from dried blood spots for determination of malaria to determine specificity and cost effectiveness.

Authors:  Barsa Baisalini Panda; Arup Shankar Meher; Rupenangshu Kumar Hazra
Journal:  J Parasit Dis       Date:  2019-06-24

Review 6.  PCR diagnosis of invasive candidiasis: systematic review and meta-analysis.

Authors:  Tomer Avni; Leonard Leibovici; Mical Paul
Journal:  J Clin Microbiol       Date:  2010-11-24       Impact factor: 5.948

7.  Rapid detection of Salmonella based on loop-mediated isothermal amplification.

Authors:  Hongling Ou; Yan Wang; Jiayi Gao; Jing Bai; Qiaoyun Zhang; Lei Shi; Xinru Wang; Chengbin Wang
Journal:  Ann Palliat Med       Date:  2021-06

8.  Risk factors for albicans and non-albicans candidemia in the intensive care unit.

Authors:  Jennifer K Chow; Yoav Golan; Robin Ruthazer; Adolf W Karchmer; Yehuda Carmeli; Deborah A Lichtenberg; Varun Chawla; Janet A Young; Susan Hadley
Journal:  Crit Care Med       Date:  2008-07       Impact factor: 7.598

Review 9.  The use of mannan antigen and anti-mannan antibodies in the diagnosis of invasive candidiasis: recommendations from the Third European Conference on Infections in Leukemia.

Authors:  Malgorzata Mikulska; Thierry Calandra; Maurizio Sanguinetti; Daniel Poulain; Claudio Viscoli
Journal:  Crit Care       Date:  2010-12-08       Impact factor: 9.097

10.  Evaluation of 11 DNA Automated Extraction Protocols for the Detection of the 5 Mains Candida Species from Artificially Spiked Blood.

Authors:  Estelle Menu; Jordi Landier; Elsa Prudent; Stéphane Ranque; Coralie L'Ollivier
Journal:  J Fungi (Basel)       Date:  2021-03-19
View more