PURPOSE: To investigate the possible effect of an implantable collamer lens (ICL) on ocular biometrics and intraocular lens (IOL) power calculation.Entities:
Year: 2021 PMID: 34970453 PMCID: PMC8714341 DOI: 10.1155/2021/8988479
Source DB: PubMed Journal: J Ophthalmol ISSN: 2090-004X Impact factor: 1.909
The biometric parameters measured by IOLMaster.
| Parameters | Preoperative | 3-month postoperative |
|
|---|---|---|---|
| ACD (mm) | 3.69 ± 0.23 | 3.49 ± 0.38 | < 0.001 |
| LT (mm) | 3.59 ± 0.22 | 3.75 ± 0.35 | < 0.001 |
| AL (mm) | 26.21 ± 1.23 | 26.20 ± 1.22 | 0.233 |
| WTW (mm) | 11.86 ± 0.38 | 11.88 ± 0.40 | 0.100 |
| K1 (D) | 43.27 ± 1.24 | 43.31 ± 0.14 | 0.117 |
| K2 (D) | 44.71 ± 1.62 | 44.54 ± 1.66 | < 0.001 |
| CCT ( | 516.09 ± 30.51 | 515.69 ± 31.14 | 0.648 |
| IOL (D) | 12.54 ± 2.69 | 12.61 ± 2.62 | 0.013 |
ACD, anterior chamber depth; LT, lens thickness; AL, axial length; WTW, white to white; K1, corneal flat axis curvature; K2, corneal steep axis curvature; CCT, central corneal thickness; IOL, intraocular lens. P < 0.05.
Figure 1(a) The distribution of preoperative and postoperative anterior chamber depth deviations (∆ = pre minus post). (b) The distribution of preoperative and postoperative lens thickness deviations (∆ = pre minus post).
Figure 2The IOLMaster images of a 28 year-old woman. The anterior surface of the right eye's lens (above) was accurately identified by IOLMaster, while in the left eye (below), the anterior surface of ICL was mistaken as the anterior surface of lens.
The biometric parameters measured by IOLMaster excluding 11 misjudged eyes.
| Parameters | Preoperative | 3-month postoperative |
|
|---|---|---|---|
| ACD (mm) | 3.71 ± 0.22 | 3.61 ± 0.22 | < 0.001 |
| LT (mm) | 3.58 ± 0.23 | 3.64 ± 0.22 | < 0.001 |
| AL (mm) | 26.27 ± 1.21 | 26.26 ± 1.20 | 0.310 |
| WTW (mm) | 11.86 ± 0.37 | 11.89 ± 0.40 | 0.038 |
| K1 (D) | 43.20 ± 1.14 | 43.25 ± 1.17 | 0.082 |
| K2 (D) | 44.64 ± 1.61 | 44.46 ± 1.64 | < 0.001 |
| CCT ( | 518.93 ± 28.96 | 517.91 ± 30.26 | 0.271 |
| IOL (D) | 12.46 ± 2.68 | 12.55 ± 2.60 | 0.004 |
ACD, anterior chamber depth; LT, lens thickness; AL, axial length; WTW, white to white; K1, corneal flat axis curvature; K2, corneal steep axis curvature; CCT, central corneal thickness; IOL, intraocular lens. P < 0.05.
Regression analysis of preoperative and postoperative IOL power difference.
| Values | ACD deviation | LT deviation | AL deviation | K1 deviation | K2 deviation | WTW deviation |
|---|---|---|---|---|---|---|
| Standard coefficient beta | 0.287 | 0.108 | −0.504 | −0.561 | −0.652 | 0.066 |
|
| 0.002 | 0.218 | < 0.001 | < 0.001 | < 0.001 | 0.015 |
ACD, anterior chamber depth; LT, lens thickness; AL, axial length; WTW, white to white; K1, corneal flat axis curvature; K2, corneal steep axis curvature. P < 0.05.
Sirius measurements of anterior segment before and after ICL implantation.
| Parameters | Preoperative | 3-month postoperative |
|
|---|---|---|---|
| ACD (mm) | 3.26 ± 0.23 | 2.32 ± 0.28 | < 0.001 |
| CCT ( | 520.58 ± 31.93 | 520.47 ± 32.32 | 0.872 |
| K1 (D) | 43.18 ± 1.26 | 43.28 ± 1.31 | 0.057 |
| K2 (D) | 44.50 ± 1.70 | 44.40 ± 1.76 | 0.072 |
| IOL (D) | 11.49 ± 2.93 | 11.43 ± 2.85 | 0.653 |
ACD, anterior chamber depth; CCT, central corneal thickness; K1, K on flat axis; K2, K on steep axis. P < 0.05.