[Effects of posterior chamber phakic intraocular lens on axial length measurements].

Ayong Yu, Qinmei Wang, Shuangqian Zhu, Anquan Xue, Yanfeng Su, Rui Pan.   

Abstract

OBJECTIVE: To study the effects of posterior chamber phakic intraocular lens (IOL) on axial length (AL) measurements and whether a correction factor for later IOL power calculation before cataract surgery is necessary.
METHODS: Twenty-two high myopic patients (39 eyes) with mean spherical equivalent -19.00 +/- 4.03 D received phakic refractive lens (PRL) implantation, and 24 patients (40 eyes) with mean spherical equivalent (-13.10 +/- 2.53) D received implantable Collamer lens (ICL) implantation. Pre-and postoperative AL measurements with the IOLMaster were compared. Results The pre-and postoperative ALs were (30.87 +/- 2.18) mm and (30.94 +/- 2.16) mm, respectively (t = -2.635, P = 0.012) in the PRL group, and (28.35 +/- 1.20) mm and (28.45 +/- 1.22) mm, respectively ( t = -3.97, P < 0.001) in the ICL group. Linear regression analysis showed: in the PRL group, preoperative AL = 1.01 x postoperative AL-0.24 (R2 = 0.99, P < 0.001), or postoperative AL--preoperative AL = 0.24-0.0055 x postoperative AL; in the ICL group, preoperative AL = 0.96 x postoperative AL +1.17 (R2 = 0.99, P < 0.001), or postoperative AL--preoperative AL = 0.0154 x postoperative AL -0.3763. The mean differences between pre-and postoperative ALs in the PRL and, ICL groups were 0.075 mm and 0.054 mm, respectively.
CONCLUSIONS: Based on the SRK formula, a < 0.1 mm AL difference would result in a <0.25 D change in the calculated IOL power, possibly a < 0.20 D change in manifest refraction. Furthermore, for PRL, the longer the AL, the smaller the difference between pre-and postoperative ALs. These differences are so small that it is unnecessary to correct IOL power calculation.

Entities:  

Mesh:

Year:  2015        PMID: 26268644

Source DB:  PubMed          Journal:  Zhonghua Yan Ke Za Zhi        ISSN: 0412-4081


  2 in total

1.  Full-vision maintenance in extra-high myopia from implantable collamer lens to trifocal intraocular lens implantation.

Authors:  Shu-Yang Chen; Chen Xie; Yang Wang; Ye Shen
Journal:  Int J Ophthalmol       Date:  2018-07-18       Impact factor: 1.779

2.  Effect of Implantable Collamer Lens on Anterior Segment Measurement and Intraocular Lens Power Calculation Based on IOLMaster 700 and Sirius.

Authors:  Xiaoyong Chen; Di Zhang; Ziyuan Liu; Yinan Liu; Hongyuan Cai; Qianru Wu; Yu Zhang
Journal:  J Ophthalmol       Date:  2021-12-21       Impact factor: 1.909

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.