Lenstar® LS 900 vs Pentacam®-AXL: Comparative study of ocular biometric measurements and intraocular lens power calculation.

Joana Maria Mendes Pereira1, Arminda Neves1, Pedro Alfaiate1, Mónica Santos1, Henrique Aragão1, João C Sousa2.   

Abstract

PURPOSE: : Comparison of biometric measurements and calculation of intraocular lens with a new biometer (Pentacam®-AXL, Oculus, Germany) and a reference biometer (Lenstar LS 900®, Haag-Streit AG, Switzerland), in order to assess the agreement between these two devices.
SETTING: : Centro Hospitalar de Leiria, Portugal.
MATERIALS AND METHODS: : Prospective, institutional study, in which measurements of axial length, anterior chamber depth from the corneal epithelium and endothelium to the anterior surface of the lens (anterior chamber depth ext and anterior chamber depth int), central corneal thickness and keratometry readings of the flattest and steepest meridians (K1 and K2) were obtained with the two systems. Intraocular lens calculation was also performed, using the Haigis, SRK/T and HofferQ formulas.
RESULTS: : The study sample included 136 eyes of 79 patients. Of these, 42 were women and 37 were men. Statistically significant differences were found (p < 0.05, paired T test) in K1, K2 and central corneal thickness between the 2 biometers. Intraocular lens calculation with the Lenstar® and the Pentacam® with Haigis, SRK/T and HofferQ formulas showed statistically significant differences (p < 0.05 Paired T test).
CONCLUSION: : Axial length measurements obtained with the Pentacam® and Lenstar® appear to be interchangeable, while measurements of anterior chamber depth, K1 and K2, and central corneal thickness do not appear to be interchangeable between different devices. Statistically significant differences were found in the calculation of intraocular lenses in all formulas used.

Entities:  

Keywords:  Axial length; Lenstar Pentacam; biometry

Mesh:

Year:  2018        PMID: 29788776     DOI: 10.1177/1120672118771844

Source DB:  PubMed          Journal:  Eur J Ophthalmol        ISSN: 1120-6721            Impact factor:   2.597


  4 in total

Review 1.  A review of imaging modalities for detecting early keratoconus.

Authors:  Xuemin Zhang; Saleha Z Munir; Syed A Sami Karim; Wuqaas M Munir
Journal:  Eye (Lond)       Date:  2020-07-16       Impact factor: 3.775

2.  Lenstar LS 900 versus Pentacam-AXL: analysis of refractive outcomes and predicted refraction.

Authors:  Henrique Aragão Arruda; Joana M Pereira; Arminda Neves; Maria João Vieira; Joana Martins; João C Sousa
Journal:  Sci Rep       Date:  2021-01-14       Impact factor: 4.379

3.  Effect of Implantable Collamer Lens on Anterior Segment Measurement and Intraocular Lens Power Calculation Based on IOLMaster 700 and Sirius.

Authors:  Xiaoyong Chen; Di Zhang; Ziyuan Liu; Yinan Liu; Hongyuan Cai; Qianru Wu; Yu Zhang
Journal:  J Ophthalmol       Date:  2021-12-21       Impact factor: 1.909

4.  Comparison of Biometry Measurements Using Standard Partial Coherence Interferometry versus New Scheimpflug Tomography with Integrated Axial Length Capability.

Authors:  Jorge Selem Haddad; Eliza Barnwell; Karolinne Maia Rocha; Renato Ambrosio; George O Waring Iv
Journal:  Clin Ophthalmol       Date:  2020-02-04