Long-term In Vivo Stability of Posterior Chamber Phakic Intraocular Lens: Properties and Light Transmission Characteristics of Explants.

Tomoaki Nakamura1, Naoki Isogai2, Takashi Kojima3, Yoko Yoshida2, Yasuo Sugiyama2, Yoshiki Tanaka4, Kazuo Ichikawa4.   

Abstract

PURPOSE: To evaluate the in vivo durability of the surface and optical properties of the implantable Collamer lens (ICL).
DESIGN: Retrospective case series.
METHODS: We included patients who developed cataracts after having undergone ICL implantation from March 2003 to May 2014 and underwent ICL explantation followed by cataract surgery from March 2017 to December 2019 at the Nagoya Eye Clinic. ICL explants were submitted to Chukyo Medical Co, Ltd (Nagoya City, Japan) for laboratory analysis using ultraviolet-visible light spectroscopy, light microscopy (LM), and scanning electron microscopy. Patients' demographic and clinical data were collected and reviewed.
RESULTS: Thirteen eyes from 10 patients were studied. The average age at ICL explantation was 50.5 ± 8.5 years (range, 34.5-66.3 years). The average length of ICL stay in the eye (from implantation to explantation) was 10.5 ± 2.7 years (range, 4.4-13.7 years). No opacification or coloring of the ICL explants was observed by LM. The ICL explants showed almost the same light transmittance as that of unused ICLs. Scanning electron microscopy revealed no irregularities at the surface of the center and periphery of the optic and haptic footplate. The positioning holes did not show any deposition.
CONCLUSION: The ICLs remained in-eye for >10 years without any deterioration in the surface and optical properties of the ICL, despite their contact with the ciliary body and iris tissues and the continuous interaction with the aqueous humor components. The present study shows long-term in vivo stability of the ICL.
Copyright © 2020 Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2020        PMID: 32622670     DOI: 10.1016/j.ajo.2020.06.025

Source DB:  PubMed          Journal:  Am J Ophthalmol        ISSN: 0002-9394            Impact factor:   5.258


  3 in total

1.  Benefits of using corneal topography to choose subjective refraction technique in keratoconus (RE-CON): a prospective comparative crossover clinical study.

Authors:  Margaux Metzger; Valentin Navel; Jean-Vincent Barrière; Fabrice Kwiatkowski; Jérémy Hébraud; Aurélien Mulliez; Laurence Béral; Frédéric Chiambaretta; Frédéric Dutheil
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2021-08-20       Impact factor: 3.117

2.  Assessment of biosafety and implantation feasibility of novel phakic refractive lens.

Authors:  Shaohua Zhang; Chang Huang; Huamao Miao; Junyao Wu; Chao Xing; Zhaoxing Dai; Jianguo Sun
Journal:  Int Ophthalmol       Date:  2022-05-13       Impact factor: 2.029

3.  Effect of Implantable Collamer Lens on Anterior Segment Measurement and Intraocular Lens Power Calculation Based on IOLMaster 700 and Sirius.

Authors:  Xiaoyong Chen; Di Zhang; Ziyuan Liu; Yinan Liu; Hongyuan Cai; Qianru Wu; Yu Zhang
Journal:  J Ophthalmol       Date:  2021-12-21       Impact factor: 1.909

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.