Ahmed Jibril Abdi1,2, Bo Mussmann1,3, Alistair Mackenzie4, Oke Gerke1,5, Gitte Maria Jørgensen3, Thor Eriksen Bechsgaard3, Janni Jensen1,3, Lone Brunshøj Olsen3, Poul Erik Andersen1.
Abstract
The purpose of this study was to assess the image quality of the low dose 2D/3D slot scanner (LDSS) imaging system compared to conventional digital radiography (DR) imaging systems. Visual image quality was assessed using the visual grading analysis (VGA) method. This method is a subjective approach that uses a human observer to evaluate and optimise radiographic images for different imaging technologies. METHODS AND MATERIALS: ten posterior-anterior (PA) and ten lateral (LAT) images of a chest anthropomorphic phantoms and a knee phantom were acquired by an LDSS imaging system and two conventional DR imaging systems. The images were shown in random order to three (chest) radiologists and three experienced (knee) radiographers, who scored the images against a number of criteria. Inter- and intraobserver agreement was assessed using Fleiss' kappa and weighted kappa.Entities:
Keywords: DR imaging systems; chest examination; image quality assessment; knee examination; slot scanner imaging system; visual grading analysis
Year: 2021 PMID: 34679630 PMCID: PMC8534907 DOI: 10.3390/diagnostics11101932
Source DB: PubMed Journal: Diagnostics (Basel) ISSN: 2075-4418
Figure 1The frontal view of (a) photographic image of the chest anthropomorphic phantom (b) photographic image of the knee phantom (c) radiographic image of the chest anthropomorphic phantom and (d) radiographic images of the knee phantom.
The list of study protocols and radiation dose level settings used in both patient dose and image quality comparisons of the LDSS and the DR systems.
| Systems | Projections | Dose Level | Protocol |
|---|---|---|---|
| DR system 1 | PA/LAT | Standard (default) | Chest (thorax) |
| DR system 2 | PA/LAT | Standard (default) | |
| LDSS | PA/LAT | Medium dose (speed 6 sec.) | |
| DR system 1 | PA/LAT | Standard (default) | Extremity (knee) |
| DR system 2 | PA/LAT | Standard (default) | |
| LDSS | PA/LAT | High dose (Speed 8 sec.) |
Exposure settings in chest and knee examinations for all imaging systems.
| Imaging Systems | Tube Voltage [kV] | SID (cm) | Tube Current (mA) | Tube Load (mAs) | Exposure Mode | AF Al/Cu (mm) | Examination Protocols |
|---|---|---|---|---|---|---|---|
| LDSS speed 6 | 90 | 130 | 280 | n/a | Manual | 0/0.1 | Chest |
| DR system 1 | 133 | 250 | n/a | 1.6 | AEC | 1/0.2 | |
| DR system 2 | 145 | 300 | n/a | 1.8 | AEC | 0/0.2 | |
| LDSS speed 8 | 68 | 130 | 400 | n/a | Manual | 0/0 | Knee |
| DR system 1 | 57 | 110 | n/a | 8.5 | Manual | 0/0 | |
| DR system 2 | 63 | 115 | n/a | 6.3 | Manual | 0/0 |
n/a = not applicable, AEC = automatic exposure control, SID = source to image distance, AF = additional filtration, Al = aluminium filtration and Cu = copper.
VGA-based scale for scoring image quality.
| Scale | Image Scoring Scale | Description |
|---|---|---|
| 1 | Poor image quality: | Image not usable, loss of information |
| 2 | Restricted image quality: | Relevant limitations for clinical use, clear loss of information |
| 3 | Sufficient image quality: | Moderate limitations for clinical use, no substantial loss of information |
| 4 | Good image quality: | Minimal limitations for clinical use |
| 5 | Excellent image quality: | No limitations for clinical use |
Image quality criteria for the chest examination protocols in both PA and LAT projections.
| Criteria | Technical Image Quality Criteria | Projections | |
|---|---|---|---|
| 1 | Visualisation of: | Vascular pattern of lungs and peripheral vessels | PA |
| 2 | Trachea and proximal bronchi | ||
| 3 | Borders of the heart and the aorta | ||
| 4 | Diaphragm and lateral costophrenic angles | ||
| 5 | Spine through the heart shadow | ||
| 1 | Visualisation of: | Thoracic spine | LAT |
| 2 | Trachea | ||
| 3 | Costophrenic angles | ||
| 4 | Sternum | ||
| 5 | Diaphragm | ||
| 6 | Mediastinum | ||
| 7 | Posterior border of the heart and aorta | ||
Image quality criteria for the knee examination protocol in both PA and LAT projections.
| Criteria | Image Quality Criteria Definitions | Technical Image Quality | Projection |
|---|---|---|---|
| 1 | Homogeneity in soft tissue, lateral to the femoral condyle | Noise | knee PA |
| 2 | Sharpness of trabeculae in the medial femoral condyle | Spatial resolution | |
| 3 | Sharpness of the demarcation between substantia spongiosa and substantia compacta in the femur above patella | Contrast small structure | |
| 4 | Visualization of patella through the femur | Low-contrast resolution and noise | |
| 5 | Visualization of the lateral intra-articular femoral condyle | Contrast large structure | |
| 1 | Homogeneity in soft tissue posterior to the knee joint | Noise | knee LAT |
| 2 | Visualization of head of fibulae behind the tibia | Low-contrast resolution and noise | |
| 3 | Visualization of patella | Contrast large structure | |
| 4 | Sharpness of trabeculae in the tibial metaphysis | Spatial resolution | |
| 5 | Sharpness of the demarcation between substantia spongiosa and substantia compacta in the anterior femur above patella | Contrast small structure |
Fleiss’ kappa assessing interobserver agreement with an asymptotic 95% confidence interval (CI) for each criterion in chest PA/LAT protocols.
| Criteria | 1 | 2 | 3 | 4 | 5 | 6 | 7 | Projections |
|---|---|---|---|---|---|---|---|---|
|
|
|
|
|
|
| n/a | chest PA | |
|
| 0.39–0.80 | 0.51–0.85 | 0.52–0.93 | 0.03–0.35 | 0.61–0.93 | |||
|
| <0.001 | <0.001 | <0.001 | 0.02 | <0.001 | |||
|
|
|
|
|
|
|
|
| chest LAT |
|
| 0.49–0.61 | 0.19–0.52 | 0.41–0.72 | 0.30–0.63 | 0.26–0.61 | 0.45–81 | 0.13–0.41 | |
|
| <0.001 | <0.001 | <0.001 | <0.001 | <0.001 | <0.001 | <0.001 | |
n/a = not applicable.
Fleiss’ kappa assessing interobserver agreement with an asymptotic 95% CI for each criterion in knee PA/LAT protocols.
| Criteria | 1 | 2 | 3 | 4 | 5 | Projections |
|---|---|---|---|---|---|---|
|
|
|
|
|
|
| knee PA |
|
| 0.26–0.64 | 0.09–0.49 | 0.13–0.48 | 0.19–0.59 | 0.10–0.36 | |
|
| <0.001 | 0.004 | <0.001 | <0.001 | <0.001 | |
|
|
|
|
|
|
| knee LAT |
|
| 0.06–0.42 | 0.03–0.33 | 0.01–0.33 | 0.05–0.29 | 0.07–0.44 | |
|
| 0.009 | 0.016 | 0.002 | 0.007 | 0.007 |
Weighted kappa assessing intraobserver agreement for each image quality criterion in the chest PA/LAT protocol.
| Criteria | 1 | 2 | 3 | 4 | 5 | 6 | 7 | Projections |
|---|---|---|---|---|---|---|---|---|
|
|
|
|
|
|
| n/a | chest PA | |
|
| 0.07–0.44 | −0.11–0.42 | −0.09–0.64 | −0.06–0.61 | 0.49–0.78 | |||
|
| 0.02 | 0.01 | <0.001 | <0.001 | <0.001 | |||
|
|
|
|
|
|
|
|
| chest LAT |
|
| −0.22–0.48 | 0.11–0.51 | −0.15–0.70 | 0.05–0.46 | −0.02–0.36 | −0.10–0.29 | −0.09–0.60 | |
|
| 0.10 | <0.001 | 0.04 | 0.04 | 0.05 | 0.05 | 0.02 | |
n/a = not applicable.
Weighted kappa assessing intraobserver agreement for each image quality criterion in the knee PA/LAT protocol.
| Criteria | 1 | 2 | 3 | 4 | 5 | Projections |
|---|---|---|---|---|---|---|
|
|
|
|
|
|
| knee PA |
|
| 0.16–0.60 | −0.13–0.35 | −0.23–0.64 | −0.01–0.42 | n/a | |
|
| <0.001 | 0.41 | 0.07 | 0.04 | <0.001 | |
|
|
|
|
|
|
| knee LAT |
|
| 0.01–0.54 | −0.05–0.44 | −0.12–0.68 | −0.10–0.53 | −0.07–0.75 | |
|
| 0.03 | 0.01 | <0.001 | 0.02 | 0.01 |
n/a = not applicable.
Means of VGA value for all imaging systems for the chest PA/LAT protocol.
| Systems | Mean | 95% CI | Projections |
|---|---|---|---|
|
| 4.73 | 4.64–4.83 | chest PA |
|
| 3.90 | 3.80–3.99 | |
|
| 3.69 | 3.59–3.78 | |
|
| 4.33 | 4.24–4.42 | chest LAT |
|
| 3.86 | 3.37–3.95 | |
|
| 3.40 | 3.31–3.49 |
Comparison of the mean VGA across the imaging systems in chest PA and LAT protocol.
| System Comparison | Mean Difference | Std. Error | Sig. b | 95% CI | Projections |
|---|---|---|---|---|---|
| LDSS vs. DR system 1 | 0.83 | 0.07 | <0.001 | 0.67–0.99 | chest PA |
| LDSS vs. DR system 2 | 1.05 | 0.07 | <0.001 | 0.88–1.20 | |
| DR system 1 vs. DR system 2 | 0.21 | 0.07 | 0.006 | 0.05–0.38 | |
| LDSS vs. DR system 1 | 0.47 | 0.07 | <0.001 | 0.32–0.63 | chest LAT |
| LDSS vs. DR system 2 | 0.94 | 0.07 | <0.001 | 0.78–1.09 | |
| DR system 1 vs. DR system 2 | 0.46 | 0.07 | <0.001 | 0.31–0.62 |
Based on estimated marginal means. b Adjustment for multiple comparisons: Bonferroni.
Figure 2VGA estimated marginal mean comparison across the imaging systems and image quality criteria for (a) the chest PA projection and (b) the chest LAT projection.
Means of mean VGA values, upper and lower bound of 95% CI for all imaging systems in the knee PA/LAT protocol.
| Systems | Mean | 95% CI | Projections |
|---|---|---|---|
|
| 3.86 | 3.78–3.94 | knee PA |
|
| 4.07 | 3.99–4.14 | |
|
| 3.95 | 3.88–4.03 | |
|
| 3.85 | 3.79–3.92 | knee LAT |
|
| 3.72 | 3.65–3.79 | |
|
| 3.92 | 3.85–3.99 |
Comparison of the mean VGA across the imaging systems in knee PA and LAT protocols.
| System Comparison | Mean Difference | Std. Error | Sig. b | 95% CI | Projections |
|---|---|---|---|---|---|
|
| −0.21 | 0.069 | 0.001 | −0.34–0.07 | knee PA |
|
| −0.09 | 0.069 | 0.27 | −0.23–0.04 | |
|
| 0.11 | 0.069 | 0.12 | −0.02–0.25 | |
|
| 0.13 | 0.07 | 0.016 | 0.02–0.25 | knee LAT |
|
| −0.07 | 0.07 | 0.49 | −0.18–0.05 | |
|
| −0.2 | 0.07 | <0.001 | −0.31–−0.09 |
Based on estimated marginal means. b Adjustment for multiple comparisons: Bonferroni.
Figure 3VGA estimated marginal mean comparison across the imaging systems and image quality criteria, (a) for the knee PA protocol and (b) knee LAT protocol.
Calculated patient ED for the chest and knee protocols in all imaging systems, ESD = entrance surface dose and ED differences between the LDSS imaging system and DR imaging systems.
| Imaging Systems | ESD (mGy) | DAP (mGycm2) | ED | ED | Protocol | Projections |
|---|---|---|---|---|---|---|
|
| 0.14 | 142.21 | 19.76 | chest | PA | |
| 0.15 | 153.22 | 21.55 | LAT | |||
|
|
|
|
| |||
|
| 0.16 | 168.93 | 39.04 | PA | ||
| 0.19 | 193.45 | 47.71 |
| LAT | ||
|
|
|
|
| |||
|
| 0.15 | 158.00 | 21.00 | PA | ||
| 0.17 | 179.86 | 40.90 |
| LAT | ||
|
|
|
|
| |||
|
| 0.17 | 54.91 | 0.0063 | knee | PA | |
| 0.17 | 55.40 | 0.0064 | LAT | |||
|
|
|
|
| |||
|
| 0.18 | 58.25 | 0.0099 | PA | ||
| 0.19 | 59.33 | 0.0089 |
| LAT | ||
|
|
|
|
| |||
|
| 0.17 | 56.24 | 0.0086 | PA | ||
| 0.18 | 57.05 | 0.0096 |
| LAT | ||
|
|
|
|
|