Comparison of image quality in chest, hip and pelvis examinations between mobile equipment in nursing homes and static indirect radiography equipment in the hospital.

H Precht1, D L Hansen2, B M Ring-Pedersen3, L F Møller Hansen3, D Waaler4, A Tingberg5, M Midtgaard6, M G Jensen Ohlsen6, S T Juhl Hankelbjerg6, P Ravn6, I E Jensen6, J K Christensen6, P A Blackburn Andersen6.   

Abstract

INTRODUCTION: A hospital environment can be a significant burden and a health risk especially for dementia patients. Mobile x-ray equipment (ME) is used to enable imaging of these patients at home. The aim was to compare image quality (IQ) of chest, hip and pelvis images from ME to the stationary equipment (SE) used in a hospital department.
METHODS: We analysed examinations of the chest (n = 20), hip (n = 64) and pelvis (n = 32). Images were equally obtained from each setting of ME and SE. All images were graded using Visual Grading Analysis (VGA) by three radiographers (hip and pelvis) and three radiologists (chest). Technical IQ assessment was done by 80 additional images of a Contrast-Detail Radiography phantom (CDRAD).
RESULTS: All chest images were approved for diagnostic use, as well as the hip AP and pelvis images from SE. 'Approved proportion of ME images was for HIP antero-posterior (AP) and pelvis, 78% [95% CI: 52-94%] and 81% [95% CI: 54-96%] respectively. Hip axial had an overall low, but not significant different approval rate. Ordered logistic regression indicated higher IQ of HIP AP and pelvic images from SE. This contrasts that the CDRAD substudy indicated better IQ, expressed as IQFinv, from ME.
CONCLUSION: The VGA showed higher IQ for the SE system, while the CDRAD showed higher IQ for the ME system. IMPLICATIONS FOR PRACTICE: Dementia patients can be examined at their home if the acquisition is optimised according to image quality in conjunct to radiation dose. Performing imaging out of the hospital and coordinating the patients' further treatment are new work areas for radiographers and requires excellent communication skills.
Copyright © 2019 The College of Radiographers. Published by Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Image quality; Indirect digital radiography; Mobile radiography; Nursing homes

Year:  2019        PMID: 32052778     DOI: 10.1016/j.radi.2019.10.004

Source DB:  PubMed          Journal:  Radiography (Lond)        ISSN: 1078-8174


  3 in total

1.  COMPARISON OF CONVENTIONAL HAND EXAMINATION ON SIX OPTIMISED DR SYSTEMS.

Authors:  Helle Precht; Claus Bjørn Outzen; Martin Weber Kusk; Malene Bisgaard; Dag Waaler
Journal:  Radiat Prot Dosimetry       Date:  2021-05-31       Impact factor: 0.972

2.  Bringing radiology to patient's home using mobile equipment: A weapon to fight COVID-19 pandemic.

Authors:  Moreno Zanardo; Simone Schiaffino; Francesco Sardanelli
Journal:  Clin Imaging       Date:  2020-06-18       Impact factor: 1.605

3.  Visual Evaluation of Image Quality of a Low Dose 2D/3D Slot Scanner Imaging System Compared to Two Conventional Digital Radiography X-ray Imaging Systems.

Authors:  Ahmed Jibril Abdi; Bo Mussmann; Alistair Mackenzie; Oke Gerke; Gitte Maria Jørgensen; Thor Eriksen Bechsgaard; Janni Jensen; Lone Brunshøj Olsen; Poul Erik Andersen
Journal:  Diagnostics (Basel)       Date:  2021-10-19
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.