Literature DB >> 32599376

Survey of chest radiography systems: Any link between contrast detail measurements and visual grading analysis?

Sunay Rodríguez Pérez1, Nicholas W Marshall2, Joke Binst3, Johan Coolen3, Lara Struelens4, Hilde Bosmans2.   

Abstract

PURPOSE: To evaluate image quality of chest radiography for a number of systems in Belgium, using a contrast-detail (c-d) test object and Visual Grading Analysis (VGA) of an anthropomorphic phantom.
METHODS: The study comprised 22 chest imaging systems in Belgium. C-d data were measured using Leeds TO20 test object, imaged using poly(methyl methacrylate) (PMMA) thicknesses of 9, 13 and 16 cm. Images of the Lungman phantom, with additional tissue-equivalent chest plates to represent different patient sizes, were then acquired. Perceived image quality was evaluated using VGA by three radiologists. Images were acquired at a patient equivalent position with system-specific exposure settings for Posterior-Anterior chest protocol. Incident air kerma (IAK) was measured using a solid-state dosemeter.
RESULTS: C-d results showed large differences between the systems. Total number of visible discs ranged from 38 to 83 (for 9 cm PMMA) with a consistent average drop of 10% as PMMA thickness was systematically increased. However, no correlation was found between number of visible discs and IAK. Perceived image quality scored by the readers from the Lungman images decreased with increasing phantom thickness, however no correlation of VGA score with IAK was seen. Moderate correlation was found between the VGA score of one of the readers and the TO20 results, and no correlation for the rest.
CONCLUSIONS: The spread in dose and image quality measures was high and no correlation was seen between either image quality measure and IAK, suggesting the need for optimization. A more powerful tool is required for task-based optimization in chest radiography.
Copyright © 2020 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Anthropomorphic phantom; Chest radiography; Contrast-detail object; Visual grading analysis

Mesh:

Year:  2020        PMID: 32599376     DOI: 10.1016/j.ejmp.2020.06.014

Source DB:  PubMed          Journal:  Phys Med        ISSN: 1120-1797            Impact factor:   2.685


  1 in total

1.  Visual Evaluation of Image Quality of a Low Dose 2D/3D Slot Scanner Imaging System Compared to Two Conventional Digital Radiography X-ray Imaging Systems.

Authors:  Ahmed Jibril Abdi; Bo Mussmann; Alistair Mackenzie; Oke Gerke; Gitte Maria Jørgensen; Thor Eriksen Bechsgaard; Janni Jensen; Lone Brunshøj Olsen; Poul Erik Andersen
Journal:  Diagnostics (Basel)       Date:  2021-10-19
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.