Jan J Lang1,2, Mirjam Bastian1, Peter Foehr1,2, Michael Seebach3, Jochen Weitz4, Constantin von Deimling1,5, Benedikt J Schwaiger6, Carina M Micheler1,3, Nikolas J Wilhelm1, Christian U Grosse2, Marco Kesting7, Rainer Burgkart1. 1. Department of Orthopedics and Sports Orthopedics, Klinikum rechts der Isar, School of Medicine, Technical University of Munich, Munich, Germany. 2. Department of Mechanical Engineering, Chair of Non-destructive Testing, Technical University of Munich, Munich, Germany. 3. Department of Mechanical Engineering, Institute for Machine Tools and Industrial Management, Technical University of Munich, Munich, Germany. 4. Department of Oral and Maxillofacial Surgery, Josefinum, and Private Practice for Oral and Maxillofacial Surgery at Pferseepark, Augsburg, Germany. 5. Department of Mechanical Engineering, Chair of Applied Mechanics, Technical University of Munich, Munich, Germany. 6. Department of Diagnostic and Interventional Neuroradiology, Klinikum rechts der Isar, School of Medicine, Technical University of Munich, Munich, Germany. 7. Department of Oral and Cranio-Maxillofacial Surgery, School of Medicine, Friedrich-Alexander University Erlangen-Nuernberg, Erlangen, Germany.
Abstract
In this study, topology optimized, patient specific osteosynthesis plates (TOPOS-implants) are evaluated for the mandibular reconstruction using fibula segments. These shape optimized implants are compared to a standard treatment with miniplates (thickness: 1.0 mm, titanium grade 4) in biomechanical testing using human cadaveric specimen. Mandible and fibula of 21 body donors were used. Geometrical models were created based on automated segmentation of CT-scans of all specimens. All reconstructions, including cutting guides for osteotomy as well as TOPOS-implants, were planned using a custom-made software tool. The TOPOS-implants were produced by electron beam melting (thickness: 1.0 mm, titanium grade 5). The fibula-reconstructed mandibles were tested in static and dynamic testing in a multi-axial test system, which can adapt to the donor anatomy and apply side-specific loads. Static testing was used to confirm mechanical similarity between the reconstruction groups. Force-controlled dynamic testing was performed with a sinusoidal loading between 60 and 240 N (reconstructed side: 30% reduction to consider resected muscles) at 5 Hz for up to 5 · 105 cycles. There was a significant difference between the groups for dynamic testing: All TOPOS-implants stayed intact during all cycles, while miniplate failure occurred after 26.4% of the planned loading (1.32 · 105 ± 1.46 · 105 cycles). Bone fracture occurred in both groups (miniplates: n = 3, TOPOS-implants: n = 2). A correlation between bone failure and cortical bone thickness in mandible angle as well as the number of bicortical screws used was demonstrated. For both groups no screw failure was detected. In conclusion, the topology optimized, patient specific implants showed superior fatigue properties compared to miniplates in mandibular reconstruction. Additionally, the patient specific shape comes with intrinsic guiding properties to support the reconstruction process during surgery. This demonstrates that the combination of additive manufacturing and topology optimization can be beneficial for future maxillofacial surgery.
In this study, topology optimized, patient specific osteosynthesis plates (TOPOS-implants) are evaluated for the mandibular reconstruction using fibula segments. These shape optimized implants are compared to a standard treatment with miniplates (thickness: 1.0 mm, titanium grade 4) in biomechanical testing using human cadaveric specimen. Mandible and fibula of 21 body donors were used. Geometrical models were created based on automated segmentation of CT-scans of all specimens. All reconstructions, including cutting guides for osteotomy as well as TOPOS-implants, were planned using a custom-made software tool. The TOPOS-implants were produced by electron beam melting (thickness: 1.0 mm, titanium grade 5). The fibula-reconstructed mandibles were tested in static and dynamic testing in a multi-axial test system, which can adapt to the donor anatomy and apply side-specific loads. Static testing was used to confirm mechanical similarity between the reconstruction groups. Force-controlled dynamic testing was performed with a sinusoidal loading between 60 and 240 N (reconstructed side: 30% reduction to consider resected muscles) at 5 Hz for up to 5 · 105 cycles. There was a significant difference between the groups for dynamic testing: All TOPOS-implants stayed intact during all cycles, while miniplate failure occurred after 26.4% of the planned loading (1.32 · 105 ± 1.46 · 105 cycles). Bone fracture occurred in both groups (miniplates: n = 3, TOPOS-implants: n = 2). A correlation between bone failure and cortical bone thickness in mandible angle as well as the number of bicortical screws used was demonstrated. For both groups no screw failure was detected. In conclusion, the topology optimized, patient specific implants showed superior fatigue properties compared to miniplates in mandibular reconstruction. Additionally, the patient specific shape comes with intrinsic guiding properties to support the reconstruction process during surgery. This demonstrates that the combination of additive manufacturing and topology optimization can be beneficial for future maxillofacial surgery.
Successful reconstruction of a highly defective mandible is a complex and demanding surgery. Functionality as well as aesthetics are important factors for the outcome and the quality of patients’ life. There exist several indications for mandibular reconstruction like resection of tumorous tissue, osteomyelitis, osteonecrosis or trauma [1]. In many cases a free fibular flap is used as substitute for the lost bone [2-5]. This transplantation of a fibular segment includes soft tissue cover and reconnection to the vascular system. For the fixation of the bone material either a reconstruction plate or miniplates are used in most cases.Reconstruction plates are used as a more rigid alternative to miniplates. However, reconstruction plates have a high volume and adjustment to the mandible curvature is difficult, due to their increased rigidity [6]. Additionally, postoperative plate exposure is often reported with reconstruction plates [7,8]. Miniplates on the other hand come with opposite characteristics. They have a reduced thickness and the surgeon is able to adjust the implants by intraoperative bending to the bone surface. In case of infection single plates can be removed, in contrast to reconstruction plates, which have to be taken out as a whole [9]. Even though, the miniplates come with several advantages, the relatively high failure rate proofs that there is still potential for improvement. It is reported that in around 1 out of 10 reconstructions with miniplates implant failure is detected [2,3]. To increase the routine level and to better support the surgeon in recreating a physiological jawline, virtual planning software is used. Weitz et al. reported that pre-surgical planning significantly improves the outcome of the reconstruction in terms of bone consolidation and reproduction of the native mandibular angle [2]. The common tools for surgical planning need the geometrical data of the patient as input, which is obtained by computed tomography (CT). Consequently, it is a logical step to further exploit this data. Not only the resection of the bone and the positioning of the fibular segments can be planned by using the three-dimensional models of fibula and mandible but also patient individual implants can be created. During planning, virtual implants can be fitted and adapted to the surface of the bone segments. With modern production methods from the field of additive manufacturing of metals for medical application an on-time fabrication of these specific implants can be provided. Using patient specific implants for mandible reconstructions has a positive effect on the treatment process. The guiding properties of these implants lead to a reduced surgery time and costs [10-13]. Additionally, they enable precise reconstructions [14,15] with short duration of postoperative care [16] and fewer complications [10,17].In this study a new approach for improving the mandibular reconstruction is developed and evaluated. In an interdisciplinary project topology optimized, patient specific implants are created, which combine advantages of miniplates and reconstruction plates. The patient specific implants have a small volume, optimized mechanics and modular exchangeability. Topology optimization is a powerful mathematical tool which allows creating an optimal structural design within prescribed loading and boundary conditions. The material distribution method for topology optimization is used in a variety of fields like aviation or structural engineering to create specialized construction parts with an optimized design based on an predefined design space [18]. These parts often have a high mechanical stability despite a reduced volume compared to conventional parts. Crudely, this is achieved by iteratively evaluating finite element simulations containing a design space, which has a homogeneous material distribution at the start. After every iteration the material distribution is altered depending on the distribution of the internal stresses in the design space. To obtain the final design, this is continued as long as the boundary conditions (e.g. maximum stress) are not hurt.For optimization of the evaluated implants in this study, an approach described by Seebach et al. is used [19]. This algorithm creates osteosynthesis plates with a reduced volume while maintaining a high stiffness. In addition, emphasis was placed on an evenly distributed loading of the fixation screws, to prevent stress shielding and screw failure by overloading. The geometry of the implant with the optimized volume leads to a reduced contact area to the bone, which improves the healing capacities due to less periost disturbance. The shape of the implants fits to the bone geometry of the patient and serves the surgeon as guides for the reconstruction during surgery. This can help to reduce the surgery time for mandibular reconstruction. Topology optimization has been described for the design of osteosynthesis plates in maxillofacial surgery in rare cases as virtual design studies [20-22]. But the biomechanical evaluation with biological specimen for these topology optimized implants is still missing in literature.The aim of this study is a biomechanical evaluation of the of the newly developed implants for mandible reconstruction. This is also used for validation of the production process starting from medical imaging until final reconstruction. Next to the production of the implants with additive manufacturing, this includes virtual planning of the reconstruction as well as automatic design of the implants and cutting guides. For biomechanical evaluation, two different treatment methods for fibula-reconstructed mandibles are compared by static and dynamic testing: standard miniplates versus topology optimized, patient specific osteosynthesis plates (TOPOS-implants).
Materials and methods
For this biomechanical study, 21 pairs of fresh frozen (-28°C) cadaveric human mandibles and respective right fibulas were used (Medcure Inc., Portland/OR, USA;). The study was approved by the Technical University ethics committee (607/20 S-KH). The accredited company providing the specimen adheres to AATB standards (American Association of Tissue Banks) and all donors have to provide informed written consent prior to death to be accepted as donors. The average donor age was 67.8 ± 8.8 years (mean ± standard deviation). Ten donors were female (70.0 ± 10.8 years) and eleven donors were male (65.8 ± 6.3 years). Considering an equal distribution of sex and age, two groups were created—one for treatment with miniplates (six male, five female; 68.9 ± 8.1 years) and one for treatment with TOPOS-implants (five male, five female; 66.6 ± 9.8 years). Within these two groups, specimens were split up for static (n = 3; one female, two male) and dynamic (n = 8 resp. n = 7; 4 female, 4 resp. 3 male) testing. All specimens underwent two freezing and thawing cycles during the whole process of preparation, reconstruction and testing. While testing, all specimens were wrapped in moist sheets to avoid dehydration.High-resolution CT-scans (IQon—Spectral CT) of all frozen specimens were taken using the following standard settings for clinical head scans: tube current intensity 403 mA, tube voltage 120 kV and tube current-time product 300 mAs. Based on the DICOM (Digital Imaging and Communications in Medicine) data of the CT-scans, segmentation was performed using a Matlab-script (The MathWorks Inc., Natick, Massachusetts, USA) to create geometrical models of all mandibles and fibulae. Thereby, artifacts from implants or other dental treatments were reduced and finally all geometries existed in STL (standard tessellation language) file format. The open-source software blender (Blender Foundation, Amsterdam, Netherlands) was used to fix holes in the mesh data and to make sure there was a continuous surface for each 3D model.A large mandibular LC-defect on the right mandibular side was chosen for the reconstruction. This defect was named according to the classification of Jewer et al. reaching from the right mandibular angle to the left edge of the anterior segment [23]. All individual reconstructions of the mandibles were pre-planned by an experienced surgeon, using a newly developed virtual software tool (Fig 1). This interactive software tool uses the 3D models of the patient’s mandible and fibula and enables the surgeon to plan each reconstruction individually, including resection and repositioning. Matching resection guides are created automatically. The recently published software of Raith et al. includes an algorithm that focuses on helping the surgeon to find the best fit of defined fibula segments determined by distinctive parameter setting [24]. In contrast to this, the planning tool used in our study allows a specific adaption of the reconstruction by shifting and rotating all cutting planes and segments freely. After manual placement of three resection planes at the intact mandible, two fibula segments are automatically projected on the resected mandible model. Furthermore, it should be mentioned, that the resection plane in the middle only marks the contact plane of both fibular segments. The mandible does not have to be cut at this position. Manual adaption of the reconstruction can be done freely by movement and rotation of both fibular segments, by changing the position and the angle of the resection planes as well as by altering the harvesting position on the fibula. Doing the virtual reconstruction, the blood supply of the fibula has to be considered because the vessel is supposed to be connected to the corresponding vascular system of the mandible during transplantation for better healing conditions. The performing surgeon did consider this during planning of the reconstruction for this study.
Fig 1
Virtual reconstruction planning with a custom-made, newly developed software tool.
Patient specific 3D models of mandibula and fibula are used. (a) Resection planes are adjusted freely for the mandible by the surgeon. (c) Fibula segments are automatically projected into (b) the defect and can be adjusted for optimal harvesting site.
Virtual reconstruction planning with a custom-made, newly developed software tool.
Patient specific 3D models of mandibula and fibula are used. (a) Resection planes are adjusted freely for the mandible by the surgeon. (c) Fibula segments are automatically projected into (b) the defect and can be adjusted for optimal harvesting site.In addition, a data set of specific cutting guides for each reconstruction was created from the geometrical information of the resection planes in STL-format. There was one individual cutting guide for each fibula with four saw blade insertion to create the two virtually planned segments. For each mandible two cutting guides were used, one for each jaw angle. All cutting guides were produced via additive manufacturing using standard thermoplastics in fused filament fabrication.For the newly developed implants, the reconstruction was planned with the use of three master implants, one for each osteosynthesis site of the LC-defect reconstruction. The set of master implants was topology optimized for a LC-defect reconstruction using the geometry of synthetic standard mandible (mandible 8900 from Synbone AG, Malans, Switzerland). The optimization algorithm includes an iterative finite element analysis using Hyperworks and OptiStruct (Altair Engineering, Troy, USA) to achieve an implant with high stiffness despite having a reduced volume to 45% on average (compared to design space). In other words, the optimization is a material reduction process that results in the least possible loss of stiffness for the implant. Another aim for the optimization was an even stress distribution among the bone screws to prevent screw failure. This optimization process was described before by Seebach et al [19]. For the presented study the shape of these topology optimized master implants was fitted to ten virtual reconstructions of the TOPOS group based on the CT data of the specimens (Fig 2). With this consecutive optimization of topology and shape, 30 patient specific plates (three per reconstruction) with a thickness of 1 mm were created and produced by electron beam melting (FIT Production GmbH, Lupburg, Germany) using medical-grade titanium alloy (Ti6Al4V, ISO 5832-3/ASTM F1472) as described by Seebach et al. [25]. For post-processing, grinding and polishing were applied.
(Top) Virtual planning and fitting of the osteosynthesis plates for the fixation of the fibular segments. (Bottom) Application of the implants to the specimen produced with electron beam melting.
(Top) Virtual planning and fitting of the osteosynthesis plates for the fixation of the fibular segments. (Bottom) Application of the implants to the specimen produced with electron beam melting.The preparation of all specimens included removing of soft tissue including muscle attachments, blood vessels, fat and periosteum. This was done for the cutting guides to fit accurately because they are created on the base of CT-data and soft tissue was not included in the segmentation. For osteotomy the specific cutting guides were applied on the mandible and fibula (Fig 3). During the reconstruction of a mandible, the first step was to arrange and combine both fibular segments. Afterwards they were connected with the corresponding mandible segment on each side.
Fig 3
Process from bone to reconstruction.
Stepwise process of creating the reconstruction specimen with topology optimized patient specific implants from the intact bone to the reconstructed mandible for testing. An experienced surgeon performed virtual planning and reconstruction. For reconstruction with miniplates the steps are similar but without the implant fitting.
Process from bone to reconstruction.
Stepwise process of creating the reconstruction specimen with topology optimized patient specific implants from the intact bone to the reconstructed mandible for testing. An experienced surgeon performed virtual planning and reconstruction. For reconstruction with miniplates the steps are similar but without the implant fitting.In case of the reconstruction with miniplates (cranial plate, straight, rigid; plate thickness 1.0 mm; material: titanium grade 4 (ASTM F67); Modus 2.0, Medartis AG, Basel), the implants for joining fibular and mandibular segments had to be aligned to the individual bone geometry by an experienced oral and maxillofacial surgeon. Plates with four holes (bar: 9 mm) and plates with six holes (bar: 9 mm or 12 mm) as well as appropriate screws (cross; material: titanium grade 4 (ASTM F67); Modus 2.0, Medartis AG, Basel) with a length 6 mm or 9 mm were used. The fixation of corresponding bone segments was performed with standard surgical equipment. The position of the screws was chosen for every reconstruction individual by surgeon. The TOPOS implants were fixated to the bone with the same kind of screws as used for the miniplates. For all reconstructed mandibles, a standardized hole (d = 5 mm) at the middle of each ramus mandibulae near the foramen mandibulae was drilled for fixation of the specimens in the test system and the processus coronoideus was removed.After reconstruction of the mandibles either using miniplates or TOPOS-implants, static and dynamic testing was performed on a custom-made multi-axial test system. The system allows adapting to the anatomical conditions of the donor and applying side specific loads [26]. For the determination of the applicable load direction and appropriate bearings for the test setup, a model with all relevant muscles that are part of chewing was developed. The orientation of the muscle forces of the major muscle groups were taken from literature [27-29]. The combination of the muscle orientations for the mandible was used to derive the direction of the force for testing on each side of the mandible. The resulting direction of the force, which was applied in the test system, is mainly orthogonal to the chewing plane with a pivoting of 7° anterior. Tilting of the force direction of calculated 0.4° to lateral was neglected for the setup. Three bearing points were used for the model and derived to the testing setup. The bite point was defined as the anterior end of the left segment (first premolar if present), since the incisors are missing for the reconstructions. Choosing this bearing point position and the front edge of the mandibular segment created high bending moments for the reconstruction. This bite point was only constrained in vertical direction [30]. This allows frictional displacement in the horizontal plane for the bite point. The capita mandibulae were the remaining two bearings, which were constrained in two translatory directions. Allowing a limited, frictional displacement in medio-lateral direction is in contrast to the conditions in numerical models [29,31,32]. But the risk of asymmetrical load distribution due to asymmetrical bearing is reduced.The custom-made test system, which allowed mechanical testing of mandibular reconstructions to analyze the fatigue behavior of the osteosynthesis plates according to DIN 50100, was force-controlled and it was possible to test two specimens at the same time. Load was applied through two bowden cables per mandible (one for each side, Fig 4), that followed the orientation of the resulting force vector and lead to hydraulic pistons via deflection rollers. For each of the cables a load cell was incorporated into the loading line, which was used for force control and documentation. Aluminum bolts with semicircular grooves at the lower end were used to support the mandibular condyles. These bearings were adjustable to the geometry of the specimens by altering the height and orientation of the groove position. At the opposite side, there was a metallic angle bracket that simulates the maxilla and which could be moved in three directions to adjust the position. That means the left front part of the mandible was in contact with the angle so movement in vertical direction was inhibited at this point (Fig 4), similar to the setup of Schupp et al. [33].
Fig 4
Biomechanical test setup for dynamic and static testing.
(a) test setup for dynamic testing of the constructed specimen, force is applied through the cables fixed to the ramen mandibulae (b) front view of the dynamic test setup with an exemplary reconstruction using miniplates (c and d) static testing of the reconstruction in the same test rig as for dynamic testing with additional displacement sensors measuring (c) the vertical displacement on the mandible (red circle) at the contact plane of the fibula segments and (d) the lateral displacement at the ramen mandibulae.
Biomechanical test setup for dynamic and static testing.
(a) test setup for dynamic testing of the constructed specimen, force is applied through the cables fixed to the ramen mandibulae (b) front view of the dynamic test setup with an exemplary reconstruction using miniplates (c and d) static testing of the reconstruction in the same test rig as for dynamic testing with additional displacement sensors measuring (c) the vertical displacement on the mandible (red circle) at the contact plane of the fibula segments and (d) the lateral displacement at the ramen mandibulae.To analyze the deformation and to gain information about stiffness and maximum load up to failure during static testing, two linear potentiometric position sensors (MMR10-11, Megatron, Germany; range: 10 mm) connected to a 16-bit-analog-digital converter were attached. Sensor 1 was used to measure displacement in vertical direction at the contact point of the fibular segments while sensor 2 registered lateral displacement on the right mandibular angle (Fig 4C and 4D). At the beginning of each test run, both sides of the mandible were preloaded with 20 N.For static testing (n = 3 per implant group; one female, two male), a force ramp was applied to the mandibular angles up to 500 N. Failure of the bone-implant-construct was determined as fracture or obvious deformation of bone or implant. Sensor 1 and sensor 2 were measuring the vertical and lateral displacement of the reconstructed mandible. The end of the linear elastic region (elastic limit) was determined by considering the load-displacement-diagrams of both sensors. The results from the static testing were used to set the maximum load for the dynamic testing of the reconstructions and to validate the mechanical similarity of the reconstruction methods.The maximum force for dynamic testing F was defined as 85% of the force at elastic limit, which was derived from the displacement measurements during static testing. The maximum load on the reconstructed side of the mandible F was reduced by 30%, in order to consider the resection of muscles during surgery and reduced chewing loads, similar to the approach of Schupp at al. [33]. Sinusoidal loading between F = 60 N and F = 240 N (F = 168 N) at a frequency of 5 Hz was performed for up to 5 · 105 cycles to analyze the fatigue properties of both treatment methods (miniplates: n = 8; TOPOS-implants: n = 7). The frequency was chosen to reduce the effect of biological degradation as far as possible. The maximum piston speed for load changes at 5 Hz was sufficient to provide the expected displacements for the applied load interval. After N = 100 cycles all subsidence processes are finished and stable testing can be supposed. For all following cycles, a change in the maximum displacement of more than 5 mm, related to the position at N, was considered as failure of the reconstructed mandible and led to an automatic stopping.To validate the variation in bone quality from the different donors the thickness of the cortical bone was measured at the ramen mandibulae according to Heibel et al. with a caliber [34]. Three measurement sites were taken at the resection plane of the right mandibular angle for evaluation (Fig 5).
Fig 5
Measurement of the cortical thickness of the mandible at the resection plane of the right mandibular angle.
Measurement of the cortical thickness of the mandible according to Heibel et al. [34].
Measurement of the cortical thickness of the mandible at the resection plane of the right mandibular angle.
Measurement of the cortical thickness of the mandible according to Heibel et al. [34].Since the distribution of the results in the different groups deviated from a normal distribution, which was tested with a Shapiro-Wilk normality test, the conditions for a standardized Student’s t-test were not met. Due to that, a nonparametric Mann-Whitney U test was used to analyze the statistical significance of the difference between the groups. Standard significance thresholds of p < 0.05 and p < 0.01 were used. The evaluation was performed with the software tool Prism 6.07 (GraphPad Software, Inc., San Diego, USA).
Results
In static testing, the elastic limit of bone-implant-constructs for the group with miniplates was reached at 287 ± 57 N (mean ± standard deviation) while the TOPOS-implant group had their mean value at 276 ± 125 N. In the elastic deformation range a vertical stiffness of 181 ± 33 N/mm was derived for the reconstruction with miniplates and 151 ± 16 N/mm for reconstruction with TOPOS-implants. At the time of failure, a vertical displacement of 2.6 ± 1.3 mm was measured for reconstructions with miniplates and 2.5 ± 1.4 mm for reconstructions with TOPOS-implants. Lateral displacement of the right mandibular angle was smaller for specimens treated with miniplates (0.7 ± 0.3 mm) than for those with TOPOS-implants (2.0 ± 1.6 mm). Failure of the reconstructions occurred at 413 ± 115 N (miniplates) and 400 ± 217 N (TOPOS-implants). Occurring failure types were bone or tooth failure and massive plastic deformation of the implant. This deformation of the implant happened only to one reconstruction with miniplates (S1 Table).Even though the reconstructions of both groups showed a very similar mechanical behavior under static loading, in cyclic testing a significant difference between the groups was observed. A distinction was made between three different outcome types: bone fractures, plate fractures and specimens that passed all applied cycles without failure (Fig 6). Screw failure was not observed for all tests. Passing all applied cycles without failure only occurred within the group of TOPOS-implants, where five reconstructions passed 5 · 105 cycles without failures. Plate fractures solely happened in the miniplate group in five out of eight cases (Fig 7C). Three of these fractures were localized at a plate that connected the right mandibular segment with the lateral fibular segment. Two plate fractures occurred at a miniplate holding on to the left mandibular segment and the central fibular segment. Bone fractures at the right ramus mandibulae appeared in both groups (for miniplates: n = 3, for TOPOS-implants: n = 2). In these cases the fracture line was always running through the upper screw holes in the bone for the corresponding implant (Fig 7A and 7B) When disregarding bone fractures, it is shown that reconstructions with miniplates reached an average number of 1.32 · 105 ± 1.46 · 105 cycles (n = 5). A highly significant difference between the groups can be determined because no TOPOS-implant failed for the applied cycles (Mann-Whitney U = 0; p = 0.008<0,01 two-tailed). Considering all specimens, reconstruction failure (including bone fracture) is noted after 3.97 · 105 ± 1.94 · 105 load cycles in the TOPOS-implant group and at 8.37 · 104 ± 1.29 · 105 load cycles after treatment with miniplates. Statistically this also corresponds to a significant difference between the groups (Mann-Whitney U = 8; p = 0.026<0,05 two-tailed). Statistical analysis was performed according to two non-normally distributed groups (Shapiro-Wilk: miniplates group p = 0.002; TOPOS-group p = 0.0008).
Fig 6
Results of the dynamic tests.
Graphical visualization of the biomechanical comparison of miniplates with patient specific, topology optimized (TOPOS-)implants. The bars show the average of the passed cycles per group with 500 000 cycles at maximum. For comparing of fatigue properties of the implant types only test runs without bone fracture are used (statistics: significant (*) p<0.05, highly significant (**) p<0.01).
Fig 7
Failure cases of the biomechanical testing.
Bone fracture occurred at the right mandibular angle with the fracture line running across the upper fixation screws for (a) miniplates (two views) and (b) TOPOS-implants (two views); (c)Failure of the implant only occurred for reconstructions with miniplates at both fixation sites of the mandibular segment to the fibular segment(left: left mandibular segment to central fibular segment; right: right mandibular segment to right fibular segment).
Results of the dynamic tests.
Graphical visualization of the biomechanical comparison of miniplates with patient specific, topology optimized (TOPOS-)implants. The bars show the average of the passed cycles per group with 500 000 cycles at maximum. For comparing of fatigue properties of the implant types only test runs without bone fracture are used (statistics: significant (*) p<0.05, highly significant (**) p<0.01).
Failure cases of the biomechanical testing.
Bone fracture occurred at the right mandibular angle with the fracture line running across the upper fixation screws for (a) miniplates (two views) and (b) TOPOS-implants (two views); (c)Failure of the implant only occurred for reconstructions with miniplates at both fixation sites of the mandibular segment to the fibular segment(left: left mandibular segment to central fibular segment; right: right mandibular segment to right fibular segment).The results for the measuring of the cortical thickness over all specimen are shown in Table 1. The measurement was performed at three sites, while b corresponds to the site were all fractures of the ramen mandibulae occurred.
Table 1
Measurement results of the cortical thickness of the mandible for all specimen (mean ± standard deviation).
A
B
C
All specimen
2.2 ± 0.6 mm
1.5 ± 0.5 mm
1.9 ± 0.5 mm
Specimen with bone fracture
1.6 ± 0.3 mm
1.0 ± 0.2 mm
1.7 ± 0.4 mm
Specimen without bone fracture
2.7 ± 0.3 mm
2.0 ± 0.3 mm
2.2 ± 0.5 mm
Measurement locations A, B and C can be seen in Fig 5.
Measurement locations A, B and C can be seen in Fig 5.
Discussion
The biomechanical testing in this study was performed on an in-house developed testing setup. Using a custom-made solution like this comes with the disadvantage of limited reproducibility for third parties. But a commercially available solution does not exist for this testing scenario. And due to the detailed description of the testing and the publication presenting the experimental setup by Foehr et al. [26], distinct transparency, comprehensibility and adaptability of the testing conditions are ensured.In static testing the reconstructions with the TOPOS-implants as well as miniplates behaved quite similar. Elastic limit and maximum force until failure differ only 4.2% and 3.1% respectively. Furthermore, fractures occur at the same site and the vertical displacement is almost the same. However, the lateral displacement is larger for reconstructions with TOPOS-implants. We assume, that the shape of the implant, which is fitted to the bone geometry, creates a more rigid structure of all four bony segments. This enables an increased force transmission between the segments that are close to each other. As a consequence, the left side with the higher load would have more influence on the reconstructed right side, at which the displacement had been measured. Consequently, the local loading and thus the displacement at the right mandibular angle would be higher. However, the measured difference in the average displacements of 1.33 mm also can be seen as clinically uncritical. Furthermore, a group size of n = 3 is too small to make reliable statements on this. Which is why the standard deviation in both groups for the maximum load is relatively high. In both groups the female mandibles, which were noticeably smaller, failed distinctly earlier (280 N (miniplates) resp. 150 N (TOPOS-implants); n = 1 for each group) compared to the male specimens (at 480 ± 14 N (miniplates) resp. 525 ± 21 N (TOPOS-implants); n = 2 for each group). This caused a high deviation regarding the failure load. The difference in the determined stiffness for the reconstruction types in static testing suggests that a slightly stiffer reconstruction of the mandible can be achieved with miniplates. But the actuator displacement during cyclic testing, which is range that the actuator moves from minimum load to the maximum load in one cycle, cannot confirm that (S2 Table). This displacement, which is also a measure for the stiffness of the reconstruction, is very similar for both reconstruction groups. The mean actuator displacement for the reconstruction with TOPOS implants was 2.1 ± 0.3 mm for the right side and 2.7 ± 0,24 mm for the left side at cycle 100. For the reconstruction with miniplates an actuator displacement of 2.2 ± 0.3 mm on the right side and 2.9 ± 0.3 mm on the left side was determined at the same cycle. This underlines the mechanical comparability of the two reconstruction approaches. The mechanical similarity also confirms that there is no excessive stiffening caused by the TOPOS implants. If the bone fragments are positioned too rigidly for osteosynthesis, it can result in inhibition of sufficient micromotions in the fracture gap [35,36]. Stress shielding like effects can lead to impaired healing in this context due to low interfragmentary strains at the fracture site [37]. But as already described there is no indication for this found in mechanical testing.The mechanical behavior of both specimen groups shows to be quite similar in static testing, but for the dynamic testing, the differences are more distinctive. The cyclic testing demonstrated clearly, that the TOPOS-implants, which had been optimized considering maximum stiffness and evenly distributed stresses, showed significantly better fatigue properties compared to miniplates. This is still true when considering only specimens without bone fracture. None of the TOPOS-implants failed during the 5 · 105 cycles, while the miniplates only reached 26% of these cycles on average. It is likely that the reconstructions with TOPOS-implants would have lasted significantly more cycles. Which means stopping the dynamic loading after half a million cycles puts the miniplates in a favourable light. The difference is therefore even more distinctive than shown here.Considering the diversity in shape and geometry of human bones, it also becomes apparent in Fig 6, that there is a huge variation in proportions of mandibles and corresponding fibulae. While in the Fig 6A mandible and fibula of the bone combination are approximately of the same width, the combination in Fig 6B shows a mandible that is in total (at the pictured resection plane) only as wide as the cortical shell of the corresponding fibula. However, only combinations of the same donor were used.Overall, nine out of 21 specimens failed because of bone fracture at the right ramus mandibulae. The fracture line ran along the line of the drilled holes for the upper miniplate respectively the upper holes of the corresponding TOPOS-implant. Consequently, it seems to be obvious that there is a link between the weakening of the bone structure through screws and the failure of the reconstructed mandible. It could be observed that in case of a bone fracture the performing surgeon used more often 9 mm screws instead of 6 mm at the right mandibular segment for fixation of the implants. This was done subjectively as compensation of structural instability. The longer screws enabled a bicortical fixation, while the 6 mm screws functioned monocortical. Considering the whole number of screws at the right mandible (six screws for miniplates, three screws for TOPOS-implants), 42% of screws were the long version in case of bone fracture, whereas only 16% were the long version when there was no bone fracture. Nevertheless, the question arises if either the usage of many long screws weakens the bone or if on an already weak bone with a high risk of fracture more 9 mm screws are used. According to Heibel et al., the corresponding region of the ramus mandibulae is the area with lowest thickness of the cortical bone in the mandible [34]. Also, 23–42% of all clinically seen mandible fractures happen at the mandibular angle, which underlines the natural mechanical instability of this area [20,38]. Additionally, the specimens with bone fracture showed a lower mean value of cortical bone thickness (1.0 ± 0.1 mm) than the average over all mandibles in the corresponding area (1.5 ± 0.5 mm) as well as the specimens, that did not fail because of bone fracture (2.0 ± 0.4 mm) (Table 1B). Due to this the area at the ramen mandibulae seems to be especially prone to fracture. Consequently, fixation in this area should be avoided or the number of used screws should be reduced for clinical mandibular reconstruction in cases of poor bone quality.The approach of using patient specific osteosynthesis plates to improve mandibular reconstruction is not new. Since 2013 a patient individual concept is provided by DePuy Synthes (DePuy Synthes, J&J Medical Devices, Warsaw, USA) in cooperation with Materialize (Materialize, Leuven, Belgium) which supports the presurgical planning of reconstruction plates to the needs of the surgeon [39,40]. Rendenbach et al. used a similar defect model as presented here in synthetic mandibles and compared a CAD/CAM designed reconstruction plate to treatment with miniplates and conventional reconstruction plates [40]. The standard reconstructions showed inferior mechanical properties in fatigue and stiffness compared to the CAD/CAM design. Also other companies like Stryker (Stryker, Kalamazoo, USA) and KLS Martin Group (Gebrüder Martin GmbH & Co. KG, Tuttlingen, Germany) are commercially providing patient individual implant planning and production [41,42]. Gutwald et al. even performed an optimization of the patient specific reconstruction plate by using a sensitivity analysis of 72 predefined options of the implant [42]. But the ultimate goal for the design of osteosynthesis plates should be that they provide stability to ideally support the healing of the bone, while interfering with it as little as possible. Topology optimization is a suitable option to reach this goal. Our study shows the first experimental evaluation of a topology optimization approach on osteosynthesis plates for reconstruction, but topology optimization has been seen in maxillofacial surgery before. Lovald et al. used topology optimization for creation of osteosynthesis plates for fracture treatment at the symphysis and lateral body of the mandible [21,22]. These optimized implants showed better results in plate stresses and fracture strain in comparison with standard treatments including miniplates using finite element analysis. Liu et al. also used topology optimization to design a V-shaped implant for the treatment of mandibular angle fracture [20]. The plate was numerically evaluated against standard osteosynthesis with miniplates of the same thickness. The simulations with the topology optimized implants had lower von-Mises stresses on the implant as well as a reduced fracture displacement. Finite element analysis is an important tool in the field of biomechanics, which is widely used to analyse the complex biomechanical behavior of the mandible. But disadvantages come with an evaluation of the optimized implants with finite element analysis. Simplifying the screw geometry with a cylinder leads to a distinct underestimation of the peak stresses in screws [43]. Furthermore, the reproduction of the complex reality of bone tissue is difficult in numerical models. In most simulations bone is divided in cortical and cancellous bone with isotropic material parameters each. In more sophisticated approaches the material parameters are dependent on the mineral content of the bone using CT data. But in reality, bone is a highly complex structure with anisotropic and viscoelastic material properties, which are not considered in numerical models. Also, an evaluation of the optimized implant on the same model and loading condition as it was created on is only the first step in the validation process of the implant. Comparing it numerically to standard implants that can be used universally, validates the topology optimization algorithm rather than the functional performance of the implant. Due to this we evaluated our topology optimized, patient specific implants on cadaveric models. This makes an application of the physiological loading conditions more difficult but provides a better representation of the material properties and the diversity of bone quality and geometry. Furthermore, we used a practical approach that puts the topology optimization before the adapting of the implants to the patient. This generalized optimization is very helpful in increasing the speed of generating an implant for different patients, which makes the integration of topology optimization into a fast and flexible treatment process much easier.For the evaluation of the whole process from the planning with the software to the production of the implants, reconstruction of the mandibles and finally biomechanical testing, the analysis was focused on LC-defects. But the concept of topology optimization and fitting of the shape to the patient’s bone geometry can be adapted for all kind of application of osteosynthesis plates.
Conclusions
This study displays a significantly increased cyclic loading capacity of mandible reconstructions with topology optimized and patient specific implants in comparison with miniplates as clinical standard. Due to their shape fitted to the bone surface they also function as guides for the surgeon to create the planned reconstruction, which can reduce surgery time. The presented combination of generalized optimization and consecutive shape adapting to the patient specific surface creates a time-efficient possibility to create optimized, patient specific implants. Additionally, the introduced implants share the advantages of miniplates by being site-specific exchangeable and not protruding as much as reconstruction plates. As additive manufacturing techniques are becoming more and more established in medical applications, the production costs will fall. Consequently, patient specific osteosynthesis plates, which are optimized to increase stability, could become widely used in mandibular reconstruction in future.
Measurement results of static testing and the cortical thickness of the mandible.
(DOCX)Click here for additional data file.
Measurement results dynamic testing and the cortical thickness of the mandible.
(DOCX)Click here for additional data file.21 Jan 2021PONE-D-20-39788Improving mandibular reconstruction by using topology optimization, patient specific design and additive manufacturing? – A biomechanical comparison against miniplates on human specimenPLOS ONEDear Dr. Lang,Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process.The manuscript is a well-conducted study providing novel findings in a relatively new field. Please address the comments/suggestions from the academic editor and reviewer 1.Please submit your revised manuscript by March, 1st 2021. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file.Please include the following items when submitting your revised manuscript:A rebuttal letter that responds to each point raised by the academic editor and reviewer(s). You should upload this letter as a separate file labeled 'Response to Reviewers'.A marked-up copy of your manuscript that highlights changes made to the original version. You should upload this as a separate file labeled 'Revised Manuscript with Track Changes'.An unmarked version of your revised paper without tracked changes. You should upload this as a separate file labeled 'Manuscript'.If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter.If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: http://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocolsWe look forward to receiving your revised manuscript.Kind regards,Luis CordovaAcademic EditorPLOS ONEJournal Requirements:When submitting your revision, we need you to address these additional requirements.1. Please ensure that your manuscript meets PLOS ONE's style requirements, including those for file naming. The PLOS ONE style templates can be found athttps://journals.plos.org/plosone/s/file?id=wjVg/PLOSOne_formatting_sample_main_body.pdf andhttps://journals.plos.org/plosone/s/file?id=ba62/PLOSOne_formatting_sample_title_authors_affiliations.pdf2. In the ethics statement in the manuscript and in the online submission form, please provide additional information about the human tissues used in this study. Specifically, please ensure that you have discussed whether next-of-kin provided informed written consent for the use of the tissues. If patients provided informed written consent prior to death to have their bodies used in medical research, please include this information.3. Please list the name and version of any software package used for statistical analysis, alongside any relevant references. For more information on PLOS ONE's expectations for statistical reporting, please see https://journals.plos.org/plosone/s/submission-guidelines.#loc-statistical-reporting.4. Please ensure you have thoroughly discussed any potential limitations of this study within the Discussion section, including the potential impact of confounding factors.5. We note that you have included the phrase “data not shown” in your manuscript. Unfortunately, this does not meet our data sharing requirements. PLOS does not permit references to inaccessible data. We require that authors provide all relevant data within the paper, Supporting Information files, or in an acceptable, public repository. Please add a citation to support this phrase or upload the data that corresponds with these findings to a stable repository (such as Figshare or Dryad) and provide and URLs, DOIs, or accession numbers that may be used to access these data. Or, if the data are not a core part of the research being presented in your study, we ask that you remove the phrase that refers to these data.6.Thank you for stating the following in the Financial Disclosure section:"This work was supported by the German Research Foundation (DFG) and the Technical University of Munich (TUM) in the framework of the Open Access Publishing Program.The research project “TOPOS - Development, Manufacturing and Testing of Topology Optimized Osteosynthesis Plates” (AZ-1019-12), in whose context the presented study was conducted, is funded by the Bavarian Research Foundation (BFS).The funders had no role in study design, data collection and analysis, decision to publish, or preparation of the manuscript."We note that one or more of the authors are employed by a commercial company: Josefinum, and private practice for Oral and Maxillofacial Surgery at Pferseeparka) Please provide an amended Funding Statement declaring this commercial affiliation, as well as a statement regarding the Role of Funders in your study. If the funding organization did not play a role in the study design, data collection and analysis, decision to publish, or preparation of the manuscript and only provided financial support in the form of authors' salaries and/or research materials, please review your statements relating to the author contributions, and ensure you have specifically and accurately indicated the role(s) that these authors had in your study. You can update author roles in the Author Contributions section of the online submission form.Please also include the following statement within your amended Funding Statement.“The funder provided support in the form of salaries for authors [insert relevant initials], but did not have any additional role in the study design, data collection and analysis, decision to publish, or preparation of the manuscript. The specific roles of these authors are articulated in the ‘author contributions’ section.”If your commercial affiliation did play a role in your study, please state and explain this role within your updated Funding Statement.b) Please also provide an updated Competing Interests Statement declaring this commercial affiliation along with any other relevant declarations relating to employment, consultancy, patents, products in development, or marketed products, etc.Within your Competing Interests Statement, please confirm that this commercial affiliation does not alter your adherence to all PLOS ONE policies on sharing data and materials by including the following statement: "This does not alter our adherence to PLOS ONE policies on sharing data and materials.” (as detailed online in our guide for authors http://journals.plos.org/plosone/s/competing-interests) . If this adherence statement is not accurate and there are restrictions on sharing of data and/or materials, please state these. Please note that we cannot proceed with consideration of your article until this information has been declared.Please include both an updated Funding Statement and Competing Interests Statement in your cover letter. We will change the online submission form on your behalf.Please know it is PLOS ONE policy for corresponding authors to declare, on behalf of all authors, all potential competing interests for the purposes of transparency. PLOS defines a competing interest as anything that interferes with, or could reasonably be perceived as interfering with, the full and objective presentation, peer review, editorial decision-making, or publication of research or non-research articles submitted to one of the journals. Competing interests can be financial or non-financial, professional, or personal. Competing interests can arise in relationship to an organization or another person. Please follow this link to our website for more details on competing interests: http://journals.plos.org/plosone/s/competing-interestsAdditional Editor Comments:The authors propose a well-conducted study to compare topology optimized, patient specific osteosynthesis plates (TOPOS-implants) versus 1.0 mm miniplates by biomechanical testing using fibula-reconstructed cadaveric mandibles.Introduction: To increase the clinical relevance, please include data about the benefits of PSI versus plates, referring to some clinical outcomes: recovery time for patients, shorten surgical and hospital time, etc.The aim of the study should be statedThe topology optimization of implants, the new technique tested in this manuscript, is defined as “…a powerful mathematical tool which allows creating an optimal structural design within prescribed loading and boundary conditions.”. This definition was supported by just one reference (#10) in page 4 line 76. This is a key point for this manuscript, so please add a more detailed definition and features of this specific technique to lead medical readers.Conclusions: They should state more precisely the benefits of TOPOS-implants over miniplates rather than “significant superiority”, which is an unprecise and subjective word.[Note: HTML markup is below. Please do not edit.]Reviewers' comments:Reviewer's Responses to QuestionsComments to the Author1. Is the manuscript technically sound, and do the data support the conclusions?The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented.Reviewer #1: YesReviewer #2: Yes**********2. Has the statistical analysis been performed appropriately and rigorously?Reviewer #1: YesReviewer #2: Yes**********3. Have the authors made all data underlying the findings in their manuscript fully available?The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.Reviewer #1: NoReviewer #2: Yes**********4. Is the manuscript presented in an intelligible fashion and written in standard English?PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.Reviewer #1: YesReviewer #2: Yes**********5. Review Comments to the AuthorPlease use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)Reviewer #1: AbstractLine 26: Please mention that “1.0mm” refers to thickness. Define “miniplates” and disclose their titanium grade.Line 29: Thickness and other dimensions of the TOPOS plates should be mentioned.Line 30: Titanium alloy? Grade?In Lines 212-215 you explain that the TOPOS plates fractured under lower static load than the miniplates. Also that the fibula fractured more quickly, but SD probably did not allow to reach statistical significance. Should this not be mentioned?The shape of the implant seems more important than any other characteristics (line 279). This can be mentioned in the abstract.In general, the abstract can reflect the findings in a much better and much understandable way.M&MLine 135: is high stiffness (= high brittleness) correctly the goal? Would reduced plasticity and increased elasticity not be a better criterion? Please explain in your text.Line 164: what do you mean with “surgical procedures”?Line 165: why did you choose for the same (small diameter) screws? Luckily none failed. Is the position of the screws not important? The configuration should be explained and the influence of the latter on bone fracture.Line 176-182: this is difficult to understand. Please rephrase. Did you take laterotrusion (= horizontal shear forces at the occlusal level in the molar area) and protrusion (= vertical shear forces at the incisors) into account?Line 180: name of device, name of company, city, countryLines 212-218 belong to the Results section.Line 230: could Young modulus not be determined per specimen after static and cyclic loading, to test differences between mandibles, being probably a confounding factor?Line 236: did you perform a Shapiro-Wilk test to conclude this?ResultsLine 253: I read “No TOPOS-implant failed for the applied cycles” and then I read in Line 254: “Failure is noted in the TOPOS implant group”, which is confusing.Overall a good paper, but difficult to read for a surgeon. Its impact may increase of you make it more accessible.Reviewer #2: According to the study design by the authors, the research work it is according to be in the right way by the literature available . Could be interesting in the future to use the same design of study to evaluate the properties of PSI vs reconstruction plates. The only thing that I miss was the blind selection of the groups of miniplates vs PSI.**********6. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.If you choose “no”, your identity will remain anonymous but your review may still be made public.Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy.Reviewer #1: Yes: Maurice Yves MommaertsReviewer #2: Yes: Rolando Carrasco[NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.]While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at figures@plos.org. Please note that Supporting Information files do not need this step.16 Apr 20211. Please ensure that your manuscript meets PLOS ONE's style requirements, including those for file naming.In agreement with the PLOS ONE style templates the section names were removed from the first page.2. In the ethics statement in the manuscript and in the online submission form, please provide additional information about the human tissues used in this study. Specifically, please ensure that you have discussed whether next-of-kin provided informed written consent for the use of the tissues. If patients provided informed written consent prior to death to have their bodies used in medical research, please include this information.Additional information about company providing the specimen and the informed written consent was included in the online submission form and the manuscript.3. Please list the name and version of any software package used for statistical analysis, alongside any relevant references. For more information on PLOS ONE's expectations for statistical reporting, please see https://journals.plos.org/plosone/s/submission-guidelines.#loc-statistical-reporting.The name and version of the software for statistical analysis was added to the manuscript.4. Please ensure you have thoroughly discussed any potential limitations of this study within the Discussion section, including the potential impact of confounding factors.Several factors and their influence on the results are identified and discussed in this study. The parameters sex and age were taken into account by distributing them evenly among the implant groups. Bone quality is another limiting factor that comes together with individual biological specimen. It is shown that the quality of the bone varied greatly. In addition, it was investigated whether the mechanical stability, in particular the stiffness, of the reconstructions differed between the implant groups. This was highlighted in more detail in an additional section of the manuscript. No relevant difference between the stiffnesses of the reconstructions could be found in the static testing as well as in the dynamic testing. Thus, this parameter can be excluded as a confounding factor. In addition, for the reconstruction of the mandible, an area on the mandible was identified that is easily weakened by the application of screws. Bone fracture in this area is a prominent failure type in this study. It indicates that for further development of the implant, avoiding the fixation in this area could be beneficial for stability in case of low bone quality.5. We note that you have included the phrase “data not shown” in your manuscript. Unfortunately, this does not meet our data sharing requirements. PLOS does not permit references to inaccessible data. We require that authors provide all relevant data within the paper, Supporting Information files, or in an acceptable, public repository. Please add a citation to support this phrase or upload the data that corresponds with these findings to a stable repository (such as Figshare or Dryad) and provide and URLs, DOIs, or accession numbers that may be used to access these data. Or, if the data are not a core part of the research being presented in your study, we ask that you remove the phrase that refers to these data.The phrase referring to these data was removed from the manuscript.6.Thank you for stating the following in the Financial Disclosure section:"This work was supported by the German Research Foundation (DFG) and the Technical University of Munich (TUM) in the framework of the Open Access Publishing Program.The research project “TOPOS - Development, Manufacturing and Testing of Topology Optimized Osteosynthesis Plates” (AZ-1019-12), in whose context the presented study was conducted, is funded by the Bavarian Research Foundation (BFS).The funders had no role in study design, data collection and analysis, decision to publish, or preparation of the manuscript."We note that one or more of the authors are employed by a commercial company: Josefinum, and private practice for Oral and Maxillofacial Surgery at Pferseeparka) Please provide an amended Funding Statement declaring this commercial affiliation, as well as a statement regarding the Role of Funders in your study. If the funding organization did not play a role in the study design, data collection and analysis, decision to publish, or preparation of the manuscript and only provided financial support in the form of authors' salaries and/or research materials, please review your statements relating to the author contributions, and ensure you have specifically and accurately indicated the role(s) that these authors had in your study. You can update author roles in the Author Contributions section of the online submission form.Please also include the following statement within your amended Funding Statement.“The funder provided support in the form of salaries for authors [insert relevant initials], but did not have any additional role in the study design, data collection and analysis, decision to publish, or preparation of the manuscript. The specific roles of these authors are articulated in the ‘author contributions’ section.”If your commercial affiliation did play a role in your study, please state and explain this role within your updated Funding Statement.Updated Funding Statement:This work was supported by the German Research Foundation (DFG) and the Technical University of Munich (TUM) in the framework of the Open Access Publishing Program.The research project “TOPOS - Development, Manufacturing and Testing of Topology Optimized Osteosynthesis Plates” (AZ-1019-12), in whose context the presented study was conducted, is funded by the Bavarian Research Foundation (BFS).The funders had no role in study design, data collection and analysis, decision to publish, or preparation of the manuscript.The funder Josefinum, and private practice for Oral and Maxillofacial Surgery at Pferseepark provided support in the form of salaries for authors JW, but did not have any additional role in the study design, data collection and analysis, decision to publish, or preparation of the manuscript. The specific roles of these authors are articulated in the ‘author contributions’ section.b) Please also provide an updated Competing Interests Statement declaring this commercial affiliation along with any other relevant declarations relating to employment, consultancy, patents, products in development, or marketed products, etc.Within your Competing Interests Statement, please confirm that this commercial affiliation does not alter your adherence to all PLOS ONE policies on sharing data and materials by including the following statement: "This does not alter our adherence to PLOS ONE policies on sharing data and materials.” (as detailed online in our guide for authors http://journals.plos.org/plosone/s/competing-interests ) . If this adherence statement is not accurate and there are restrictions on sharing of data and/or materials, please state these. Please note that we cannot proceed with consideration of your article until this information has been declared.Please include both an updated Funding Statement and Competing Interests Statement in your cover letter. We will change the online submission form on your behalf.Please know it is PLOS ONE policy for corresponding authors to declare, on behalf of all authors, all potential competing interests for the purposes of transparency. PLOS defines a competing interest as anything that interferes with, or could reasonably be perceived as interfering with, the full and objective presentation, peer review, editorial decision-making, or publication of research or non-research articles submitted to one of the journals. Competing interests can be financial or non-financial, professional, or personal. Competing interests can arise in relationship to an organization or another person. Please follow this link to our website for more details on competing interests: http://journals.plos.org/plosone/s/competing-interests Updated Competing Interests Statement:The authors have declared that no competing interests exist.The affiliation Josefinum, and private practice for Oral and Maxillofacial Surgery at Pferseepark of JW does not alter our adherence to PLOS ONE policies on sharing data and materials.Additional Editor Comments:The authors propose a well-conducted study to compare topology optimized, patient specific osteosynthesis plates (TOPOS-implants) versus 1.0 mm miniplates by biomechanical testing using fibula-reconstructed cadaveric mandibles.Introduction: To increase the clinical relevance, please include data about the benefits of PSI versus plates, referring to some clinical outcomes: recovery time for patients, shorten surgical and hospital time, etc.A corresponding paragraph was added to the Introduction.The aim of the study should be statedA paragraph stating the aim of the study was added to the end of the Introduction.The topology optimization of implants, the new technique tested in this manuscript, is defined as “…a powerful mathematical tool which allows creating an optimal structural design within prescribed loading and boundary conditions.”. This definition was supported by just one reference (#10) in page 4 line 76. This is a key point for this manuscript, so please add a more detailed definition and features of this specific technique to lead medical readers.Thank you very much for pointing this out. More details about topology optimization and the advantages for osteosynthesis plates are added to the introduction.Conclusions: They should state more precisely the benefits of TOPOS-implants over miniplates rather than “significant superiority”, which is an unprecise and subjective word.The sentence was changed to:This study displays a significantly increased cyclic loading capacity of mandible reconstructions with topology optimized and patient specific implants in comparison with miniplates as clinical standardReviewers' comments:Reviewer's Responses to QuestionsComments to the Author1. Is the manuscript technically sound, and do the data support the conclusions?The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented.Reviewer #1: YesReviewer #2: Yes________________________________2. Has the statistical analysis been performed appropriately and rigorously?Reviewer #1: YesReviewer #2: Yes________________________________3. Have the authors made all data underlying the findings in their manuscript fully available?The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.Reviewer #1: NoReviewer #2: Yes________________________________4. Is the manuscript presented in an intelligible fashion and written in standard English?PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.Reviewer #1: YesReviewer #2: Yes________________________________5. Review Comments to the AuthorPlease use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)Reviewer #1: AbstractLine 26: Please mention that “1.0mm” refers to thickness. Define “miniplates” and disclose their titanium grade.Line 29: Thickness and other dimensions of the TOPOS plates should be mentioned.Line 30: Titanium alloy? Grade?The information about the materials of the used plates and the thickness of 1.0 mm was added to the abstract.In Lines 212-215 you explain that the TOPOS plates fractured under lower static load than the miniplates. Also that the fibula fractured more quickly, but SD probably did not allow to reach statistical significance. Should this not be mentioned?We agree that the information about the results from static testing would be interesting for the abstract. But due to the limited amount of words and in order to make the abstract more understandable, we reduced it to the sentence: “Static testing was used to confirm mechanical similarity between the reconstruction groups.”Concerning the results of static testing there must be a misunderstanding, neither TOPOS-implant nor miniplates fractured. The reconstructions were tested in a whole and bone fracture was the main failure type, only once plastic deformation of miniplates occurred. A sentence referencing the result table for the static testing in the supplementary data was added to the manuscript. The differences between the mean values of the groups (elastic limit: 11 N, reconstruction failure: 13 N) are small compared to the SD and the absolute mean value for static testing. This is the reason why we concluded comparable mechanical properties in static testing for the two groups in the abstract.The shape of the implant seems more important than any other characteristics (line 279). This can be mentioned in the abstract.In general, the abstract can reflect the findings in a much better and much understandable way.To further emphasize the importance of the shape of the implants, it was included in the introductory words of the abstract.M&MLine 135: is high stiffness (= high brittleness) correctly the goal? Would reduced plasticity and increased elasticity not be a better criterion? Please explain in your text.Brittleness as a material property often comes together with a high material stiffness. But in this case, the optimization goal was to maintain a high stiffness for the whole implant despite significant material reduction. The Young’s moduli, which are a measure for elasticity and can be seen as the material stiffness, are very similar for titanium grade 4 and grade 5. So, the material properties are not changed and the implant still has a ductile mechanical behavior. The created geometry is optimized to allow as little deformation as possible with the given material. This is criterion is chosen to have a better distribution of internal stress in the implant and to improve fatigue properties. An optimization of the implant to higher elastic deformation would have been possible, but this always comes with a decreased stiffness for the implant. In consequence the reconstruction is not as stable and the risk of interfragmentary motion is higher, which impedes proper bone healing.Line 164: what do you mean with “surgical procedures”?Thank you for pointing that out. The sentence can be misleading to the reader, because the in-vitro reconstruction is not comparable with the intraoperative situation. The sentence was changed to: “The fixation of corresponding bone segments was performed with standard surgical equipment. The position of the screws was chosen for every reconstruction individual by surgeon.”Line 165: why did you choose for the same (small diameter) screws? Luckily none failed. Is the position of the screws not important? The configuration should be explained and the influence of the latter on bone fracture.The same diameter of screws was chosen to have comparable conditions for both reconstruction techniques. With only two length of screws the variational parameter were kept at a minimum, while maintaining the capability to create stable reconstructions. The position of the screws was chosen by the operating surgeon either at planning for TOPOS-implants or during reconstruction for the miniplates. That the position of the screws plays an important role is part of the results of this paper. This is especially important for the mandible angle, where a weakening effect of the screws becomes obvious.Line 176-182: this is difficult to understand. Please rephrase. Did you take laterotrusion (= horizontal shear forces at the occlusal level in the molar area) and protrusion (= vertical shear forces at the incisors) into account?We did a restructuring of the paragraph with additional information added for better understanding. The force direction for the testing setup was derived from the major muscle groups. The in-vivo forces that are resulting from the muscular movement are dependent on the contact situation of the teeth. This includes shear forces at the incisors and the molar area. Due to the fact that we have only a bite point contact at the premolar, which is constraint in vertical direction for the setup, shear forces at the incisors are not present. Horizontal shear forces can occur for the setup at the bite point due to frictional contact. But this was not emphasized for this study. The loading situation was derived from different numerical and experimental models for testing of the mandible. The bearing at the end of the left mandible segment creates the biggest lever arm for the loading, which is like a worst-case scenario for possible chewing loads considering the remaining teeth.Line 180: name of device, name of company, city, countryThank you very much for this hint. The test system has already been described as custom made in a previous paragraph. To avoid confusion for the reader, this has been pointed out again in the manuscript.Lines 212-218 belong to the Results section.The paragraph was moved to the Result section.Line 230: could Young modulus not be determined per specimen after static and cyclic loading, to test differences between mandibles, being probably a confounding factor?The information about the stiffness of the reconstruction (static: stiffness in elastic region, dynamic: actuator displacement) was added to the supplementary information and also mentioned in the Result and Discussion sections. Especially the displacement of the actuators shows that the both reconstruction types show very similar mechanical reaction to the applied load. This eliminates the stiffness of the reconstruction as confounding factor for this study.Line 236: did you perform a Shapiro-Wilk test to conclude this?Yes, due to the fact that the specimen number was too small for a D’Agostino & Pearson omnibus test, a Shapiro-Wilk normality test was performed. This indicated that the results of the dynamic testing were not normally distributed. (Shapiro-Wilk normality test: miniplates group p = 0.002; TOPOS-group p = 0.0008). This information was added to the manuscript.ResultsLine 253: I read “No TOPOS-implant failed for the applied cycles” and then I read in Line 254: “Failure is noted in the TOPOS implant group”, which is confusing.Thank you very much for pointing that out. The second sentence was changed to “…, reconstruction failure (including bone fractures) is noted (…) in the TOPOS implant group…”.Overall a good paper, but difficult to read for a surgeon. Its impact may increase of you make it more accessible.Thank you very much for this hint. We hope the revision made it easier accessible for the readers.Reviewer #2: According to the study design by the authors, the research work it is according to be in the right way by the literature available. Could be interesting in the future to use the same design of study to evaluate the properties of PSI vs reconstruction plates. The only thing that I miss was the blind selection of the groups of miniplates vs PSI.Thank you very much. A comparison with reconstruction plates does sound like an interesting consecutive project.The specimen in this study were first separated in two groups with a similar distribution of the parameters sex and age. This was done so that the result could be considered independent of these factors. Otherwise age and sex could have been confounding factors.Submitted filename: Response to Reviewers.docxClick here for additional data file.6 May 2021PONE-D-20-39788R1Improving mandibular reconstruction by using topology optimization, patient specific design and additive manufacturing? – A biomechanical comparison against miniplates on human specimenPLOS ONEDear Dr. LANG,Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process.==============================Dear authors, thank you for submitting this revised version of the manuscript: "Improving mandibular reconstruction by using topology optimization, patient-specificdesign and additive manufacturing? – A biomechanical comparison against mini plateson human specimen" by Lang et al.As a general comment, the revision of the final revised version (PONE-D-20-39788R1) was complex because it doesn't include changes highlighted in red or yellow color. All changes were presented at the end of the compiled PDF file document as raw tracked changes. TO MAKE THE NEXT REVISION EASY, PLEASE, BE SURE TO HIGHLIGHT FINAL MODIFICATIONS IN RED COLOR ONLY IN THE FINAL REVISED VERSION.Reviewer 1 has 2 comments/suggestions. Please, address them in the next 30 days.==============================Please submit your revised manuscript by June 5th, 2021. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file.Please include the following items when submitting your revised manuscript:A rebuttal letter that responds to each point raised by the academic editor and reviewer(s). You should upload this letter as a separate file labeled 'Response to Reviewers'.A marked-up copy of your manuscript that highlights changes made to the original version. You should upload this as a separate file labeled 'Revised Manuscript with Track Changes'.An unmarked version of your revised paper without tracked changes. You should upload this as a separate file labeled 'Manuscript'.If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter.If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: http://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols. Additionally, PLOS ONE offers an option for publishing peer-reviewed Lab Protocol articles, which describe protocols hosted on protocols.io. Read more information on sharing protocols at https://plos.org/protocols?utm_medium=editorial-email&utm_source=authorletters&utm_campaign=protocols.We look forward to receiving your revised manuscript.Kind regards,Dr. Luis CordovaAcademic EditorPLOS ONEJournal Requirements:Please review your reference list to ensure that it is complete and correct. If you have cited papers that have been retracted, please include the rationale for doing so in the manuscript text, or remove these references and replace them with relevant current references. Any changes to the reference list should be mentioned in the rebuttal letter that accompanies your revised manuscript. If you need to cite a retracted article, indicate the article’s retracted status in the References list and also include a citation and full reference for the retraction notice.Additional Editor Comments (if provided):Dear authors, thank you for submitting this revised version of the manuscript: "Improving mandibular reconstruction by using topology optimization, patient-specificdesign and additive manufacturing? – A biomechanical comparison against mini plateson human specimen" by Lang et al.As a general comment, the revision of the final revised version (PONE-D-20-39788R1) was complex because it doesn't include changes highlighted in red or yellow color. All changes were presented at the end of the compiled PDF file document as raw tracked changes. TO MAKE REVISION EASY, PLEASE, BE SURE TO HIGHLIGHT FINAL MODIFICATIONS IN RED COLOR IN THE THE FINAL REVISED VERSION.Reviewer 1 has two comments/suggestions located at the end of the paragraphs copied below. Please, address them in the next 30 days.Best regardsProf. Luis CordovaLine 135: is high stiffness (= high brittleness) correctly the goal? Would reducedplasticity and increased elasticity not be a better criterion? Please explain in your text.Brittleness as a material property often comes together with a high material stiffness.But in this case, the optimization goal was to maintain a high stiffness for the wholeimplant despite significant material reduction. The Young’s moduli, which are ameasure for elasticity and can be seen as the material stiffness, are very similar fortitanium grade 4 and grade 5. So, the material properties are not changed and theimplant still has a ductile mechanical behavior. The created geometry is optimized toallow as little deformation as possible with the given material. This is criterion is chosento have a better distribution of internal stress in the implant and to improve fatigueproperties. An optimization of the implant to higher elastic deformation would havebeen possible, but this always comes with a decreased stiffness for the implant. Inconsequence the reconstruction is not as stable and the risk of interfragmentary motionis higher, which impedes proper bone healing.Reviewer 1: Then please explain the importance of stress shielding, the micro strains involved related to the Utah paradigm.Line 180: name of device, name of company, city, countryThank you very much for this hint. The test system has already been described ascustom made in a previous paragraph. To avoid confusion for the reader, this has beenpointed out again in the manuscript.Reviewer 1: Difficult to confirm the results by a re-test by a third party. Please describe the flaw.[Note: HTML markup is below. Please do not edit.]Reviewers' comments:[NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.]While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at figures@plos.org. Please note that Supporting Information files do not need this step.21 May 2021Dear Mr. Cordova,dear Mr. Mommaerts,dear Mr. Carrasco,thank you very much für the second revision of our manuscript entitled “Improving mandibular reconstruction by using topology optimization, patient specific design and additive manufacturing? – A biomechanical comparison against miniplates on human specimen”. In the following I would like to respond to each point mentioned in your revision.1. Please review your reference list to ensure that it is complete and correct. If you have cited papers that have been retracted, please include the rationale for doing so in the manuscript text, or remove these references and replace them with relevant current references. Any changes to the reference list should be mentioned in the rebuttal letter that accompanies your revised manuscript. If you need to cite a retracted article, indicate the article’s retracted status in the References list and also include a citation and full reference for the retraction notice.The reference list has been checked. All references are accessible and no retractions are present. During the two revision cycles the following references position have been added to the reference list: 10-18, 35-372. As a general comment, the revision of the final revised version (PONE-D-20-39788R1) was complex because it doesn't include changes highlighted in red or yellow color. All changes were presented at the end of the compiled PDF file document as raw tracked changes. TO MAKE THE NEXT REVISION EASY, PLEASE, BE SURE TO HIGHLIGHT FINAL MODIFICATIONS IN RED COLOR ONLY IN THE FINAL REVISED VERSIONI am sorry to hear about these inconveniences. I did use the normal Track changes function in Word, which includes a highlighted mark-up, as stated in the requirements for submission of revisions. Somehow, in the transition to the compiled pdf there must have been a problem with these markups. Nevertheless, in my pdf-summary of the upload the mark-up was tracked and highlighted normally. I did upload now tracked changes files as pdf in order to prevent this complication. If there is still a problem with the mark-up, please explain me, how exactly to solve this.Comments by Reviewer 13. Reviewer 1: Line 135: is high stiffness (= high brittleness) correctly the goal? Would reduced plasticity and increased elasticity not be a better criterion? Please explain in your text.Brittleness as a material property often comes together with a high material stiffness.But in this case, the optimization goal was to maintain a high stiffness for the whole implant despite significant material reduction. The Young’s moduli, which are a measure for elasticity and can be seen as the material stiffness, are very similar for titanium grade 4 and grade 5. So, the material properties are not changed and the implant still has a ductile mechanical behavior. The created geometry is optimized to allow as little deformation as possible with the given material. This is criterion is chosen to have a better distribution of internal stress in the implant and to improve fatigue properties. An optimization of the implant to higher elastic deformation would have been possible, but this always comes with a decreased stiffness for the implant. In consequence the reconstruction is not as stable and the risk of interfragmentary motion is higher, which impedes proper bone healing.Reviewer 1: Then please explain the importance of stress shielding, the micro strains involved related to the Utah paradigm..Thank you very much for pointing this out. To emphasize for the reader, that stress shielding and bone loss due to reduced micro strains is not an issue for these implants, two sections were added to the manuscript. First, the optimization process was rephrased to show that an optimization goal for high stiffness does not result in an extraordinary stiffness for the implant. Secondly it was highlighted that the mechanical similarity of the reconstruction does also mean similar mechanical stresses for the fracture area.The following sentence was added to the Material&Methods section for clarification:“… In other words, the optimization is a material reduction process that results in the least possible loss of stiffness for the implant.”The topic concerning stress shielding and micromotions in the fracture is now addressed with the following paragraph in the discussion section:“The mechanical similarity also confirms that there is no excessive stiffening caused by the TOPOS implants. If the bone fragments are positioned too rigidly for osteosynthesis, it can result in inhibition of sufficient micromotions in the fracture gap [35,36]. Stress shielding like effects can lead to impaired healing in this context due to low interfragmentary strains at the fracture site [37]. But as already described there is no indication for this found in mechanical testing. “4. Reviewer 1: Line 180: name of device, name of company, city, countryThank you very much for this hint. The test system has already been described as custom made in a previous paragraph. To avoid confusion for the reader, this has been pointed out again in the manuscript.Reviewer 1: Difficult to confirm the results by a re-test by a third party. Please describe the flaw.The following sentence was added to the manuscript for clarification:“The biomechanical testing in this study was performed on an in-house developed testing setup. Using a custom-made solution like this comes with the disadvantage of limited reproducibility for third parties. But a commercially available solution does not exist for this testing scenario. And due to the detailed description of the testing and the publication presenting the experimental setup by Foehr et al. [26], distinct transparency, comprehensibility and adaptability of the experiments are ensured.”We hope the revised version of our article meets all the conditions for publication in your journal.Submitted filename: Response to Reviewers.docxClick here for additional data file.27 May 2021Improving mandibular reconstruction by using topology optimization, patient specific design and additive manufacturing? – A biomechanical comparison against miniplates on human specimenPONE-D-20-39788R2Dear Dr. Lang,We’re pleased to inform you that your manuscript has been judged scientifically suitable for publication and will be formally accepted for publication once it meets all outstanding technical requirements.Within one week, you’ll receive an e-mail detailing the required amendments. When these have been addressed, you’ll receive a formal acceptance letter and your manuscript will be scheduled for publication.An invoice for payment will follow shortly after the formal acceptance. To ensure an efficient process, please log into Editorial Manager at http://www.editorialmanager.com/pone/, click the 'Update My Information' link at the top of the page, and double check that your user information is up-to-date. If you have any billing related questions, please contact our Author Billing department directly at authorbilling@plos.org.If your institution or institutions have a press office, please notify them about your upcoming paper to help maximize its impact. If they’ll be preparing press materials, please inform our press team as soon as possible -- no later than 48 hours after receiving the formal acceptance. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information, please contact onepress@plos.org.Kind regards,Dr. Luis CordovaAcademic EditorPLOS ONEAdditional Editor Comments (optional):Reviewers' comments:31 May 2021PONE-D-20-39788R2Improving mandibular reconstruction by using topology optimization, patient specific design and additive manufacturing? – A biomechanical comparison against miniplates on human specimenDear Dr. Lang:I'm pleased to inform you that your manuscript has been deemed suitable for publication in PLOS ONE. Congratulations! Your manuscript is now with our production department.If your institution or institutions have a press office, please let them know about your upcoming paper now to help maximize its impact. If they'll be preparing press materials, please inform our press team within the next 48 hours. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information please contact onepress@plos.org.If we can help with anything else, please email us at plosone@plos.org.Thank you for submitting your work to PLOS ONE and supporting open access.Kind regards,PLOS ONE Editorial Office Staffon behalf ofDr. Luis CordovaAcademic EditorPLOS ONE
Authors: Florian Andreas Probst; Marc Metzger; Michael Ehrenfeld; Carl-Peter Cornelius Journal: J Oral Maxillofac Surg Date: 2016-03-24 Impact factor: 1.895
Authors: Steven M Roser; Srinivasa Ramachandra; Henry Blair; William Grist; Grant W Carlson; Andrew M Christensen; Katherine A Weimer; Martin B Steed Journal: J Oral Maxillofac Surg Date: 2010-09-09 Impact factor: 1.895
Authors: Ashley B Robey; Michael L Spann; Timothy M McAuliff; Jane L Meza; Ronald R Hollins; Perry J Johnson Journal: Plast Reconstr Surg Date: 2008-12 Impact factor: 4.730
Authors: D D Jewer; J B Boyd; R T Manktelow; R M Zuker; I B Rosen; P J Gullane; L E Rotstein; J E Freeman Journal: Plast Reconstr Surg Date: 1989-09 Impact factor: 4.730