| Literature DB >> 33891228 |
Li-Bin Shi1,2, Yu-Chuan Wang2,3, Song-Yun Chu2,3, Alessandro De Bortoli4, Peter Schuster1,2, Eivind Solheim2, Jian Chen5,6.
Abstract
BACKGROUND: This study aimed to clarify the interrelationship and additive effects of contact force (CF), power and application time in both conventional and high-power short-duration (HPSD) settings.Entities:
Keywords: Ablation index; Contact force; High-power short-duration; Power; Pulmonary vein isolation
Mesh:
Year: 2021 PMID: 33891228 PMCID: PMC9399015 DOI: 10.1007/s10840-021-00990-4
Source DB: PubMed Journal: J Interv Card Electrophysiol ISSN: 1383-875X Impact factor: 1.759
Number of applications at different levels of power and contact force and distribution of points reaching an impedance drop of 10 Ω (number and percentage in parenthesis)
| Mean contact force | ||||||
|---|---|---|---|---|---|---|
| CF < 5 g | CF 5–10 g | CF 10–20 g | CF ≥ 20 g | Total | ||
| 25 W | 31 (7, 22.6%) | 53 (25, 47.2%) | 44 (35, 79.5%) | 13 (13, 100%) | 141 (80, 56.7%) | |
| Power | 30 W | 16 (5, 31.3%) | 55 (42, 76.4%) | 55 (49, 89.1%) | 14 (14, 100%) | 140 (110, 78.6%) |
| 35 W | 22 (10, 45.5%) | 60 (45, 75.0%) | 44 (44, 100%) | 12 (12, 100%) | 138 (111, 80.4%) | |
| 50 W | 16 (2, 12.5%) | 157 (78, 49.7%) | 157 (114, 72.6%) | 38 (32, 84.2%) | 368 (226, 61.4%) | |
| Total | 85 (24, 28.2%) | 325 (190, 58.5%) | 300 (242, 80.7%) | 77 (71, 92.2%) | 787 (527, 67.0%) | |
CF contact force
Fig. 1Additive impacts of contact force, power and application time on impedance drop. With CF < 5 g, ID does not reach the threshold of 10 Ω within 60 s at any power setting. When CF ≥ 5 g, power and application time compensate for each other within restricted ranges. Compared with the conventional sub-groups at different time points, ID levels at 10 s of HPSD ablation lie between those of the 25 and 30W sub-groups under the same CF level. ID, impedance drop; CF, contact force; HPSD, high power short duration; †P < 0.05 compared to 25 W, ‡P < 0.01 compared to 25 W, *P < 0.01 compared to 50W
Time (in seconds) to reach impedance drop of 10 Ω and to the maximal impedance drop (in parentheses)
| Mean contact force | |||||
|---|---|---|---|---|---|
| CF < 5 g | CF 5–10g | CF 10–20 g | CF ≥ 20 g | ||
| Power | 25 W | ― | ― | 40 (60) | 10 (40) |
| 30 W | ― | 20 (50) | 10 (50) | 10 (20) | |
| 35 W | ― | 10 (20) | 10 (20) | 10 (20) | |
| 50 W | ― | 10a | 10a | 10a | |
CF contact force
aTotal application time 10 s
Fig. 2Correlation between impedance drop and ablation index under different power settings. Changes of ID with increasing AI values are similar at settings of 30, 35 and 50 W. At 25 W, they show the same trend, but with lower ID at the same AI level (P < 0.01)