Literature DB >> 24798957

Contact force and force-time integral in atrial radiofrequency ablation predict transmurality of lesions.

Fabien Squara1, Decebal Gabriel Latcu, Youssef Massaad, Marouane Mahjoub, Sok-Sithikun Bun, Nadir Saoudi.   

Abstract

AIMS: During radiofrequency (RF) delivery, lesion volume is highly dependent on contact force (CF). It has recently been shown that changes of bipolar electrogram (EGM) predict transmurality. We hypothesized that there is a correlation between CF and EGM criteria of transmural lesion (TL) during RF. METHODS AND
RESULTS: We prospectively studied consecutive 512 RF applications from atrial fibrillation ablation procedures. A force-sensing ablation catheter (Tacticath(®), Endosense) was used to continuously measure CF and force-time integral (FTI) during each RF application. Distal bipolar EGM was analysed before, during, and after each RF application. Depending on initial EGM morphology, transmurality of lesions was defined by: (i) disappearance of the positivity after RF when there was QR morphology, (ii) diminution >75% of the positivity when there was QRS morphology, or (iii) disappearance of the R' positivity when there was RSR' morphology. Electrogram criteria were validated by electrophysiologists blinded to force measurements. Force-time integral was higher in TL than in non-transmural lesions (NTLs): 652 ± 248 vs. 212 ± 140 gs (P < 0.001). Mean CF per RF pulse was higher in TL than in NTL: 26.3 ± 12.5 vs. 11.3 ± 10.3 g (P < 0.001). The best cut-off to predict TL was an FTI ≥ 392 gs [sensitivity 0.89, specificity (Sp) 0.93, positive predictive value (PPV) 0.98, and negative predictive value 0.67] and a higher FTI (>700 gs) warrants transmurality of RF atrial lesions (100% Sp and PPV).
CONCLUSION: Contact force and FTI during RF are correlated with TL. During RF delivery, a target FTI > 392 gs can be used as an endpoint.

Entities:  

Keywords:  Ablation; Atrial fibrillation; Contact force; Force-time integral; Transmural lesion

Mesh:

Year:  2014        PMID: 24798957     DOI: 10.1093/europace/euu068

Source DB:  PubMed          Journal:  Europace        ISSN: 1099-5129            Impact factor:   5.214


  40 in total

1.  General anesthesia improves contact force and reduces gap formation in pulmonary vein isolation: a comparison with conscious sedation.

Authors:  Akio Chikata; Takeshi Kato; Takanori Yaegashi; Satoru Sakagami; Chieko Kato; Takahiro Saeki; Keiichi Kawai; Shin-Ichiro Takashima; Hisayoshi Murai; Soichiro Usui; Hiroshi Furusho; Shuichi Kaneko; Masayuki Takamura
Journal:  Heart Vessels       Date:  2017-03-04       Impact factor: 2.037

2.  Contact force monitoring during catheter ablation of intraatrial reentrant tachycardia in patients with congenital heart disease.

Authors:  Ulrich Krause; David Backhoff; Sophia Klehs; Heike E Schneider; Thomas Paul
Journal:  J Interv Card Electrophysiol       Date:  2016-01-07       Impact factor: 1.900

3.  Clinical utility of automated ablation lesion tagging based on catheter stability information (VisiTag Module of the CARTO 3 System) with contact force-time integral during pulmonary vein isolation for atrial fibrillation.

Authors:  Yasuo Okumura; Ichiro Watanabe; Kazuki Iso; Koichi Nagashima; Kazumasa Sonoda; Naoko Sasaki; Rikitake Kogawa; Keiko Takahashi; Kimie Ohkubo; Toshiko Nakai; Shiro Nakahara; Yuuichi Hori; Atsushi Hirayama
Journal:  J Interv Card Electrophysiol       Date:  2016-06-09       Impact factor: 1.900

4.  Electrogram characteristics of the coronary sinus in cases requiring epicardial ablation within the coronary sinus for creating a conduction block at the left lateral mitral isthmus.

Authors:  Daisuke Sato; Hiroki Mani; Yu Makihara; Hiroki Kitajima; Yuji Nishikawa; Seno Keitaro; Yeong-Hwa Chun
Journal:  J Interv Card Electrophysiol       Date:  2018-06-27       Impact factor: 1.900

Review 5.  Recent Advances in Lesion Formation for Catheter Ablation of Atrial Fibrillation.

Authors:  Adam S Barnett; Tristram D Bahnson; Jonathan P Piccini
Journal:  Circ Arrhythm Electrophysiol       Date:  2016-05

Review 6.  [Catheter ablation of cardiac arrhythmias : Forms of energy and biophysical principles].

Authors:  Thomas Beiert; Jan W Schrickel
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2019-11-06

7.  Clinical significance of induced left atrial macro-reentrant tachycardia after pulmonary vein isolation.

Authors:  Hideyuki Hara; Masahiro Yoshinaga; Yumie Matsui; Satoshi Yamamoto; Takahiro Ishido; Kotaro Yutaka; Tomonori Kasuu; Masahiro Karakawa
Journal:  J Interv Card Electrophysiol       Date:  2015-09-21       Impact factor: 1.900

Review 8.  Atrial fibrillation ablation and left appendage closure in heart failure patients.

Authors:  Minesh R Patel; Angelo B Biviano
Journal:  Curr Opin Cardiol       Date:  2015-05       Impact factor: 2.161

Review 9.  Better Lesion Creation And Assessment During Catheter Ablation.

Authors:  Saurabh Kumar; Chirag R Barbhaiya; Samuel Balindger; Roy M John; Laurence M Epstein; Bruce A Koplan; Usha B Tedrow; William G Stevenson; Gregory F Michaud
Journal:  J Atr Fibrillation       Date:  2015-10-31

Review 10.  Improvements In AF Ablation Outcome Will Be Based More On Technological Advancement Versus Mechanistic Understanding.

Authors:  Chen-Yang Jiang Md; Ru-Hong Jiang Ms
Journal:  J Atr Fibrillation       Date:  2014-08-31
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.