Impedance monitoring during radiofrequency catheter ablation in humans.

M Harvey1, Y N Kim, J Sousa, R el-Atassi, F Morady, H Calkins, J J Langberg.   

Abstract

Radiofrequency catheter ablation of accessory pathways and the atrioventricular junction often requires multiple applications of energy. The inability to determine the effects of any given application on the underlying tissue may contribute to this problem. In the present study, impedance was monitored in 20 patients undergoing radiofrequency catheter ablation, and the relationship between an initial decrease in impedance and subsequent effects were examined. An initial fall in impedance of more than 10 omega was 78% sensitive and 88% specific for predicting subsequent evidence of tissue heating (interruption of conduction or an abrupt rise in impedance due to coagulum formation). In contrast, initial values of voltage, current, or impedance did not distinguish between effective and ineffective applications of radiofrequency energy. Continuous monitoring of impedance may facilitate radiofrequency catheter ablation.

Entities:  

Mesh:

Year:  1992        PMID: 1370996     DOI: 10.1111/j.1540-8159.1992.tb02897.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  4 in total

1.  [Guidelines for catheter ablation].

Authors:  Karl-Heinz Kuck; Sabine Ernst; Uwe Dorwarth; Ellen Hoffmann; Heinz Pitschner; Jürgen Tebbenjohanns; Hans Kottkamp
Journal:  Clin Res Cardiol       Date:  2007-11       Impact factor: 5.460

2.  Characteristics of Very High-Power, Short-Duration Radiofrequency Applications.

Authors:  Gábor Orbán; Zoltán Salló; Péter Perge; Pál Ábrahám; Katalin Piros; Klaudia Vivien Nagy; István Osztheimer; Béla Merkely; László Gellér; Nándor Szegedi
Journal:  Front Cardiovasc Med       Date:  2022-07-13

3.  A computational comparison of radiofrequency and pulsed field ablation in terms of lesion morphology in the cardiac chamber.

Authors:  Mario Gómez-Barea; Tomás García-Sánchez; Antoni Ivorra
Journal:  Sci Rep       Date:  2022-09-27       Impact factor: 4.996

4.  The impacts of contact force, power and application time on ablation effect indicated by serial measurements of impedance drop in both conventional and high-power short-duration ablation settings of atrial fibrillation.

Authors:  Li-Bin Shi; Yu-Chuan Wang; Song-Yun Chu; Alessandro De Bortoli; Peter Schuster; Eivind Solheim; Jian Chen
Journal:  J Interv Card Electrophysiol       Date:  2021-04-23       Impact factor: 1.759

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.