Daniel Stocker1, Andrei Manoliu1,2,3,4, Anton S Becker1, Borna K Barth1, Daniel Nanz1,5, Markus Klarhöfer6, Olivio F Donati1.
Abstract
PURPOSE: To evaluate the influence of body phased-array (BPA) receive coil setups on signal-to-noise ratio (SNR) and image quality (IQ) in prostate MRI.Entities:
Keywords: 18CH, BPA 18-channel body array coil; 30CH, BPA 30-channel body array coil; 60CH, BPA 60-channel body array coil; ANOVA, Analysis of variances; BPA, Body phased-array; ERC, Endorectal coil; ICC, Intra-class correlation coefficient; IQR, Interquartile range; Magnetic resonance imaging; PSTT, Post-hoc paired-sample t-tests; Prostate imaging; ROIs, Region of interests; SD, Standard deviation; SNR, Signal to noise ratio; Signal-to-noise ratio; T2W-TSE, T2-weighted turbo spin echo; mpMRI, Multi-parametric magnetic resonance imaging; ss-DWI-EPI, Single-shot diffusion-weighting spin-echo echo-planar imaging
Year: 2021 PMID: 33644263 PMCID: PMC7889823 DOI: 10.1016/j.ejro.2021.100327
Source DB: PubMed Journal: Eur J Radiol Open ISSN: 2352-0477
Fig. 1Voxel-wise SNR maps for a representative participant. T2W-TSE and DWI images were postprocessed individually together with the corresponding noise data, yielding SNR maps on a voxel-wise basis. SNR maps are shown for T2W-TSE / DWI sequences using 18CH, 30CH as well as 60CH coils, respectively. SNR values are color-coded from 5 (black) to 40 (yellow/white). (For interpretation of the references to colour in this figure legend, the reader is referred to the web version of this article).
Abbreviations: SNR, signal-to-noise ratio; T2W-TSE, T2-weighted turbo spin echo; DWI, diffusion weighted images; 18CH, 18-channel body phased-array receive coil setup; 30CH, 30-channel body phased-array receive coil setup; 60CH, 60-channel body phased-array receive coil setup.
Signal-to-noise ratio analysis. To evaluate potential differences between pairs of phased array receive coil setups in SNR within the peripheral and transition zone of the prostate for T2-weighted turbo spin echo and diffusion weighted images, post-hoc paired-sample t-tests (PSTT) were performed. After correction for multiple comparisons (Benjamini-Hochberg procedure, n = 12 to account for all performed PSTT) a p-value <0.004 was considered statistically significant. Italics indicate p-values <0.004.
| Peripheral Zone | Transition Zone | ||||||
|---|---|---|---|---|---|---|---|
| SNR | T-value | p-value | SNR | T-value | p-value | ||
| 18.16 ± 6.04 vs. 22.45 ± 4.52 | −2.182 | 0.061 | 7.68 ± 2.37 vs. 11.54 ± 4.44 | −3.623 | 0.007 | ||
| 18.16 ± 6.04 vs. 24.61 ± 5.62 | −1.983 | 0.083 | 7.68 ± 2.37 vs. 15.20 ± 4.22 | −4.965 | |||
| 22.45 ± 4.52 vs. 24.61 ± 5.62 | −1.089 | 0.308 | 11.54 ± 4.44 vs. 15.20 ± 4.22 | −2.166 | 0.062 | ||
| 17.32 ± 4.24 vs. 16.09 ± 5.95 | 0.586 | 0.574 | 11.43 ± 2.44 vs. 10.47 ± 2.02 | 0.940 | 0.375 | ||
| 17.32 ± 4.24 vs. 30.24 ± 15.45 | −2.626 | 0.030 | 11.43 ± 2.44 vs. 15.19 ± 4.32 | −2.434 | 0.041 | ||
| 16.09 ± 5.95 vs. 30.24 ± 15.45 | −2.782 | 0.024 | 10.47 ± 2.02 vs. 15.19 ± 4.32 | −2.770 | 0.024 | ||
Abbreviations: SNR, signal-to-noise ratio; T2W-TSE, T2-weighted turbo spin echo; DWI, diffusion weighted images; 18CH, 18-channel body phased-array receive coil setup; 30CH, 30-channel body phased-array receive coil setup; 60CH, 60-channel body phased-array receive coil setup.
Fig. 2Boxplots for ADC measurements in the peripheral and transition zone. ADC measurements in the peripheral and transition zone were similar between the 18CH BPA, 30CH BPA and 60CH BPA receive coil setup without significant differences between the coil setups (ANOVA peripheral zone F = 0.594, p = 0.55; transition zone F = 1.790, p = 0.186).
Abbreviation: ADC, apparent diffusion coefficient; 18CH BPA, 18-channel body phased-array receive coil setup; 30CH BPA, 30-channel body phased-array receive coil setup; 60CH BPA, 60-channel body phased-array receive coil setup; ANOVA, Analysis of variances.
Qualitative results of conventional reading (Likert score ± standard deviation), number of artifacts, diagnostic sequences and results from Wilcoxon signed rank test. After correction for multiple comparisons (Bonferroni correction, n = 3 to account for all body phased-array receive coil setups) a p-value <0.017 was considered statistically significant.
| T2W-TSE | p-value | DWI | p-value | ||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| 18CH | 30CH | 60CH | 18CH vs. 30CH | 18CH vs. 60CH | 30CH vs. 60CH | 18CH | 30CH | 60CH | 18CH vs. 30CH | 18CH vs. 60CH | 30CH vs. 60CH | ||||||
| Resolution | R1 | 4.3 ± 0.7 | 3.9 ± 1.1 | 4.1 ± 0.7 | 0.102 | 0.414 | 0.527 | 4.1 ± 0.7 | 4.2 ± 0.6 | 4.2 ± 0.6 | 0.564 | 0.655 | 1 | ||||
| R2 | 4.6 ± 0.7 | 4.4 ± 0.8 | 4.7 ± 0.5 | 0.414 | 0.655 | 0.317 | 4.4 ± 1.0 | 4.1 ± 0.9 | 4.3 ± 0.7 | 0.180 | 0.564 | 0.414 | |||||
| Capsule demarcation | R1 | 4.8 ± 0.4 | 4.4 ± 0.8 | 4.8 ± 0.4 | 0.102 | 1 | 0.102 | 4.7 ± 0.5 | 4.4 ± 1.0 | 4.5 ± 0.7 | 0.180 | 0.157 | 0.564 | ||||
| R2 | 4.8 ± 0.4 | 4.6 ± 0.5 | 4.9 ± 0.3 | 0.157 | 0.564 | 0.083 | 4.8 ± 0.6 | 4.2 ± 0.9 | 4.6 ± 1.0 | 0.063 | 0.157 | 0.234 | |||||
| Zonal anatomy | R1 | 4.0 ± 0.8 | 3.9 ± 1.1 | 4.2 ± 0.8 | 0.705 | 0.480 | 0.257 | 4.2 ± 0.8 | 3.9 ± 0.7 | 3.8 ± 0.8 | 0.257 | 0.046 | 0.705 | ||||
| R2 | 4.4 ± 0.7 | 4.6 ± 0.5 | 4.8 ± 0.4 | 0.317 | 0.102 | 0.480 | 4.6 ± 1.0 | 4.0 ± 1.2 | 4.3 ± 1.1 | 0.034 | 0.180 | 0.180 | |||||
| Overall IQ | R1 | 4.3 ± 0.7 | 4.3 ± 1.1 | 4.4 ± 0.7 | 1 | 0.564 | 0.655 | 4.2 ± 0.8 | 4.1 ± 0.7 | 3.9 ± 0.9 | 0.655 | 0.180 | 0.480 | ||||
| R2 | 4.6 ± 0.7 | 4.3 ± 0.8 | 4.7 ± 0.5 | 0.180 | 0.655 | 0.102 | 4.4 ± 0.8 | 4.3 ± 0.7 | 4.5 ± 0.7 | 0.705 | 0.739 | 0.157 | |||||
| Periurethral clarity | R1 | 4.4 ± 0.7 | 4.3 ± 1.1 | 4.5 ± 0.7 | 0.739 | 0.655 | 0.317 | ||||||||||
| R2 | 4.7 ± 0.5 | 4.7 ± 0.7 | 4.9 ± 0.3 | 1 | 0.317 | 0.414 | |||||||||||
| Geometric distortion | R1 | 1.6 ± 1.0 | 2.0 ± 1.1 | 1.7 ± 1.1 | 0.194 | 0.705 | 0.334 | ||||||||||
| R2 | 1.7 ± 0.9 | 1.7 ± 0.9 | 2.3 ± 1.6 | 1 | 0.131 | 0.206 | |||||||||||
| Number of artifacts | R1 | 5 | 6 | 8 | 0.564 | 0.180 | 0.317 | ||||||||||
| R2 | 4 | 5 | 5 | 0.655 | 0.655 | 1 | |||||||||||
| Artifacts influencing diagnostic evaluation | R1 | 1.3 ± 0.9 | 1.6 ± 1.1 | 1.4 ± 1.0 | 0.414 | 0.317 | 0.705 | ||||||||||
| R2 | 1.2 ± 0.6 | 1.5 ± 1.1 | 1.3 ± 0.9 | 0.180 | 0.317 | 0.317 | |||||||||||
| Artifacts influencing IQ | R1 | 1.8 ± 1.2 | 2.2 ± 1.3 | 2 ± 0.8 | 0.336 | 0.414 | 0.739 | ||||||||||
| R2 | 1.6 ± 1.0 | 1.8 ± 1.0 | 1.7 ± 1.1 | 0.414 | 0.705 | 0.739 | |||||||||||
| Diagnostic (yes/no) | R1 | 10/0 | 9/1 | 10/0 | 0.317 | 1 | 0.317 | 9/1 | 9/1 | 9/1 | 1 | 1 | 1 | ||||
| R2 | 10/0 | 10/0 | 10/0 | 1 | 1 | 1 | 10/0 | 10/0 | 10/0 | 1 | 1 | 1 | |||||
Abbreviations: T2W-TSE, T2-weighted turbo spin echo; DWI, diffusion weighted images; 18CH, 18-channel body phased-array receive coil setup; 30CH, 30-channel body phased-array receive coil setup; 60CH, 60-channel body phased-array receive coil setup; R1, reader 1; R2, reader 2; IQ, image quality.
Fig. 3Exemplary images acquired with three different coil setups in the same volunteer side-by-side. Images of the prostate at the midgland level of the T2W-TSE images from the 18CH, 30CH and 60CH coil setup and the DWI are shown. Qualitative image features for the different coil setups were rated with similar subjective scores for T2W-TSE and DWI. The overall image quality for the 18CH, 30CH and 60CH coil setup was rated 5, 5 and 5 for T2W-TSE images and 5, 4 and 4 for DWI by both readers. All T2W-TSE and DWI sequences were rated diagnostic by both readers.
Abbreviations: T2W-TSE, T2-weighted turbo spin echo; DWI, diffusion weighted images; 18CH, 18-channel body phased-array receive coil setup; 30CH, 30-channel body phased-array receive coil setup; 60CH, 60-channel body phased-array receive coil setup.