Wenjun Kou1, Dustin A Carlson2, Neelesh A Patankar3, Peter J Kahrilas2, John E Pandolfino2. 1. Feinberg School of medicine, Northwestern University, 676 North Saint Clair Street, Chicago, IL 60611, USA. 2. Feinberg School of Medicine, Northwestern University, Chicago, IL, USA. 3. Department of Mechanical Engineering, Northwestern University, Evanston, IL, USA.
Abstract
BACKGROUND: This study aimed to introduce a novel analysis paradigm, referred to as 4-dimensional (4D) manometry based on biophysical analysis; 4D manometry enables the visualization of luminal geometry of the esophagus and esophagogastric junction (EGJ) using high-resolution-impedance-manometry (HRIM) data. METHODS: HRIM studies from two asymptomatic controls and one type-I achalasia patient were analyzed. Concomitant fluoroscopy images from one control subject were used to validate the calculated temporal-spatial luminal radius and time-history of intraluminal bolus volume and movement. EGJ analysis computed diameter threshold for emptying, emptying time, flow rate, and distensibility index (DI), which were compared with bolus flow time (BFT) analysis. RESULTS: For normal control, calculated volumes for 5 ml swallows were 4.1 ml-6.7 ml; for 30 ml swallows 21.3 ml-21.8 ml. With type-I achalasia, >4 ml of intraesophageal bolus residual was present both pre- and post-swallow. The four phases of bolus transit were clearly illustrated on the time-history of bolus movement, correlating well with the fluoroscopic images. In the control subjects, the EGJ diameter threshold for emptying was 8 mm for 5 ml swallows and 10 mm for 30 ml swallows; emptying time was 1.2-2.2 s for 5 ml swallows (BFT was 0.3-3 s) and 3.25-3.75 s for 30 ml swallows; DI was 2.4-3.4 mm2/mmHg for 5 ml swallows and 4.2-4.6 mm2/mmHg for 30 ml swallows. CONCLUSIONS: The 4D manometry system facilitates a comprehensive characterization of dynamic esophageal bolus transit with concurrent luminal morphology and pressure from conventional HRIM measurements. Calculations of flow rate and wall distensibility provide novel measures of EGJ functionality.
BACKGROUND: This study aimed to introduce a novel analysis paradigm, referred to as 4-dimensional (4D) manometry based on biophysical analysis; 4D manometry enables the visualization of luminal geometry of the esophagus and esophagogastric junction (EGJ) using high-resolution-impedance-manometry (HRIM) data. METHODS: HRIM studies from two asymptomatic controls and one type-I achalasia patient were analyzed. Concomitant fluoroscopy images from one control subject were used to validate the calculated temporal-spatial luminal radius and time-history of intraluminal bolus volume and movement. EGJ analysis computed diameter threshold for emptying, emptying time, flow rate, and distensibility index (DI), which were compared with bolus flow time (BFT) analysis. RESULTS: For normal control, calculated volumes for 5 ml swallows were 4.1 ml-6.7 ml; for 30 ml swallows 21.3 ml-21.8 ml. With type-I achalasia, >4 ml of intraesophageal bolus residual was present both pre- and post-swallow. The four phases of bolus transit were clearly illustrated on the time-history of bolus movement, correlating well with the fluoroscopic images. In the control subjects, the EGJ diameter threshold for emptying was 8 mm for 5 ml swallows and 10 mm for 30 ml swallows; emptying time was 1.2-2.2 s for 5 ml swallows (BFT was 0.3-3 s) and 3.25-3.75 s for 30 ml swallows; DI was 2.4-3.4 mm2/mmHg for 5 ml swallows and 4.2-4.6 mm2/mmHg for 30 ml swallows. CONCLUSIONS: The 4D manometry system facilitates a comprehensive characterization of dynamic esophageal bolus transit with concurrent luminal morphology and pressure from conventional HRIM measurements. Calculations of flow rate and wall distensibility provide novel measures of EGJ functionality.
The primary function of the esophagus is to transport ingested food from the mouth to
the stomach utilizing delicately balanced neuromuscular control and biomechanical
tissue properties. This manifests as temporal-spatial variation of esophageal wall
morphology, stress, and luminal pressure during bolus transit.[1-3] Consequently, extracting the
details of esophageal pressurization with concurrent luminal morphology is crucial
for the understanding and management of esophageal dysfunction. Currently, the
dominant technology utilized in the assessment of esophageal motility is
high-resolution manometry (HRM) along with the Chicago Classification,[4] but this falls short in assessing the corresponding aberrations in esophageal
morphology. Hence, intraluminal ultrasound, concomitant fluoroscopy, computed
tomography, and magnetic resonance imaging have been utilized experimentally to
enhance the assessment of esophageal luminal geometry.[5-9] However, for technical and
practical reasons, none of these have been applied widely.High-resolution impedance manometry (HRIM) offers an alternative strategy to
simultaneously monitor intraluminal pressure and luminal geometry. The functional
luminal imaging probe (FLIP) dynamically monitors luminal geometry within a
cylindrical bag of conducting fluid but provides limited intraluminal pressure
resolution.[10,11] HRIM integrates high-resolution pressure and impedance
measurement into a single catheter but, thus far, has suffered from a very limited,
dichotomous analysis of bolus present/bolus absent assessment with respect to the
impedance data.[12,13] However, HRIM, could offer truly concurrent temporal-spatial
dynamics of luminal geometry and pressure with more sophisticated analysis
techniques. In this work, we propose such a biophysical analysis technique, referred
to as 4-dimensional (4D) manometry assessing HRIM data dynamically along the
dimensions of pressure, cross-sectional area, esophageal length, and time. We
hypothesize that the proposed 4D manometry analytic technique could extract
comprehensive details of esophageal morphology and pressurization and enable
subsequent analysis of bolus flow rate, flow velocity, and distensibility index (DI)
of the esophageal-gastric junction (EGJ) during esophageal emptying using only
conventional HRIM data. Thus, the goal of the current study was to introduce 4D
manometry methodology and evaluate its efficacy in extracting biophysical
information. Also included was a preliminary validation of this technique.
Materials and methods
Subjects
The 4D manometry analysis paradigm was developed and tested by comparison with
concurrent fluoroscopic imaging, bolus flow time (BFT) analysis, and
distensibility analysis obtained from a FLIP study. Three cases were analyzed:
(1) an asymptomatic control with concurrent HRIM and fluoroscopy studied in our
previous report referred to as Control 1[6]; (2) a second asymptomatic control studied with HRIM and FLIP varying the
bolus volume and constituents during the HRIM study, referred to as
Control 2; and (3) a type-I achalasia patient studied with
HRIM and FLIP, referred to Type-I Ach.The study protocol was approved by the Northwestern University Institutional
Review Board (STU00096856, approved November 2014). Informed consent was
obtained from all subjects in a written format; control subjects were paid for
their participation.
Study protocol
The HRIM catheter utilized was a 4.2-mm outer diameter solid-state assembly with
36 circumferential pressure sensors at 1-cm interval, and 18 impedance segments
at 2-cm intervals (Given Imaging, Los Angeles, CA, USA). Transducers were
calibrated at 0–300 mmHg using externally applied pressure. The assembly was
placed transnasally and positioned to record from the hypopharynx to the stomach
with about three intragastric pressure sensors. HRIM data and fluoroscopic
imaging were acquired simultaneously using the HRIM video system (model A400,
Given Imaging), which synchronized the fluoroscopic images with manometric data
during acquisition. Fluoroscopy was performed with a multipurpose X-ray system
(Artis MP, Siemens, Malvern, PA, USA). This information was then displaced on a
computer screen in real time and stored on a hard drive for further analysis.
The HRIM protocol included a first 5-min baseline recording and then 20-s to
30-s interval for each swallow. FLIP studies were done in conjunction with
endoscopy. The 16-cm FLIP (EndoFLIP® EF-322N; Medtronic, Inc., Shoreview, MN,
USA) was calibrated to atmospheric pressure prior to trans-oral probe placement.
With the endoscope withdrawn, the FLIP was positioned within the esophagus such
that 1–3 impedance sensors were observed beyond the EGJ with this positioning
maintained throughout the FLIP study. Stepwise, 10-ml balloon distensions
beginning with 20 ml and increasing to target volume of 70 ml were then
performed; each stepwise distension volume was maintained for 30–60 s.For Control 1, concurrent HRIM and videofluoroscopy were done during six barium
swallows (5 ml of barium mixed with 50% normal saline) in a supine position and
four in an upright position after ⩾6-h fasting. For Control 2, 22 swallows were
performed during HRIM procedure: 10 supine swallows of 5 ml 50% normal saline, 2
supine swallows of 5 ml 100% normal saline, and 10 upright swallows (5 of 5 ml
50% normal saline, 2 of 5 ml 100% normal saline, 2 of 30 ml 50% normal saline,
and 1 multiple repetitive swallow that was not used in the analysis). The Type-I
Ach subject was studied with 10 supine swallows and 5 upright swallows of 5 ml
50% normal saline.
4D manometry analysis
Overview of 4D manometry analysis
Similar to the previous BFT algorithm, the analysis was applied to each
swallow for each case. A 30-s HRIM data block was imported from ManoView
(Medtronic Inc.) including at least 3 s of pre-swallow baseline data into a
customized MATLAB program along with notations of the upper esophageal
sphincter (UES) channel number, crural diaphragm channel number, swallow
start time, and swallow end time.
Simultaneous
rendering of space-time variation of pressure, impedance, and esophageal
luminal morphology.We first linearly interpolated the 18-channel impedance data to 36-channels
to align the impedance and pressure data both temporally and spatially and
plotted superimposed pressure and impedance color maps. We then identified
three key landmarks: the UES channel, the lower esophageal sphincter (LES)
channel, and the high-pressure zone channel number. The swallow starting
time was identified in the UES channel and set as time 0 in the ensuing
analysis. The swallow end time was based on LES restoration or at 12 s,
based on our previous experience with BFT analysis. The time range of
interest is from time –3 s to the swallow end time (see Figure 1). The spatial-temporal total
conductance, was obtained using the interpolated impedance data. Based
on biophysical principles detailed in the Appendix, the spatial-temporal values of luminal liquid
cross-sectional area (CSA), was then obtained. Note that for a normal swallow with
minimal pre-swallow residual, the mucosal conductance at channels above the
LES, was computed as the minimal value of total conductance (see Appendix). The mucosal conductance of channels below the
LES, where liquid potentially always exists, was approximated by the median
of mucosal conductance in the esophageal body. For swallows with pre-swallow
residual as with Type-I Ach, mucosal conductance at channels with non-zero
residual could be approximated as the median of mucosal conductance at
channels with minimal residual. See Appendix and Discussion for details.
Figure 1.
Pressure-impedance topography concurrent with bolus transit history
of various swallows/cases. The bolus volume within the body is
defined as the total calculated volume of liquid bolus between the
UES and LES channels; the retention volume as the volume of liquid
proximal to the CW channel but distal to the UES channel at each
time.
Pressure-impedance topography concurrent with bolus transit history
of various swallows/cases. The bolus volume within the body is
defined as the total calculated volume of liquid bolus between the
UES and LES channels; the retention volume as the volume of liquid
proximal to the CW channel but distal to the UES channel at each
time.CW, contraction wave; LES, lower esophageal sphincter; UES, upper
esophageal sphincter.With the spatial-temporal values of luminal liquid CSA, , the luminal radius at each time and channel,
was obtained based on catheter radius, , assuming a circular lumen.With the above data, a simultaneous rendering of space-time variation of
pressure, impedance, and luminal morphology, along with the time-history of
bolus volume, was generated (see supplemental movies).
Bolus
transit analysis: quantifying time-history of bolus retention volume and
bolus transit characteristics.At each instant, the channel with the maximal pressure along the esophageal
body was identified as the contraction channel. The bolus retention volume
was defined as the volume of liquid proximal to the contraction channel, and
calculated by integrating bolus CSAs from the UES channel to the contraction
channel at that instant. A time-history of bolus retention volume was then
obtained. Similarly, the time-history of bolus volume within the esophageal
body (i.e., above the crural diaphragm channel) was obtained and used to
identify various phases of bolus transit. The emptying period was defined as
the duration of the effective esophageal emptying (introduced later). The
emptying flux across the EGJ was calculated as the change of esophageal
bolus volume during emptying divided by the emptying time.
EGJ analysis: determining EGJ emptying period, flow characteristics, and
distensibility
An EGJ region of interest (ROI) extended from 1 cm above the crural diaphragm
line to 1 cm below the crural diaphragm line and from –3 s to 12 s. Pressure
data and bolus CSA data within the ROI (three channels, respectively), were then
plotted in conjunction with volume data for simultaneous analysis. An emptying
diameter was introduced as the threshold diameter allowing bolus passage. The
EGJ segment was defined as open when all the three channels within the ROI
equaled or exceeded the emptying diameter. The chosen emptying diameter was
validated during esophageal emptying using the corresponding bolus volume data,
detailed in the EGJ analysis discussion.The first validation included a comparison of calculated bolus volumes with the
actual bolus volumes in Control 2. The second validation was based on the
fluoroscopy study on Control 1 which included 10 swallows of 5 ml mixture of
barium and saline. The X-ray quality, however, was variable among swallows, such
that swallow #7 was most suitable for detailed analysis to delineate the phases
of bolus transit (see Figure
2).
Figure 2.
4D manometry analysis on Control 1 against X-ray image to delineate bolus
transit. (Top) A simultaneous plot of impedance, pressure, bolus volume,
and retention volume history obtained by 4D manometry on Swallow #7 of
Control 1. The delineation of the four-phases of bolus transit was
illustrated based on bolus volume history. (Bottom) 3D rendering of
esophageal luminal morphology corresponding to each phase, compared with
the X-ray images below. Note that the lumen is simplified as a straight
tube during the geometric construction.
3D, three-dimensional; 4D, four-dimensional.
4D manometry analysis on Control 1 against X-ray image to delineate bolus
transit. (Top) A simultaneous plot of impedance, pressure, bolus volume,
and retention volume history obtained by 4D manometry on Swallow #7 of
Control 1. The delineation of the four-phases of bolus transit was
illustrated based on bolus volume history. (Bottom) 3D rendering of
esophageal luminal morphology corresponding to each phase, compared with
the X-ray images below. Note that the lumen is simplified as a straight
tube during the geometric construction.3D, three-dimensional; 4D, four-dimensional.We then focused on the EGJ analysis comparing the 4D manometry analysis with BFT,
esophageal impedance integral (EII) ratio, and the DI from corresponding FLIP
study in Control 2. To quantify the EGJ emptying from 4D manometry, the emptying
diameter was first investigated. The onset of emptying occurred when bolus
volume achieved its maximal value and the emptying diameter defined both the
onset and offset of effective esophageal emptying as well as facilitating the
quantification of EGJ flow. Once the period of esophageal emptying was
identified, the emptying period (comparable with BFT), EGJ pressure, and EGJ
diameter were extracted. Consequently, the mean flow rate (bolus volume
reduction divided by emptying time), and the average emptying velocity (flow
rate divided by EGJ CSA), were evaluated. DI was computed based on EGJ pressure
and diameter during the emptying period and compared with that from the FLIP
study.
Statistical analysis
The HRIM procedure on each subject includes tests in both supine and upright
positions. Hence, for each HRIM study, we defined a subgroup as swallows of the
same volume, the same concentration of saline, and the same body position.
Within each subgroup, we computed medians and interquartile range (see
Tables 1–3).
Results
Validations of 4D manometry calculations
Dynamics of bolus transit and calculated bolus volumes
Figure 1 illustrates
the dynamics of bolus transit including the time-history of the total bolus
volume and bolus retention volume (bottom) concurrent with
pressure-impedance topography (top). For various swallow conditions in
Control 2 (Figure 1,
panels 1–5), bolus intake occurred much more rapidly
(<1.0 s) than emptying (>2.0 s).
Minimal bolus retention was observed, except for an upright swallow of 5 ml
1.0 N supine (Figure
1, panel 4). For swallows in the Type-I Ach case, bolus retention
and pre-swallow residual were observed as shown in Figure 1, panel 6. Table 1 shows the
calculated versus actual bolus volume for Control 1 and
Control 2 swallows. For 5 ml Control 2 swallows, the calculated bolus
volumes ranged from 4.1 ml–6.7 ml, with most cases within 20% error. For
5 ml Control 1 swallows, greater variation was observed (3.2–5.8 ml),
probably because the actual conductance of the barium-bolus mixture deviated
from the assumed 8.12 mS/cm of 0.5 NS. If we work backward from the known
swallowed bolus volume (i.e., 5 ml), and deduce the effective in
vivo conductivity of bolus for Control 1 5 ml upright swallows,
the actual in vivo conductivity of barium-based bolus could
be estimated as around (3.6 ml/5.0 ml) * 8.12 mS/cm = 5.77 mS/cm. For the
two 30 ml swallows, volume was under-predicted with an average error around
27%. The under prediction was likely because a 30-ml bolus was too large and
a portion was likely retained in the mouth. Table 1 also illustrates the
variability of calculated bolus volume with body position, although there
was no consistent effect between subjects. Arguably, the current study
showed that 4D manometry gave reasonable calculations of bolus volume,
especially for standard 5 ml swallows. With a barium-based bolus, 4D
manometry could deduce the actual conductivity of the in
vivo bolus mixture that matched the nominal volume.
Table 1.
Predicted bolus volume and EGJ flow onset time predicted from 4D
manometry against nominal volume and fluoroscopic evaluation,
respectively. Note that EGJ flow onset time was measured by the time
lapse from swallow onset and was computed with two criteria from 4D
manometry. One is based on onset of bolus volume decrease, referred
to as Flow onset time (4D manometry: volume criterion) and the other
uses onset of esophageal emptying based on diameter criteria,
referred to as Flow onset time (4D manometry: diameter criterion).
EGJ emptying diameter was chosen to be 8 mm and 10 mm, for 5 ml
swallow and 30 ml swallow, respectively. Only the first case,
Control 1, included concurrent fluoroscopic data.
Bolus volume from 4D manometry (ml)
Flow onset time (s) (4D manometry: volume
criterion)
Flow onset time (s) (4D manometry: diameter
criterion)
Predicted bolus volume and EGJ flow onset time predicted from 4D
manometry against nominal volume and fluoroscopic evaluation,
respectively. Note that EGJ flow onset time was measured by the time
lapse from swallow onset and was computed with two criteria from 4D
manometry. One is based on onset of bolus volume decrease, referred
to as Flow onset time (4D manometry: volume criterion) and the other
uses onset of esophageal emptying based on diameter criteria,
referred to as Flow onset time (4D manometry: diameter criterion).
EGJ emptying diameter was chosen to be 8 mm and 10 mm, for 5 ml
swallow and 30 ml swallow, respectively. Only the first case,
Control 1, included concurrent fluoroscopic data.4D, four-dimensional; EGJ, esophagogastric junction.
Timing and four phases against fluoroscopy
Figure 2 illustrates
dynamic bolus transit based on simultaneous visualization of luminal
morphology, impedance, and pressure with concurrent fluoroscopy images. The
four phases of swallow can be delineated from the time-history of bolus
transit. Phase I was esophageal filling, wherein impedance decreased and
bolus volume in the body rapidly increased. The filling phase spanned from
UES relaxation until UES tone was restored. The compartmentalization phase
occurred for the next 3.5 s during which LES relaxation occurred as evident
by the impedance and pressure measurement, but the bolus volume in the
esophagus did not decrease. The esophageal emptying began once the bolus
volume started to decrease at 3.8 s post swallow, initially rapid and then
gradual. The initial fast emptying phase corresponded to ‘esophageal
emptying phase’ driven by peristalsis, whereas the gradual emptying
corresponded to ‘ampullary emptying phase’ driven by LES reconstitution.
Interestingly, the onset of esophageal emptying coincided with peak bolus
volume in the body and the onset of the distal esophageal contraction (i.e.,
the end of the transition zone). This coincidence was also observed in other
swallows of normal controls and may have important physiological
implications.
Type-I achalasia versus normal
All 10 swallow tests in the Type I Ach patient showed
>4 ml bolus retention before and after the swallow. A
typical case is shown in Figure 1, panel 6. Also characteristic of these swallows, the
time-history of bolus volume fluctuated due to the absence of an esophageal
contraction and the delineation of four phases of bolus transit was
obscured. Hence, the 4D manometry signature of Type I achalasia is the
morphing of the time-history of pre-swallow bolus volume with post-swallow
bolus retention.
EGJ analysis
EGJ emptying diameter: a threshold defines the effective esophageal
emptying
Figure 3 (5 ml
swallows from Control 2) and Figure 4 (Type-I Ach swallow and
30 ml swallow from Control 2) illustrate examples of simultaneous plots of
bolus volume history, EGJ pressure, and EGJ diameter. Esophageal emptying
consistently occurred when the three EGJ diameters are ⩾8 mm for 5-ml
swallows, and ⩾10 mm for 30-ml swallows suggesting these to be appropriate
thresholds for delineating the period of ‘effective esophageal emptying’.
With the two criteria for esophageal emptying, one based on bolus volume and
the other based on ‘emptying diameter’, the respective emptying onset times
are shown in Table
1. For 5 of the 10 swallows, the time difference between the two
criteria is around 0.4 s, which is small compared with a general 12 s window
for a typical swallow. Table 2 also shows the onset time defined fluoroscopically on
Control 1 which exhibited a larger discrepancy of up to 1 s, likely
attributable to the difficulty of discerning emptying onset
fluoroscopically. Arguably, the proposed diameter-based criteria provided
the most objective way to delineate the period of effective esophageal
emptying.
Figure 3.
EGJ analysis of 5 ml swallows that delineate period of effective
emptying. The concurrent plotting of bolus transit, EGJ pressure and
diameters was used to delineate the period of effective esophageal
emptying and evaluate DI listed in Table 3.
EGJ analysis of one swallow from Type-I Achalasia and one 30 ml
swallow from Control 2. Concurrent plotting of bolus transit, EGJ
pressure, and diameters was used to delineate the period of
effective esophageal emptying and evaluate DI listed in Table 3
for a 30-ml swallow.
EGJ flow characteristics versus BFT, EII for Control
2. Median values and interquartile ranges are shown.BFT, bolus flow time; 4D, four-dimensional; EGJ, esophagogastric
junction; EII, esophageal impedance integral ratio.EGJ analysis of 5 ml swallows that delineate period of effective
emptying. The concurrent plotting of bolus transit, EGJ pressure and
diameters was used to delineate the period of effective esophageal
emptying and evaluate DI listed in Table 3.
Table 3.
Additional novel metrics of emptying from 4D manometry for Control 2
swallows. Median values and interquartile ranges are shown. Median
DI is based on median EGJ pressure and emptying dimeter (i.e., 8 mm
for 5 ml swallow, 10 mm for 30 ml swallow). For comparison, the
Endoflip study of the same subject showed that DI at distension
volume of 30, 40 and 50 ml was 2.3, 8.3 and 5.8 mm2/mmHg,
respectively.
DI, distensibility index; EGJ, esophagogastric junction.EGJ analysis of one swallow from Type-I Achalasia and one 30 ml
swallow from Control 2. Concurrent plotting of bolus transit, EGJ
pressure, and diameters was used to delineate the period of
effective esophageal emptying and evaluate DI listed in Table 3
for a 30-ml swallow.DI, distensibility index; EGJ, esophagogastric junction.
EGJ-flow characteristics: 4D manometry versus BFT; flow
rate and distensibility
Table 2 shows the
4D the EGJ flow characteristics that are comparable with BFT and EII ratio
for Control 2. Specifically, 4D manometry emptying time, defined as the
period of ‘effective esophageal emptying’ (see Figures 3 and 4) is analogues to the BFT. Compared
with BFT analysis, the emptying time predicted by 4D manometry was more
consistent among swallow types. Except for the 30 ml swallow and the upright
swallow with 5 ml 0.5 N saline, the emptying time ranged from 1.2 s to
2.2 s. In contrast, the BFT ranged from 0.5 s to 3 s. Within the 4D
manometry calculation, 30 ml swallows showed longer emptying time as
expected, whereas upright swallow did not necessarily lead to shorter
emptying time. An analogue to the E-II ratio is the bolus retention ratio,
defined as maximal retention volume divided by the normal bolus volume. For
all Control 2 swallows, the bolus retention ratio was consistently minimal
(<0.1 in most of cases) in contrast with the EII,
which ranged from 0.08 to 0.58. For the Type I Ach case, all swallows,
except the one showed in Figure 4, panel 1, showed no effective esophageal emptying.
Figure 4, panel
1 shows a very short period of emptying, followed by LES restoration leaving
a large amount of retention (see also Figure 1, panel 6).Additional novel metrics of emptying calculated by 4D manometry are shown in
Table 3. For
5 ml swallows, average flow rate ranged from 1.3 to 2.7 ml/s, whereas the
maximal flow rate was as high as 10.5 ml/s, likely occurring at the very
beginning of emptying. The median EGJ pressure ranged from 15 to 20 mmHg and
the EGJ DI ranged from 2.4 to 3.4 mm2/mmHg for 5 ml swallows and
4.2–4.6 mm2/mmHg for 30 ml swallows, respectively. A greater
DI for 30 ml swallow is expected due to a larger emptying diameter and these
values are more consistent with DI measured from FLIP study of the same
subject, which showed DI of 2.3 mm2/mmHg at 30 ml distension
volume, and DI = 5.8 mm2/mmHg at the 50 ml volume.Additional novel metrics of emptying from 4D manometry for Control 2
swallows. Median values and interquartile ranges are shown. Median
DI is based on median EGJ pressure and emptying dimeter (i.e., 8 mm
for 5 ml swallow, 10 mm for 30 ml swallow). For comparison, the
Endoflip study of the same subject showed that DI at distension
volume of 30, 40 and 50 ml was 2.3, 8.3 and 5.8 mm2/mmHg,
respectively.4D, four-dimensional; DI, distensibility index; EGJ,
esophagogastric junction.
Discussion
In this work, we describe a analytic methodology, 4D manometry, for post-processing
HRIM data to dynamically characterize and visualize esophageal luminal geometry,
contractility, and bolus transport during test swallows. Compared with conventional
manometry, 4D manometry provides a comprehensive characterization of
temporal-spatial dynamics of bolus transit (see supplemental movies). This approach facilitates quantification of
bolus flow within and through the esophagus as well as visualizing and quantifying
esophageal wall distensibility. Furthermore, 4D manometry requires minimal
additional information for HRIM post-processing. Specifically, if only the swallow
bolus volume is known, the current method can even deduce an effective bolus
resistance in vivo, as illustrated in Appendix Equation (7).The current work is not the first study to construct esophageal luminal geometry from
impedance measurement. FLIP is a related technique that uses impedance sensors
within a cylindrical bag filled with conductive fluid to CSA measurements. However,
with respect to motility the current FLIP procedure has only a single pressure
sensor and is used primarily to assess secondary peristalsis. More analogous to the
current study, Kim et al. used HRIM impedance data with concurrent
ultrasound imaging of the esophageal wall to study the correlation between luminal
CSA and impedance change describing a qualitative inverse relationship between
impedance and peak CSA.[14] Zifan et al. proposed a methodology to quantify CSA from
impedance data, which required swallowing boluses of two different conductivity
values in order to derive tissue conductivity.[15] With that method, they investigated contraction-distention topography along
the entire esophagus during peristalsis assisted with the ultrasound.[16] The current 4D manometry differs from these previous approaches in several
ways. First, it relies only on standard HRIM measurements of each swallow as long as
the bolus volume or conductivity is known. This provides great applicability of 4D
manometry to post-processing existing HRIM data. In particular, with known swallow
bolus volume, the current method can circumvent the challenge of measuring bolus
conductivity in vivo, which may differ from in
vitro conductivity due to mixing between the bolus luminal fluids.
Second, the proposed 4D manometry provides more comprehensive and innovate metrics
of esophageal morphology and bolus transport within the body and through the EGJ.
Specifically, 4D manometry provides concurrent spatial-temporal representations of
luminal geometry and pressure, and the time-history of bolus volume and bolus
retention volume during peristalsis. Subsequent EGJ analysis enables the
quantification of emptying time, mean bolus velocity, flow rate as well as EGJ DI.
In overall, the range of EGJ DI values calculated with 4D manometry were very
similar to the EGJ DI values calculated using FLIP in asymptomatic controls,[10] but certain discrepancies between DI from 4D manometry and FLIP are also
observed. In particular, at volume of 30 ml, DI from 4D manometry is greater than DI
from FLIP (4.4 versus 2.3). This is likely related to the
difference of EGJ physiology during normal emptying in the 4D manometry analysis and
that measured during volume-based dilation in FLIP when esophageal distention is
likely stimulating contraction.BFT and EII-ratio are other metrics developed to illustrate bolus retention and flow
based on impedance data. Compared with EII-ratio, 4D manometry provides a more
quantitative evaluation of bolus retention including its time-history and an
overview of dynamic bolus transit. In comparison with the BFT, 4D manometry
calculates the period of effective esophageal emptying from impedance data only
circumventing the difficulty of artifact pressure transients attributable to contact
between EGJ wall and the pressure sensors.Validations of 4D manometry in the current study found reasonable correlation between
calculated and actual bolus volume, but the error was not small in some cases.
Potential sources of this error include: (1) change of effective conductivity of the
bolus due to in vitro dilution, (2) approximation error from
assuming a circular luminal shape, and (3) catheter or EGJ axial movement. The first
two factors are expected to be the dominant ones, whereas the third factor could be
corrected by visual inspection of the pressure topography. In instance of known
bolus volume, the first factor can also be fixed using the deduced effective bolus
conductivity based on Appendix Equation (7). Zifan et al. mentioned
another potential source of error related to luminal air which has two aspects, its
effect on conductance [Equation (1)] and its effect on luminal CSA. The effect on
conductance can be safely ignored, as the conductivity of air (~10−9 S/m)
is about seven orders of magnitude greater than that of drinking water or saline
(~10−2 S/m). The influence of air on luminal CSA can be sorted out by
always interpreting the calculated luminal CSA as that of the
swallowed bolus (i.e., liquid CSA).As a preliminary study, one limitation of the current work is the small number of
cases included. More studies with comprehensive measurements from both HRIM and
fluoroscopy could help to refine several outcomes. However, this would require a
fairly complicated procedure of HRIM synchronized with a concurrent fluoroscopy that
we are currently unable to conduct. Consequently, this preliminary study illustrates
a new analysis algorithm, and we plan on studying larger cohorts of patients and
controls in future work.In summary, the proposed 4D manometry tool, based on simple biophysical analysis,
enables a comprehensive characterization of temporal-spatial dynamics of esophageal
bolus transit with concurrent pressure and luminal geometry. Further biomechanical
analysis can deduce additional metrics such as the flow rate, esophageal wall
distensibility, and esophageal retention. The methodology was validated against
volume predictions on fluoroscopy, and the DI prediction from Endoflip. Preliminary
studies showed 4D manometry could easily differentiate normal from Type I Achalasia,
which featured pre- and post-swallow residual and minimal EGJ opening. Large cohort
studies will be conducted in future to determine whether 4D manometry can further
refine motility diagnoses that are currently heterogeneous such as ineffective
esophageal motility and EGJ outflow obstruction.Click here for additional data file.Supplemental material, 4D_manometry_appendix_supplement for Four-dimensional
impedance manometry derived from esophageal high-resolution impedance-manometry
studies: a novel analysis paradigm by Wenjun Kou, Dustin A. Carlson, Neelesh A.
Patankar, Peter J. Kahrilas and John E. Pandolfino in Therapeutic Advances in
Gastroenterology
Authors: Sudip K Ghosh; Patrick Janiak; Werner Schwizer; Geoffrey S Hebbard; James G Brasseur Journal: Am J Physiol Gastrointest Liver Physiol Date: 2005-11-10 Impact factor: 4.052
Authors: Wenjun Kou; Boyce E Griffith; John E Pandolfino; Peter J Kahrilas; Neelesh A Patankar Journal: J Comput Phys Date: 2017-07-18 Impact factor: 3.553
Authors: Wenjun Kou; John E Pandolfino; Peter J Kahrilas; Neelesh A Patankar Journal: Am J Physiol Gastrointest Liver Physiol Date: 2015-06-25 Impact factor: 4.052
Authors: P J Kahrilas; A J Bredenoord; M Fox; C P Gyawali; S Roman; A J P M Smout; J E Pandolfino Journal: Neurogastroenterol Motil Date: 2014-12-03 Impact factor: 3.598
Authors: Wenjun Kou; Dustin A Carlson; Peter J Kahrilas; Neelesh A Patankar; John E Pandolfino Journal: Neurogastroenterol Motil Date: 2022-06-05 Impact factor: 3.960