Paramarshi Das1, D V Swapna1, Roopa R Nadig1.
Abstract
AIMS: The aim of this study is to evaluate and compare the surface deformation and cyclic fatigue resistance of Hyflex controlled memory (CM) and Hyflex electric discharge machining (EDM) files following instrumentation and heat sterilization. SUBJECTS AND METHODS: Thirty Hyflex CM and thirty Hyflex EDM files were selected and profilometric images taken. Sixty extracted molars were decoronated, cleaning and shaping done with files, and subjected to autoclaving. Profilometric images were taken after instrumentation and sterilization. Cyclic fatigue testing was done at 30° and 60° in a custom-made jig. STATISTICAL ANALYSIS USED: The Mann-Whitney test was used to compare profilometer and cyclic fatigue values between Hyflex CM and Hyflex EDM groups, whereas Friedman's and Wilcoxon signed-rank test were used for intragroup comparisons.Entities:
Keywords: Cyclic fatigue; Hyflex controlled memory; Hyflex electric discharge machining; optical profilometry; sterilization
Year: 2020 PMID: 33082663 PMCID: PMC7537757 DOI: 10.4103/JCD.JCD_39_20
Source DB: PubMed Journal: J Conserv Dent ISSN: 0972-0707
Mean and Standard Deviations of Sa, Sq, and Sz parameters before use, after use, and after sterilization of HyFlex CM and Hyflex EDM NiTi Instruments (in μm)
| Hyflex CM | Hyflex EDM | |||||||
|---|---|---|---|---|---|---|---|---|
| Cutting blade | Flute | Cutting blade | Flute | |||||
| Mean±SD | Mean±SD | Mean±SD | Mean±SD | |||||
| Sa | ||||||||
| Before use | 21456.03±3336.96x | BU versus AU (<0.001*) | 4135.03±709.95x | BU versus AU (<0.001*) | 70985.03±12777.19y | BU versus AU (0.20) | 13568±2128.06y | BU versus AU (0.18) |
| After use | 42897.03±6004.09x | BU versus AS (<0.001*) | 7988.03±940.85x | BU versus AS (<0.001*) | 77980.07±16260.27y | BU versus AS (0.20) | 14743.07±3638.36y | BU versus AS (0.18) |
| After sterilisation | 41784.93±3841.22x | AU versus AS (0.22) | 7586.97±1015.08x | AU versus AS (0.13) | 72344.97±11188.90y | AU versus AS (0.20) | 14167.03±2025.18y | AU Versus AS (0.18) |
| Sq | ||||||||
| Before use | 29878.9±4795.90x | BU Versus AU (<0.001*) | 4987±656.36x | BU versus AU (<0.001*) | 86349.03±16635.09y | BU versus AU (0.36) | 18776.07±3671.42y | BU versus AU (0.36) |
| After use | 58097.97±6374.01x | BU versus AS (<0.001*) | 10896.97±925.66x | BU versus AS (<0.001*) | 92098.03±14177.78y | BU versus AS (0.36) | 19908.1±4790.40y | BU versus AS (0.36) |
| After sterilisation | 54667.93±6427.33x | AU versus AS (>0.05) | 10196±1525.67x | AU versus AS (>0.05) | 88653.97±14221.23y | AU versus AS (0.36) | 19450.07±2366.97y | AU versus AS (0.36) |
| Sz | ||||||||
| Before use | 154675.07±25568.79x | BU versus AU (<0.001*) | 45763.07±5686.22x | BU versus AU (<0.001*) | 446873.97±86468.36y | BU versus AU (0.88) | 123359.97±16797.78y | BU versus AU (0.50) |
| After use | 272785.03±46805.70x | BU versus AS (<0.001*) | 119853.97±18968.44x | BU versus AS (<0.001*) | 460895±56461.79y | BU versus AS (0.88) | 125876.03±25218.54x | BU versus AS (0.50) |
| After sterilisation | 264668.1±35896.17x | AU versus AS (0.35) | 112433±15741.00x | AU versus AS (1.63) | 449761.07±68543.50y | AU versus AS (0.88) | 125752.9±16333.23y | AU versus AS (0.50) |
*Statistically Significant, Different superscript letters (xy) indicate statistically significant differences at 5% level for intergroup comparison between Hyflex CM and Hyflex EDM. SD: Standard deviation, EDM: Electric discharge machining, CM: Control memory
Figure 1Profilometric image of Hyflex controlled memory files: (a) Before use; (b) After use; (c) After sterilization
Figure 2Profilometric image of Hyflex electric discharge machining files: (a) Before use; (b) After use; (c) After sterilization
Mean number of cycles to failure and mean length of fractured fragment (in mm) between different curvatures in Hyflex CM and Hyflex EDM groups
| Mean±SD | ||
|---|---|---|
| NCF | Length of fractured fragment (in mm) | |
| Hyflex CM | ||
| 30° | 2228.07±140.54ax | 6.87±0.74ax |
| 60° | 1406.4±124.23bx | 7.06±0.64bx |
| Hyflex EDM | ||
| 30° | 4603.27±423.05ay | 5.58±0.52ay |
| 60° | 3373.47±306.44by | 5.73±0.33ay |
Different superscript letters indicate statistically significant differences at 5% level (ab, for intragroup comparison; xy for intergroup comparison). NCF: Number of cycles to failure, SD: Standard deviation, EDM: Electric discharge machining, CM: Control memory