Can we measure mesopic pupil size with the cobalt blue light slit-lamp biomicroscopy method?

Miguel J Maldonado1, Alberto López-Miguel, David P Piñero, José R Juberías, Juan C Nieto, Jorge L Alió.   

Abstract

BACKGROUND: The aim of this work is to assess a previously described slit-lamp biomicroscopy-based method (SLBM) for measuring pupil diameter and compare it to Colvard infrared pupillometry (CIP).
METHODS: Two examiners performed three repeated measurements with each instrument in 40 healthy eyes. We determined the agreement of SLBM and CIP, intraobserver and interobserver repeatabilities, and interobserver concordance (kappa) and SLBM ability for detecting pupil sizes over 6.0 mm.
RESULTS: The mean (±standard deviation [SD]) pupil diameter was 5.81 ± 0.70 mm with SLBM and 6.26 ± 0.68 mm with CIP (p = 0.01) averaging both examiner's results. Mean differences between the SLBM and CIP were -0.60 mm and -0.30 mm for each examiner using the average of the three readings (p = 0.02), and they were very similar using the first reading. Intraobserver reproducibility: the width of the 95% LoA ranged from 1.79 to 2.30 mm. The ICCs were 0.97 and 0.92 for SLBM, and 0.96 and 0.90 for CIP. Interobserver reproducibility: the width of the LoA ranged from 1.82 to 2.09 mm. Kappa statistics were 0.39 and 0.49 for the first and mean SLBM readings, respectively, and 0.45 for both the first and mean CIP readings. Sensitivity and specificity of SLBM for detection of pupils larger than 6 mm ranged from 55.56% to 73.68% and from 76.19% to 95.45%, respectively. The best trade-off between sensitivity and specificity ranged from 5.4 mm to 6.2 mm.
CONCLUSIONS: Although the SLBM is quite repeatable, it underestimates mesopic pupil size and shows a too wide range of agreement with CIP. SLBM shows low sensitivity in detecting pupils larger than 6 mm, which may be misleading when planning anterior segment surgery. Previous grading-consensus training strategies may increase interrater reproducibility, and compensation for the systematic underestimation could improve accuracy of the SLBM.

Entities:  

Mesh:

Year:  2012        PMID: 22227642     DOI: 10.1007/s00417-011-1909-0

Source DB:  PubMed          Journal:  Graefes Arch Clin Exp Ophthalmol        ISSN: 0721-832X            Impact factor:   3.117


  54 in total

1.  Comparison of the pupil card and pupillometer in measuring pupil size.

Authors:  Mihai Pop; Yves Payette; Emma Santoriello
Journal:  J Cataract Refract Surg       Date:  2002-02       Impact factor: 3.351

2.  Comparison of scotopic pupil measurement with slitlamp-based cobalt blue light and infrared video-based system.

Authors:  Tomy Starck; Yun Liu; Amy L Prewett; Lettie G Curup
Journal:  J Cataract Refract Surg       Date:  2002-11       Impact factor: 3.351

3.  Effect of preoperative pupil size on quality of vision after wavefront-guided LASIK.

Authors:  Annie Chan; Edward E Manche
Journal:  Ophthalmology       Date:  2010-11-20       Impact factor: 12.079

4.  Pupil measurement using the Colvard pupillometer and a standard pupil card with a cobalt blue filter penlight.

Authors:  Elyse L Chaglasian; Sabah Akbar; Louis E Probst
Journal:  J Cataract Refract Surg       Date:  2006-02       Impact factor: 3.351

5.  Comparison of Rosenbaum pupillometry card using red and blue light to Colvard and Iowa pupillometers.

Authors:  Lawrence Y Ho; Thomas M Harvey; Joseph Scherer; Mamtha Balasubramaniam; Deepinder K Dhaliwal; Francis S Mah
Journal:  J Refract Surg       Date:  2010-07       Impact factor: 3.573

6.  Functional outcome and patient satisfaction after Artisan phakic intraocular lens implantation for the correction of myopia.

Authors:  Nayyirih G Tahzib; Sander J Bootsma; Fred A G J Eggink; Rudy M M A Nuijts
Journal:  Am J Ophthalmol       Date:  2006-07       Impact factor: 5.258

7.  Measurement error and correlation coefficients.

Authors:  J M Bland; D G Altman
Journal:  BMJ       Date:  1996-07-06

8.  Statistical analysis of multi-eye data in ophthalmic research.

Authors:  W A Ray; D M O'Day
Journal:  Invest Ophthalmol Vis Sci       Date:  1985-08       Impact factor: 4.799

9.  Pre-excimer laser and post-excimer laser refractive surgery measurements of scotopic pupil diameter using 2 pupillometers.

Authors:  Leopoldo Spadea; Daniele Giammaria; Roberto Ferrante; Emilio Balestrazzi
Journal:  Ophthalmology       Date:  2005-06       Impact factor: 12.079

10.  Risk factors for night vision complaints after LASIK for myopia.

Authors:  Mihai Pop; Yves Payette
Journal:  Ophthalmology       Date:  2004-01       Impact factor: 12.079

View more
  2 in total

1.  Slitlamp Photography and Videography With High Magnifications.

Authors:  Jin Yuan; Hong Jiang; Xinjie Mao; Bilian Ke; Wentao Yan; Che Liu; Hector R Cintrón-Colón; Victor L Perez; Jianhua Wang
Journal:  Eye Contact Lens       Date:  2015-11       Impact factor: 2.018

2.  Intrasession repeatability of pupil size measurements under different light levels provided by a multidiagnostic device in healthy eyes.

Authors:  David P Piñero; Dolores de Fez; Inmaculada Cabezos; Alberto López-Navarro; María T Caballero; Vicente J Camps
Journal:  BMC Ophthalmol       Date:  2020-08-31       Impact factor: 2.209

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.