Intrasession Repeatability and Intersession Reproducibility Measurements Using VX120 Multidiagnostic Unit.

Irene Sanchez1, Sara Ortiz-Toquero, Raul Martin.   

Abstract

OBJECTIVE: The VX120 multidiagnostic unit is a multidiagnostic instrument that combines several functions: autorefraction, keratometry, corneal topography, aberrometry, pachymetry, and noncontact tonometry. The purpose of this study was to determine the intrasession repeatability and the intersession reproducibility of all parameters measured by the VX120 multidiagnostic unit in a sample of normal healthy eyes.
METHODS: Three repeated measurements in the right eye of the volunteers were taken with VX120. Repeatability of the sphere, cylinder, axis, anterior corneal powers (K flat and K steep, 3-mm pupil), high- and low-order aberration root mean square (HOA-RMS and LOA-RMS, 3-mm pupil), eccentricity, white-to-white (WTW), anterior chamber depth (ACD), and central corneal thickness (CCT) (2-mm central) was analyzed. Within-subject SD (Sw), precision, repeatability, coefficient of variation (CV), and the intraclass correlation coefficient (ICC) were calculated.
RESULTS: The CV was low for K flat and K steep, WTW, ACD, and CCT with a range from 0.34% to 1.16%. The CV was higher for sphere, cylinder, HOA-RMS, and LOA-RMS and eccentricity with a range among 6.92% to 54.24%. The ICC showed high values in all parameters except in HOA-RMS (0.720-0.776) and eccentricity (0.889) in first session with moderate agreement. Comparing the intrasession repeatability of first and second session, statistically significant differences (P<0.01) were found between both sessions just to the CV for all parameters (except cylinder values) measured with VX120. However, nonstatistically significant differences (P>0.13) were found for Sw, precision, and repeatability values.
CONCLUSIONS: The VX120 multidiagnostic unit provides repeatable measurements in anterior corneal power (K flat and K steep), WTW, ACD, and CCT. However sphere, cylinder, HOA-RMS, LOA-RMS, and eccentricity showed worse repeatability. Intersession reproducibility showed good results with little differences between sessions in healthy subjects.

Entities:  

Mesh:

Year:  2018        PMID: 29461300     DOI: 10.1097/ICL.0000000000000486

Source DB:  PubMed          Journal:  Eye Contact Lens        ISSN: 1542-2321            Impact factor:   2.018


  6 in total

1.  Near Vision Tasks and Optical Quality of the Eye.

Authors:  Jessica Rafaela Moreira Gomes; Sandra Maria de Braga Franco
Journal:  J Ophthalmic Vis Res       Date:  2021-10-25

2.  The prevalence of refractive errors in college students in Israel.

Authors:  Einat Shneor; Ravid Doron; Lisa A Ostrin; Ariela Gordon-Shaag
Journal:  J Optom       Date:  2021-12-28

3.  The associations of high academic performance with childhood ametropia prevalence and myopia development in China.

Authors:  Yahan Yang; Ruiyang Li; Daniel Ting; Xiaohang Wu; Jialing Huang; Yi Zhu; Chuan Chen; Bingsen Lin; Sijin Li; Xinliang Zhang; Kexin Chen; Tongyong Yu; Dongxuan Wu; Zijun Mo; Hongxi Wang; Shiqun Li; Haotian Lin
Journal:  Ann Transl Med       Date:  2021-05

4.  Automatic identification of myopia based on ocular appearance images using deep learning.

Authors:  Yahan Yang; Ruiyang Li; Duoru Lin; Xiayin Zhang; Wangting Li; Jinghui Wang; Chong Guo; Jianyin Li; Chuan Chen; Yi Zhu; Lanqin Zhao; Haotian Lin
Journal:  Ann Transl Med       Date:  2020-06

5.  Intrasession repeatability of pupil size measurements under different light levels provided by a multidiagnostic device in healthy eyes.

Authors:  David P Piñero; Dolores de Fez; Inmaculada Cabezos; Alberto López-Navarro; María T Caballero; Vicente J Camps
Journal:  BMC Ophthalmol       Date:  2020-08-31       Impact factor: 2.209

6.  Relationship between Axial Length and Corneo-Scleral Topography: A Preliminary Study.

Authors:  Laurent Bataille; Ainhoa Molina-Martín; David P Piñero
Journal:  Diagnostics (Basel)       Date:  2021-03-18
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.