Agreement and repeatability of infrared pupillometry and the comparison method.

B S Wachler1, R R Krueger.   

Abstract

OBJECTIVE: To evaluate the accuracy and repeatability of the widely used comparison method of measuring pupil size.
DESIGN: Cross-sectional study. PARTICIPANTS: Fourteen eyes of seven healthy myopic subjects were examined. INTERVENTION: Two examiners made two repeated measures of pupil diameters of 14 eyes using Rosenbaum card comparison pupillometry and infrared pupillometry. Subjects fixated on a distant visual acuity chart, and pupils were measured under three luminance conditions. The agreement and inter-rater repeatability of both methods were determined. MAIN OUTCOME MEASURES: Outcomes were pupil diameters, limits of agreement, and coefficient of repeatability of two examiners.
RESULTS: The mean difference between the two techniques ranged from 0.3 to 0.5 mm. The limits of agreement within two standard deviations ranged from 2.4 to 2.8 mm. Coefficient of repeatability ranged from 0.6 to 1.4 mm for infrared pupillometry and 1.0 to 1.2 mm for Rosenbaum pupillometry. Inter-rater repeatability of Rosenbaum pupillometry was consistently pupil diameter biased. Pupil diameters measured with the Rosenbaum method were consistently larger than diameters measured with the infrared technique for both examiners under all luminance conditions.
CONCLUSIONS: Results indicate that although the mean difference in techniques was small, the range of the agreement between the Rosenbaum and the infrared techniques was large. The Rosenbaum method consistently overestimated pupil diameters and was subject to inter-rater repeatability bias. Rosenbaum pupillometry may not be appropriate when accurate pupil measurements are required. The results have implications for many clinical trials in ophthalmology, including those evaluating refractive surgery that use Rosenbaum comparison pupillometry.

Entities:  

Mesh:

Year:  1999        PMID: 9951484     DOI: 10.1016/s0161-6420(99)90070-2

Source DB:  PubMed          Journal:  Ophthalmology        ISSN: 0161-6420            Impact factor:   12.079


  14 in total

1.  A simple infrared-augmented digital photography technique for detection of pupillary abnormalities.

Authors:  Tarek A Shazly; G R Bonhomme
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2015-01-15       Impact factor: 3.117

2.  Immediate effects of a thoracic spine thrust manipulation on the autonomic nervous system: a randomized clinical trial.

Authors:  Rob Sillevis; Joshua Cleland; Madeleine Hellman; Kristina Beekhuizen
Journal:  J Man Manip Ther       Date:  2010-12

3.  Clinical implications of quantitative infrared pupillometry in neurosurgical patients.

Authors:  Kostas N Fountas; Eftychia Z Kapsalaki; Theofilos G Machinis; Angel N Boev; Joe S Robinson; E Christopher Troup
Journal:  Neurocrit Care       Date:  2006       Impact factor: 3.210

4.  Can we measure mesopic pupil size with the cobalt blue light slit-lamp biomicroscopy method?

Authors:  Miguel J Maldonado; Alberto López-Miguel; David P Piñero; José R Juberías; Juan C Nieto; Jorge L Alió
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2012-01-08       Impact factor: 3.117

5.  Dysfunction of the pupillary light reflex in experimental autoimmune autonomic ganglionopathy.

Authors:  Shalini Mukherjee; Steven Vernino
Journal:  Auton Neurosci       Date:  2007-07-05       Impact factor: 3.145

6.  Clear-cornea cataract surgery: pupil size and shape changes, along with anterior chamber volume and depth changes. A Scheimpflug imaging study.

Authors:  Anastasios John Kanellopoulos; George Asimellis
Journal:  Clin Ophthalmol       Date:  2014-10-24

7.  Infrared pupillometry, the Neurological Pupil index and unilateral pupillary dilation after traumatic brain injury: implications for treatment paradigms.

Authors:  Jefferson William Chen; Kiana Vakil-Gilani; Kay Lyn Williamson; Sandy Cecil
Journal:  Springerplus       Date:  2014-09-23

8.  The Reliability, Validity, and Normative Data of Interpupillary Distance and Pupil Diameter Using Eye-Tracking Technology.

Authors:  Nicholas P Murray; Melissa Hunfalvay; Takumi Bolte
Journal:  Transl Vis Sci Technol       Date:  2017-07-03       Impact factor: 3.283

9.  Analysis of various factors affecting pupil size in patients with glaucoma.

Authors:  Ji Woong Park; Bong Hui Kang; Ji Won Kwon; Kyong Jin Cho
Journal:  BMC Ophthalmol       Date:  2017-09-16       Impact factor: 2.209

10.  B-mode ultrasound assessment of pupillary function: Feasibility, reliability and normal values.

Authors:  Felix A Schmidt; Klemens Ruprecht; Florian Connolly; Matthew B Maas; Friedemann Paul; Jan Hoffmann; Lutz Harms; Stephan J Schreiber
Journal:  PLoS One       Date:  2017-12-06       Impact factor: 3.240

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.